IOVS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


(Investigative Ophthalmology and Visual Science. 1971;10:592-600.)
© 1971 by The Association for Research in Vision and Ophthalmology, Inc.

This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by MOSES, R. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by MOSES, R. A.

Theory and Calibration of the Schiøtz Tonometer

IV. A Mathematical Model of Schiotz Tonometry Scale Reading and tonometric Pressure

ROBERT A. MOSES 1

1 Department of Ophthalmology, Washington University School of Medicine, 660 South Euclid, Saint Louis, Mo. 63110

A geometrical model of the Schiøtz tonometer indentation is proposed. Calculations based on the model predict relationships of scale reading to tonometric pressure, plunger weight, and radius of corneal curvature. Change of the general pattern of the relationships at low-scale readings is discussed.

Key Words: scale reading • tonometric pressure • plunger weight • corneal curvature

Submitted on April 27, 1971
Accepted on July 15, 1971