1From the National Eye Institute, the 4Division of Bioengineering and Physical Sciences, and the 6Pharmacy Department, Clinical Center, National Institutes of Health, Bethesda, Maryland; the 2Departments of Bioengineering and 3Chemical and Biomolecular Engineering, University of Maryland, College Park, Maryland; the 5College of Veterinary Medicine, North Carolina State University, Raleigh, North Carolina; and the 7Emmes Corp., Rockville, Maryland.
PURPOSE. To determine the short and long-term pharmacokinetics and assess the toxicity of a cyclosporine (CsA) episcleral implant for the prevention of high-risk keratoplasty rejection.
METHODS. CsA episcleral implants were made with a high (implant A) or low (implant B) release rate, and in vitro release rates were performed. Short-term pharmacokinetics were performed in rabbits using implant B, and the spatial and temporal spread of drug was observed by sampling from multiple corneal and conjunctival sites at 3 and 72 hours. Implant A was used in long-term pharmacokinetic studies in dogs aged more than 1 year. An ocular toxicity study was performed in dogs older than 1 year.
RESULTS. A high release rate was observed with both implants over the initial 5 months followed by a steady state release. The cumulative release over the 400-day assay period from implants A and B was 3.8 ± 0.3 and 2.3 ± 0.3 mg, respectively. In the short-term pharmacokinetic studies, the cornea had CsA concentrations of 0.15 ± 0.06, 0.07 ± 0.02, and 0.05 ± 0.02 µg/mg at sites centered 8, 13, and 18 mm away from the implant site, respectively. In the long-term pharmacokinetic studies, corneal CsA levels ranged from 0.18 ± 0.06 to 0.009 ± 0.004 µg/mg during the 1-year study. There were no signs of ocular toxicity at 1 year.
CONCLUSIONS. Episcleral implants are safe and effective at delivering therapeutic CsA levels to the cornea to potentially prevent corneal allograft rejection. The implant can be surgically inserted at the time of penetrating keratoplasties, since the implant achieves therapeutic levels as early as 3 hours.
This article has been cited by other articles:
![]() |
A. C. Amrite, H. F. Edelhauser, and U. B. Kompella Modeling of Corneal and Retinal Pharmacokinetics after Periocular Drug Administration Invest. Ophthalmol. Vis. Sci., January 1, 2008; 49(1): 320 - 332. [Abstract] [Full Text] [PDF] |
||||