1From the Department of Ophthalmology and the 2Horten Centre, University of Zurich, Zurich, Switzerland.
PURPOSE. The dynamic contour tonometer (DCT; Pascal tonometer) is a novel tonometer designed to measure intraocular pressure (IOP) independent of corneal properties. The purpose of this study was a comparison of the DCT with the Goldmann applanation tonometer (GAT) with respect to mean of IOP readings, the influence of ocular structural factors on IOP readings, and both intra- and interobserver variability, in a large group of healthy subjects.
METHODS. In a prospective study of 228 eyes, IOP measurements by GAT and DCT were compared, and the effects of central corneal thickness (CCT), corneal curvature, axial length, and anterior chamber depth were analyzed. To evaluate intra- and interobserver variability, IOP was measured in eight eyes by four observers.
RESULTS. There was a high concordance between the IOP readings obtained by DCT and GAT. However, IOP readings were consistently higher with DCT than with GAT (median difference: +1.7 mm Hg, interquartile range [25th75th percentile] = 0.82.7 mm Hg). In contrast to GAT, multivariable regression analysis showed no significant effect of corneal thickness, corneal curvature, astigmatism, anterior chamber depth, and axial length on DCT readings. For repeated measurements the intraobserver variability was 0.65 mm Hg for the DCT and 1.1 mm Hg for the GAT (P = 0.008). Interobserver variability was 0.44 mm Hg for the DCT and 1.28 mm Hg for the GAT (P = 0.017).
CONCLUSIONS. IOP measurements by DCT are highly concordant with IOP readings obtained from GAT but do not vary in CCT and have a lower intra- and interobserver variability. DCT seems to be an appropriate method of tonometry for routine clinical use.
This article has been cited by other articles:
![]() |
S. A. Read, M. J. Collins, and D. R. Iskander Diurnal Variation of Axial Length, Intraocular Pressure, and Anterior Eye Biometrics Invest. Ophthalmol. Vis. Sci., July 1, 2008; 49(7): 2911 - 2918. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. G. Boehm, A. Weber, L. E. Pillunat, R. Koch, and E. Spoerl Dynamic Contour Tonometry in Comparison to Intracameral IOP Measurements Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2472 - 2477. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Romppainen, L. M. Bachmann, C. Kaufmann, C. Kniestedt, M. Mrochen, and M. A. Thiel Effect of Riboflavin-UVA Induced Collagen Cross-linking on Intraocular Pressure Measurement Invest. Ophthalmol. Vis. Sci., December 1, 2007; 48(12): 5494 - 5498. [Abstract] [Full Text] [PDF] |
||||
![]() |
A R Ismail, M Lamont, S Perera, D Khan-Lim, R Mehta, J D A Macleod, and D F Anderson Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties Br. J. Ophthalmol., July 1, 2007; 91(7): 980 - 981. [Full Text] [PDF] |
||||
![]() |
M. C. Grieshaber, A. Schoetzau, C. Zawinka, J. Flammer, and S. Orgul Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle Glaucoma Arch Ophthalmol, June 1, 2007; 125(6): 740 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Cervino Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure. Br. J. Ophthalmol., December 1, 2006; 90(12): 1444 - 1446. [Full Text] [PDF] |
||||
![]() |
J M Gonzalez-Meijome, J Jorge, A Queiros, P Fernandes, R Montes-Mico, J B Almeida, and M A Parafita Age differences in central and peripheral intraocular pressure using a rebound tonometer Br. J. Ophthalmol., December 1, 2006; 90(12): 1495 - 1500. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Ravin and E. J. Higginbotham Are we there yet? Celebrating the centennial of the schiotz tonometer. Arch Ophthalmol, September 1, 2006; 124(9): 1337 - 1338. [Full Text] [PDF] |
||||
![]() |
C. Kaufmann, L. M. Bachmann, Y. C. Robert, and M. A. Thiel Ocular Pulse Amplitude in Healthy Subjects as Measured by Dynamic Contour Tonometry. Arch Ophthalmol, August 1, 2006; 124(8): 1104 - 1108. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Kotecha, E T White, J M Shewry, and D F Garway-Heath The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry Br. J. Ophthalmol., December 1, 2005; 89(12): 1572 - 1575. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kniestedt, S. Lin, J. Choe, A. Bostrom, M. Nee, and R. L. Stamper Clinical Comparison of Contour and Applanation Tonometry and Their Relationship to Pachymetry Arch Ophthalmol, November 1, 2005; 123(11): 1532 - 1537. [Abstract] [Full Text] [PDF] |
||||
![]() |
P-A Tonnu, T Ho, T Newson, A El Sheikh, K Sharma, E White, C Bunce, and D Garway-Heath The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry Br. J. Ophthalmol., July 1, 2005; 89(7): 851 - 854. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |