|
|
||||||||
1 From the Department of Ophthalmology, School of Medicine and 2 W. W. Hansen Experimental Physics Laboratory, Stanford University, Stanford, California.
| Abstract |
|---|
|
|
|---|
METHODS. A high electric field at the tip of a fine wire can, like lasers, initiate plasma formation. Micrometer-length plasma streamers are generated when an insulated 25 micron (µm) wire, exposed to physiological medium at one end, is subjected to nanosecond electrical pulses between 1 and 8 kV in magnitude. The explosive evaporation of water in the vicinity of these streamers cuts soft tissue without heat deposition into surrounding material (cold cutting). Streamers of plasma and the dynamics of water evaporation were imaged using an inverted microscope and fast flash photography. Cutting effectiveness was evaluated on both polyacrylamide gels, on different tissues from excised bovine eyes, and in vivo on rabbit retina. Standard histology techniques were used to examine the tissue.
RESULTS. Electric pulses with energies between 150 and 670 µJ produced plasma streamers in saline between 10 and 200 µm in length. Application of electric discharges to dense (10%) polyacrylamide gels resulted in fracturing of the gel without ejection of bulk material. In both dense and softer (6%) gels, layer by layer shaving was possible with pulse energy rather than number of pulses as the determinant of ultimate cutting depth. The instrument made precise partial or full-thickness cuts of retina, iris, lens, and lens capsule without any evidence of thermal damage. Because different tissues require distinct energies for dissection, tissue-selective cutting on complex structures can be performed if the appropriate pulse energies are used; for example, retina can be dissected without damage to the major retinal vessels.
CONCLUSIONS. This instrument, called the Pulsed Electron Avalanche Knife (PEAK), can quickly and precisely cut intraocular tissues without traction. The small delivery probe and modest cost make it promising for many ophthalmic applications, including retinal, cataract, and glaucoma surgery. In addition, the instrument may be useful in nonophthalmic procedures such as intravascular surgery and neurosurgery.
| Introduction |
|---|
|
|
|---|
At present the most precise tractionless and nontraumatic ophthalmic microsurgical dissections are carried out with short-pulsed (10-9 to 10-12 seconds), laser-based instruments (e.g., Nd:YAG, Nd:YLF, and Ti:Sapphire).1 2 3 These lasers produce optical breakdown of the transparent medium resulting in plasma formation in the focal area of the laser beam. Temperatures within the plasma volume reach a few thousands degrees, and pressures reach several kBar, so that tissue is rapidly ionized, evaporated, and disintegrated.4 5 High temperatures and pressures associated with plasma formation allow for dissection of both soft and rigid tissue.
However, this approach of focusing the short-pulsed laser light through the ocular media is limited by the potential hazard to photosensitive tissue caused by intense pulses of light propagating beyond the focal point of the beam.1 6 In addition, large optical aberrations, for example, while operating on peripheral retina, may increase focal spot size and, consequently, threshold energy, thus reducing the precision and safety of the procedure. Optical breakdown-based ablation is also limited by the need for high optical quality of the liquidair interface essential for tight focusing of the laser beam inside the organ of interest. Finally, short-pulsed lasers are very expensive and require a cumbersome transport system for preserving the high quality of the laser beam.
Another approach to tractionless and "cold" tissue dissection in fluids is more invasive and is based on application of the pulsed, shallow-penetrating lasers delivered into the eye via optical fibers.7 8 9 Fast overheating of a shallow layer of water10 or tissue11 by the laser pulse (ns to µs duration) results in formation of a vapor bubble and disruption and ablation of tissue.10 11 Because the vapor bubble cools down very rapidly (typically in a few microseconds), the effect of this interaction is mainly mechanical, with no detectable thermal damage in the lesion.12 The main advantage of such systems, compared with the lasers focused from outside the eye (Nd:YAG, Nd:YLF), is that the light is absorbed in a shallow layer adjacent to the tip. Thus, the tissue is treated only in close proximity to the tip and has the same accuracy at any part of the eye and in any direction. Both the Er:YAG7 8 13 and ArF9 12 excimer cutting instruments have been used in animal and human intraocular surgery but have failed so far to achieve widespread acceptance in surgical practice because of their prohibitively high cost, large size, and relatively slow pace.13 14 The pulsed electrosurgical system, which emulates the action of the Er:YAG laser by overheating of the conductive fluid using Joule heat, has been proposed15 but was found to be limited to use only for very soft tissues such as retina. Our system, which uses an electric field to generate the plasma, eliminates many of these problems.
Plasma generation in liquids allows for deposition of higher energy density than that achieved with light absorption in tissue. This mechanism provides higher temperatures and pressures and thus is more versatile in its applications to surgery of biological tissues. The ideal endosurgical cutter thus would be the one that combines the advantages of powerful but safely confined plasma-mediated interactions and a flexible microprobe delivery with a compact and portable pulse generator of reasonable cost. With such an instrument one could cut tissue efficiently at any point reachable with the microprobe without damaging tissue beyond the close proximity of the tip. We have developed a system, called the Pulsed Electron Avalanche Knife (PEAK), that may achieve these goals by using electricity, rather than laser photons, to generate plasma microstreamers in a conductive aqueous media.
| Methods |
|---|
|
|
|---|
Cutting was also performed on rabbit retina in vivo. Animals were used in accordance with the ARVO Statement for the Use of Animals in Ophthalmic and Vision Research. Two linear cuts approximately 4 mm in length were produced parallel to the medullary ray in each eye after partial vitrectomy. The eyes were enucleated immediately thereafter and fixed in glutaraldehyde/paraformaldehyde fixative, postfixed in osmium tetroxide, and embedded in Epon 812 (Electron Microscopy Sciences, Fort Washington, PA). One-micrometer sections were cut on an ultramicrotome and stained with toluidine blue for light microscopy.
Polyacrylamide gels used for testing cutting parameters were prepared from mixing 6% and 10% (wt/wt) purified polyacrylamide with phosphate-buffered saline (PBS). These gels were polymerized in a Petri dish, which was then loaded onto an inverted microscope stage. Position and orientation of the PEAK probe was controlled using a three-dimensional micromanipulator.
Physical Principles and Instrument Design
Sub-microsecond dielectric breakdown and plasma formation in
conductive fluids is a well-known phenomenon that has been studied
primarily on the macroscopic (>1 cm) level using two equal-size
electrodes in aqueous salt solutions.16
17
It occurs at
electric fields above 106 V/cm and a discharge
energy density on the order of 1 kJ/cm3. At such
conditions an electron avalanche typically develops during a few tens
of nanoseconds.16
Once the avalanche forms, its
interaction with tissue is similar to laser-based dielectric breakdown.
Ionization and explosive evaporation of liquid medium can disrupt the
adjacent tissue and result in cavitation bubble formation. The high
pressures achieved during plasma formation, along with fast expansion
(>100 m/sec) of the vapor bubble and subsequent collapse of the
cavity, can extend the zone of this interaction beyond the primary
energy deposition zone. Because the vapor bubble cools down very
rapidly (typically in a few microseconds), the effect of the
interaction is mainly mechanical, with no detectable thermal damage to
the surrounding tissue.4
12
18
To use this process in microsurgery, the size of the plasma-generating electrode must be scaled down to micrometers. Additionally, the plasma discharges must be confined to the probes tip. We achieved both these goals by designing an asymmetric bipolar microelectrode, so that a micrometer-sized electrode serves as the site of plasma formation, whereas a second, larger electrode is used simply to close the circuit via the conductive physiological medium.
Our plasma-generating electrode (microelectrode) consists of a 25-µm-diameter wire, which is sealed into a tapered insulator and polished to a hemispherical surface at the exit point, as shown in Figure 1 . A 4-cm-long concentric steel needle (20 gauge or 0.9 mm OD) surrounds the insulator for mechanical protection and is also used as the second electrode. The internal wire is connected to the output of a high-voltage pulse generator, and the external needle is connected to the common.
|
1/(r +
a)2, where a is a radius of the
electrode and r is a distance from it). In effect, then,
high electric field levels are confined to an area comparable to the
electrodes radius. This confinement allows for the generation of
micrometer-sized plasma streamers in conductive medium. | Results |
|---|
|
|
|---|
2
µm) did not change with increasing pulse voltage and energy, but the
length extended up to
200 µm when voltage was increased to 3 kV
and energy to 670 µJ. With further increases in pulse energy, the
streamers multiplied and branched but did not significantly elongate.
Each plasma streamer generated an explosive evaporation of liquid
medium, thereby forming multiple vapor cavities, as can be seen in
Figure 2
. Vapor bubbles growing around each streamer fuse into a spherical
bubble, which expands and then collapses, with its lifetime and maximal
size dependent on the pulse energy.19
For example,
at a discharge energy of 50 µJ, the bubble reaches a maximal radius
of 0.19 mm 16 µs after the pulse and then collapses during the same
period. To test the cutting characteristics of PEAK, we used several
types of ocular tissue extracted from bovine eyes. The energy
requirements and pulse repetition rate that would cut at a practical
speed varied greatly among different tissues. For example, we
found that to dissect the iris and lens at a rate of 1 mm/sec to a
depth of 100 µm/scan, pulse energies of
500 µJ and a repetition
rate of
50 Hz were required. On the other hand, bovine retina could
be dissected in vitreous to full depth (
200 µm) by a single pulse
at an energy level of
90 µJ, with only minor damage to the
underlying retinal pigment epithelium cell layer (see Fig. 3
). In all these experiments the probe was kept in contact with the
tissue. Histologic examination revealed no signs of
coagulation at the edges of the crater. With pulses of 50
µJ at a repetition rate of 10 Hz, we could cut bovine retina at
a linear rate of 1 mm/sec, while sparing the retinal blood vessels
(Fig. 4) . At a pulse energy of 90 µJ, large retinal blood vessels could also
be incised, which results in bleeding, because PEAK cutting
is not accompanied by coagulation.
|
|
|
15 pulses
(Fig. 6 , horizontal sequence). However, the depth of the ruptured zone
increased almost linearly with pulse energy (i.e., from 55 µm
at 27 µJ to 220 µm at 141 µJ on average), as seen in vertical
sequence in Figure 6
.
|
|
|
| Discussion |
|---|
|
|
|---|
2 µm and a
length varying between 10 and 200 µm. The high temperatures and
pressures associated with these plasma streamers allow for precise
dissection of soft tissue without traction, in a fashion similar to the
laser-induced dielectric breakdown. Fast expansion of the vapor cavity,
formed by high temperatures inside the plasma streamers, causes rapid
cooling and mechanical fragmentation of tissue. Histologic examination
of cuts made on bovine ocular tissue revealed no collateral thermal
damage. Mechanical effects of the plasmatissue interaction appear to
be similar to those observed with optical dielectric breakdownbased
instruments.5 Because plasma is localized in front of the probe and does not propagate forward as the tissue is removed, pulse energy level is the ultimate determinant of cutting depth. Although increasing the repetition rate increases the speed with which this depth is achieved, the depth of cutting cannot be increased past a saturation point unless the pulse energy is increased. This phenomenon is an important safety feature: inevitable variations in the number of pulses delivered to each point of tissue during surgical dissection using a continuous train of pulses will not result in variation of the cutting depth. Thus, for example, dissection of a pathologic membrane on top of the retina will not result in retinal lesion, regardless of how many pulses are applied, so long as the energy level is set to a crater depth less than the membrane thickness, and the probe is not moved in a forward direction.
We found that fragmentation of a 10% polyacrylamide gel, which mimics the more rigid structures of the eye (thick fibrotic membranes, lens, and lens capsule), is not accompanied by ejection of debris into the surrounding liquid. Importantly, this suggests that PEAK can dissect tissue of similar consistency without contamination of the medium. This effect will be important during intraocular surgery, where continued imaging of tissue through the liquid medium is essential but where the fluid exchange rate is limited.
Efficiency of tissue dissection with PEAK varies considerably
with the tissues mechanical properties. Softer tissue requires fewer
pulses and/or lower voltage to achieve the same depth of cut as more
rigid tissue. The substantial difference in mechanical strength and in
conductivity of different types of tissue makes it possible to perform
tissue-selective surgery. For example, retinal tissue (in vitro bovine
eyes) can be cut at a pulse energy of
50 µJ, with no
visible damage to the larger retinal blood vessels (see Fig. 4
),
whereas those can be dissected at pulse energies exceeding 90 µJ.
Rabbit retina in vivo is cut at even lower pulse energies:
17 µJ
(see Fig. 5
). Because tissue is dissected without substantial heat
deposition in the surrounding medium, other modalities such as
diathermy may be necessary to control bleeding from cut vessels.
The ability to control crater depth through energy deposited and the number of pulses fired may allow the development of applications such as "shaving" of soft epiretinal membranes adherent to the retina or removal of pathologic or unwanted tissue surrounding a vessel without damaging the vessel itself. The instruments ability to cut at controlled depth, combined with its capacity for tissue-selective surgery, may eventually allow surgeons a safe, efficient, and bloodless way to remove highly vascularized and tightly adherent epiretinal fibrotic membranes in conditions such as diabetic retinopathy, proliferative vitreoretinopathy, and retinopathy of prematurity.
The instruments ability to both cut without traction and potentially to "shave" tissue layer by layer makes PEAK potentially useful for cataract surgery as well. Tractionless cutting of the anterior capsule could reduce the incidence of capsule rupture. Layer by layer shaving of the lens material would then allow the surgeon a controlled method for removing the cataract without jeopardizing the integrity of the posterior capsule.
We are currently in the process of testing other ophthalmic applications for the instrument, including scleral shaving to facilitate trans-scleral drug delivery, capsulotomy and lens emulsification, initial dissection of the internal limiting membrane during macular hole surgery, dissection of retinal tissue during radical retinectomy, and modification of the trabecular meshwork in glaucoma surgery. In addition to its use in ophthalmology, PEAK microelectrodes could be attached to thin flexible cables and positioned deep inside the human body to aid in, for example, vascular surgery, neurosurgery, and controlled tumor removal. This technology could replace cumbersome and expensive short-pulsed laser systems in endosurgical applications. Its simplicity and low cost may lead to widespread acceptance in clinical practice if the unique capabilities outlined in this work are confirmed by further preclinical and clinical studies.
| Acknowledgements |
|---|
| Footnotes |
|---|
Submitted for publication January 30, 2001; revised May 24, 2001; accepted May 31, 2001.
Commercial relationships policy: C (DVP, MSB); P (DVP); R (DVP); N (all others).
The publication costs of this article were defrayed in part by page
charge payment. This article must therefore be marked
"advertisement" in accordance with 18 U.S.C.
1734
solely to indicate this fact.
Corresponding author: Daniel V. Palanker, W.W. Hansen Experimental Physics Laboratory, Stanford University, Stanford, CA, 94305-4085. palanker{at}stanford.edu
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. G Priglinger, D. Palanker, C. S Alge, T. C Kreutzer, C. Haritoglou, M. Grueterich, and A. Kampik Pulsed electron avalanche knife: new technology for cataract surgery Br. J. Ophthalmol., July 1, 2007; 91(7): 949 - 954. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. W. Chalberg, A. Vankov, F. E. Molnar, A. F. Butterwick, P. Huie, M. P. Calos, and D. V. Palanker Gene transfer to rabbit retina with electron avalanche transfection. Invest. Ophthalmol. Vis. Sci., September 1, 2006; 47(9): 4083 - 4090. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Priglinger, C. Haritoglou, D. V. Palanker, C. S. Alge, A. Gandorfer, and A. Kampik Pulsed Electron Avalanche Knife (PEAK-fc) for Dissection of Retinal Tissue Arch Ophthalmol, October 1, 2005; 123(10): 1412 - 1418. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Miller, D. V. Palanker, A. Vankov, M. F. Marmor, and M. S. Blumenkranz Precision and Safety of the Pulsed Electron Avalanche Knife in Vitreoretinal Surgery Arch Ophthalmol, June 1, 2003; 121(6): 871 - 877. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Fletcher, D. V. Palanker, P. Huie, J. Miller, M. F. Marmor, and M. S. Blumenkranz Intravascular Drug Delivery With a Pulsed Liquid Microjet Arch Ophthalmol, September 1, 2002; 120(9): 1206 - 1208. [Abstract] [Full Text] [PDF] |
||||
![]() |
Daniel. V. Palanker, M. F. Marmor, A. Branco, P. Huie, J. M. Miller, S. R. Sanislo, A. Vankov, and M. S. Blumenkranz Effects of the Pulsed Electron Avalanche Knife on Retinal Tissue Arch Ophthalmol, May 1, 2002; 120(5): 636 - 640. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |