1From the Department of Ophthalmology and the 2Horten Center for Patient-Oriented Research, University of Zurich, Zurich, Switzerland.
PURPOSE. Thinning of the corneal stroma by laser in situ keratomileusis (LASIK) results in inaccurate low intraocular pressure (IOP) readings by Goldmann applanation tonometry (GAT). Dynamic contour tonometry (DCT) is a novel measuring technique, designed to measure IOP largely independent of corneal thickness and curvature. The purpose of this study was to compare IOP measurements using GAT and DCT in eyes undergoing LASIK for correction of myopia.
METHODS. In a prospective, single-center study, central corneal thickness (CCT) and IOP were measured in patients undergoing first-time LASIK for myopia. IOP was measured before and after surgery using GAT and DCT. The untreated contralateral eyes served as paired controls.
RESULTS. There was good concordance between the two tonometers in 62 normal eyes before LASIK. Corneal ablation of 90.0 ± 49.18 µm (median ± SD) reduced IOP readings as measured by GAT by 3.0 ± 1.9 mm Hg (P < 0.001). In contrast, no significant change in IOP readings was recorded by DCT (-0.2 mm Hg ± 1.5 mm Hg, P = 0.30). There was no change in IOP in the untreated control eyes as measured by GAT and DCT.
CONCLUSIONS. Significant decreases in IOP were recorded by GAT after LASIK for myopia. Measurements by DCT, however, did not reveal any significant changes in IOP.
This article has been cited by other articles:
![]() |
T. Romppainen, L. M. Bachmann, C. Kaufmann, C. Kniestedt, M. Mrochen, and M. A. Thiel Effect of Riboflavin-UVA Induced Collagen Cross-linking on Intraocular Pressure Measurement Invest. Ophthalmol. Vis. Sci., December 1, 2007; 48(12): 5494 - 5498. [Abstract] [Full Text] [PDF] |
||||
![]() |
A R Ismail, M Lamont, S Perera, D Khan-Lim, R Mehta, J D A Macleod, and D F Anderson Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties Br. J. Ophthalmol., July 1, 2007; 91(7): 980 - 981. [Full Text] [PDF] |
||||
![]() |
M. C. Grieshaber, A. Schoetzau, C. Zawinka, J. Flammer, and S. Orgul Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle Glaucoma Arch Ophthalmol, June 1, 2007; 125(6): 740 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Martinez-de-la-Casa, J. Garcia-Feijoo, A. Fernandez-Vidal, C. Mendez-Hernandez, and J. Garcia-Sanchez Ocular Response Analyzer versus Goldmann Applanation Tonometry for Intraocular Pressure Measurements. Invest. Ophthalmol. Vis. Sci., October 1, 2006; 47(10): 4410 - 4414. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kaufmann, L. M. Bachmann, Y. C. Robert, and M. A. Thiel Ocular Pulse Amplitude in Healthy Subjects as Measured by Dynamic Contour Tonometry. Arch Ophthalmol, August 1, 2006; 124(8): 1104 - 1108. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Kotecha, E T White, J M Shewry, and D F Garway-Heath The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry Br. J. Ophthalmol., December 1, 2005; 89(12): 1572 - 1575. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kniestedt, S. Lin, J. Choe, A. Bostrom, M. Nee, and R. L. Stamper Clinical Comparison of Contour and Applanation Tonometry and Their Relationship to Pachymetry Arch Ophthalmol, November 1, 2005; 123(11): 1532 - 1537. [Abstract] [Full Text] [PDF] |
||||
![]() |
P-A Tonnu, T Ho, T Newson, A El Sheikh, K Sharma, E White, C Bunce, and D Garway-Heath The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry Br. J. Ophthalmol., July 1, 2005; 89(7): 851 - 854. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Chihara, H Takahashi, K Okazaki, M Park, and M Tanito The preoperative intraocular pressure level predicts the amount of underestimated intraocular pressure after LASIK for myopia Br. J. Ophthalmol., February 1, 2005; 89(2): 160 - 164. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kaufmann, L. M. Bachmann, and M. A. Thiel Comparison of Dynamic Contour Tonometry with Goldmann Applanation Tonometry Invest. Ophthalmol. Vis. Sci., September 1, 2004; 45(9): 3118 - 3121. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |