(Investigative Ophthalmology and Visual Science. 2002;43:371-376.)
© 2002
by The Association for Research in Vision and Ophthalmology, Inc.
Estimation of Human Corneal Oxygen Consumption by Noninvasive Measurement of Tear Oxygen Tension While Wearing Hydrogel Lenses
Joseph A. Bonanno1,
Thomas Stickel1,
Tracy Nguyen1,
Trina Biehl1,
Donna Carter1,
William J. Benjamin2 and
P. Sarita Soni1
1 From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
2 Eye Physiology and Ocular Prosthetics Laboratory, School of Optometry, University of Alabama at Birmingham, Birmingham, Alabama.
 |
Abstract
|
|---|
PURPOSE. To devise a procedure for direct estimation of corneal oxygen
consumption in human subjects.
METHODS. Tear oxygen tension (PO2) was measured at the
posterior surface of two standard hydrogel contact lenses (38% water,
0.2 and 0.06 mm thick, oxygen transmissibility [Dk/t] = 4.2 and
14 x 10-9 cm · mL O2/mL · sec ·
torr) and one newly available hydrogel-silicone polymer lens (Dk/t = 99 x 10-9). The oxygen-sensitive dye,
Pd-meso-tetra (4-carboxyphenyl) porphine, bound to bovine serum
albumin, was incubated with the lenses overnight. The lenses, coated
with the proteindye complex, were placed on four subjects eyes, and
tear PO2was measured in the open eye and after
5 minutes of eye closure, using a timedomain phosphorescence
measurement system. Given the tear
PO2, lens Dk/t, and corneal
thickness, oxygen consumption
(QC, in mL
O2/cm3 · sec) could be
calculated from established oxygen diffusion models.
RESULTS. Protein-dye complex bound to the lens surface enabled reporting of tear
PO2 for long periods. As expected,
estimated tear PO2 was higher in subjects wearing
lenses with higher Dk/t: mean open-eye
PO2 = 30.6 ± 3.1 and 8.1 ± 1.3 torr for the thin and thick hydrogel lenses, respectively, and
97.6 ± 22.9 torr for the hydrogel-silicone lens. After 5 minutes
of eye closure, tear PO2 was
significantly reduced and reached a new steady state in approximately
20 seconds after eye opening. Fitting a single exponential model to the
data and extrapolating to t = 0 provided
an estimate of PO2 under the closed
lid for the thin hydrogel (PO2 =
7 ± 2.3 torr) and the hydrogel-silicone lens
(PO2 = 22.6 ± 4 torr). After 5
minutes of eye closure with the thick hydrogel lens, tear
PO2 remained constant for
10
seconds after eye opening (mean PO2 =
3.9 ± 0.7) before increasing to a new steady state. This delay
could be accounted for by the time needed for oxygen to diffuse to the
posterior surface of the lens. Calculated
QC ranged from 2.2 x 10-4 to 3.7 x
10-6 mL
O2/cm3 · sec) at the
highest and lowest PO2s,
respectively, and is comparable to previous in vitro and in vivo
estimates.
CONCLUSIONS. Tear PO2 behind hydrogel lenses can
be measured in human subjects using the phosphorescence of the
porphyrin-protein complex bound to the lens surface. The method is
simple, fast, reliable, and noninvasive, allowing quick and direct
estimates of QC. In addition to
contact lens wear, this method should be useful for examining the
effects of disease, surgery, or topical drugs on the corneal oxygen
consumption rate.
 |
Introduction
|
|---|
Assessment of metabolic activity in vivo by noninvasive
techniques is a desirable approach for studying the normal physiology
of ocular tissues and how it is altered by disease and drugs, surgery,
or other interventions. Fluorescence- and phosphorescence-based
techniques offer high sensitivity and are applicable to studying the
physiology of many ocular structures that are optically accessible.
Autofluorescence of naturally occurring substances (e.g., the reduced
nicotinamide adenine dinucleotide-to-nicotinamide adenine
dinucleotide [NADH/NAD] ratio) can detect the tissues
metabolic state1
and is advantageous, because addition of
exogenous agents is not required. Unfortunately, the autofluorescence
signal is typically very weak or may require ultraviolet illumination,
but two-photon techniques could circumvent the radiation
hazard.2
Fluorescent and phosphorescent dyes are more
sensitive, but must be delivered in usable concentrations to the site
of interest and/or may have toxic interactions, which could limit
clinical applicability.
Previously, we have reported the use of a phosphorescence-quenching
technique to measure the tear PO2
beneath contact lenses in rabbits.3
This method has also
been used to determine anterior chamber4
and retinal
vasculature5
PO2. Although it
is of interest to contact lens researchers and manufacturers to use
tear PO2 to assess lens performance, this
approach can also be used to measure corneal oxygen consumption
(QC) in vivo. Clinical
response to contact lens wear and laboratory hypoxia-induced corneal
swelling studies have hinted that there is a wide variability in
corneal oxygen demand in the normal young population.6
Furthermore, it has been shown that oxygen uptake into corneas of
diabetic rabbits7
and humans8
is suppressed.
Thus a sensitive, easily administered, and quick measurement of
QC could be useful in studying the
effects of disease, surgery, topical drug use, or contact lens wear on
the metabolic status of the cornea.
The steady state tear PO2 under a
contact lens is determined primarily by
QC.9
10
Thus, given the
tear PO2 under a contact
lens of known oxygen transmissibility (Dk/t), it is possible to
estimate QC. A previous attempt to
directly measure in vivo human QC
required the use of tight-fitting, fluid-filled goggles and
oxygen-consuming Clark-type electrodes.11
A less invasive,
but indirect and time-consuming approach, relied on corneal swelling
responses to estimate the
PO2 in subjects wearing
contact lenses of known transmissibility.12
In contrast,
the phosphorescence technique that we report in the current study, is a
direct measurement of tear
PO2 beneath contact lenses
that requires only a few minutes of hydrogel contact lens wear.
 |
Methods
|
|---|
Subjects
Four subjects (two men, two women; mean age, 25 ± 2 years)
who were free of ocular and systemic disease and had not worn contact
lenses for at least 6 months participated in this study. The research
adhered to the tenets of the Declaration of Helsinki and was approved
by the Indiana University Human Subjects Committee.
Instrumentation
The principles of oxygen measurement by phosphorescence
quenching have been previously described in detail.3
Briefly, a 10-µsec excitation flash excites a probe whose
phosphorescence is quenched by oxygen. The relationship between
phosphorescence decay lifetime and oxygen concentration follows a
linear relation described by the Stern-Volmer equation. We used a
commercially available system (Oxyspot; Medical Instruments, Inc.;
now available through Harvard Apparatus, Holliston, MA). The flash
excitation was coupled to the illumination optics of a slit lamp (model
FS-1; Nikon, Melville, NY) through a fiber optic cable. The slit was
adjusted to provide a 2 x 2-mm2
illumination. Phosphorescence was collected by a gated photomultiplier
tube mounted on the slit lamps camera port. Gating delay, sample
number, and sampling rate were controlled by computer (Oxyspot
software; Harvard Apparatus, running on a Windows 95 [Microsoft,
Redmond, WA]compatible computer). The phosphorescence decay constant
(
) was computed after each flash, and the average
of 10
successive flashes was computed for each
PO2 data point. Also determined was
the correlation coefficient for the fit of the data to a first-order
exponential decay. At a PO2 less than 50
torr, data with correlation coefficients less than 0.9 were rejected.
At higher PO2 levels, a correlation
coefficient of 0.8 was used as the cutoff. Typically, poor correlations
occurred during blinks or eye movements.
Lenses
Three contact lenses were used: (1) a newly available
hydrogel-silicone lens with Dk/t = 99 x
10-9 (Balafilcon; Bausch & Lomb, Rochester, NY;
Dk = 99 x 10-11
cm2 · mL O2/mL · sec
· torr; thickness = 0.1 mm), a 38% water lens with Dk/t =
14 x 10-9 (Polymacon, Dk = 8.4 x 10-11, thickness = 0.06 mm; Metroptics,
Glendora, CA), and a 0.2-mm thick lens (Dk/t = 4.2 x
10-9; Polymacon; Metroptics). All lenses had
back surface radii of 8.6 mm, diameter of 13.5 mm and -0.50 D power to
achieve uniform thickness.
Preparation of Oxygen-Sensitive Dye and Lens
A 1:9 part mixture of the oxygen-sensitive phosphorescent dye Pd
meso-tetra (4-carboxyphenyl) porphine and bovine serum albumin (BSA)
was obtained from Harvard Apparatus. The powder was dissolved into
Ringers solution with final concentrations (in mM) of 140 NaCl, 2
K2HPO4, 0.61
MgCl2, 1.4 Ca+-gluconate,
and 28.5 Na+-gluconate (pH 7.5). Osmolarity was
adjusted to 300 ± 5 mOsm. The solution was forced through a
0.2-µm filter and placed in a sterile container. A new sterile
contact lens was placed in the dye solution and incubated overnight at
room temperature. The next day, the lens was rinsed with sterile saline
and placed on the subjects right eye.
It is conceivable that a protein coating on the surface of a contact
lens could provide another layer of resistance to the passage of oxygen
across the lens. This would potentially decrease the Dk/t of the lens.
Indeed, very thick, denatured albumin coatings placed on contact lenses
have been found to lower the amount of oxygen reaching the cornea, but
coatings of such thickness are not encountered in
practice.13
Protein coatings from normal wear or mild
protein applications in the laboratory do not significantly alter
oxygen transmissibility14
15
and are more representative
of the slight coatings that were applied in this study. Nevertheless,
we measured the Dk/t of 5 to 7 lenses from each of the three lens types
that were incubated with protein and dye as described earlier (18
coated lenses in all) and compared the Dk/t with that obtained with an
identical number of uncoated lenses. The polarographic method for
determining hydrogel Dk/t has been described in detail
previously.16
An electronic thickness gauge (ET-1; Rehder
Development Co., Castro Valley, CA) was used to verify that the mean
thickness of coated and uncoated lenses were the same. We found no
significant difference of Dk/t between coated lenses and uncoated
lenses of the same material and thickness.
Procedure
Once the stained lens was inserted, it was allowed to settle on
the eye for at least 10 minutes. The subject was seated at the slit
lamp and asked to fixate, using the left eye, on an LED placed across
the room. The subject was allowed to blink at will. The operator
aligned the flash illumination on the center of the cornea and adjusted
the PMT voltage to bring the signal on scale with the 12-bit
analog-to-digital (A/D) converter of the system (Oxyspot; Harvard
Apparatus). Open-eye measurements were then made at 0.5 Hz over 60
seconds and repeated 1 to 2 times to assure that the lens was
completely settled on the eye, which was judged by the successive data
sets being within ±5 torr at a
PO2 less than 50 torr and ±10 torr
at a higher PO2. Phosphorescence from the
anterior surface of the lens was avoided by including a delay between
the flash and data collection. Because the anterior surface is exposed
to air (155 torr), its phosphorescence decays very rapidly (half-life,
20 µsec). Therefore, we used a delay of at least 40 µsec.
To estimate closed-eye PO2, we asked
that subjects close their eyes for 5 minutes while continuing to hold
position in the slit lamp after an open-eye measurement. When directed
to open their eyes, they immediately took up original fixation. They
were instructed to try not to blink for the first 10 seconds and then
to blink at will thereafter. Concomitant with eye opening, data
collection was started and continued for at least 40 seconds at 1 Hz.
Alignment of the flash illumination area with the central cornea was
generally preserved, but, occasionally, small adjustments were
necessary. In the first 10 seconds after eye opening, most of the
individual data point correlation coefficients had to be acceptable to
reconstruct the PO2 change
between the closed-eye and the open-eye condition. If more than two of
the data points in this period had correlations that were not
acceptable, the procedure was repeated until acceptable measures were
obtained. The data collected after eye opening was fit to a first-order
exponential model
 | (1) |
where PO2 is
oxygen tension at any time (in torr), SS is the steady state
oxygen tension, I is the initial (t = 0)
PO2,
k is the rate constant, and t is
time. Regressions were performed using PSI-Plot software (Poly Software
International, Sandy, UT).
Estimation of QC
The steady state tear PO2
under a contact lens is primarily determined by the corneal oxygen
consumption rate (QC). Thus, given the
tear PO2 under a contact
lens of known Dk/t, it is possible to determine the oxygen flux into
the cornea, jC, which leads to an
estimate of QC.
QC (mL O2/mL ·
sec) is calculated from the following equations published by Fatt et
al.9
10
17
18
 | (2) |
and rearranging:
 | (3) |
where jC is oxygen flux (in mL
O2/cm2 · sec) into the
cornea, Pt is the
PO2 in the tears,
Pa is the
PO2 at the endothelial
surface (30 torr), LC is
corneal thickness, and DkC is the
oxygen permeability (in mL O2
cm2/mL · sec · torr) of the cornea
(2.4 x 10-10).19
20
At the
tearscornea interface, the flux into the cornea must be equal to the
flux of oxygen that is leaving the contact lens, given that the
thickness of the tears is relatively small (<10 µm). Therefore,
jC =
jCL, and
jCL is calculated by
 | (4) |
where
DkCL/LCL
is the oxygen transmissibility of the contact lens (Dk/t in the newer
notation). Pant is the
PO2 at the anterior
surface of the lens and is generally assumed to be 155 torr in the open
eye and 55 torr in the closed eye.18
Corneal thickness
(LC) integrated
over a central 3-mm diameter, was measured with a pachometer (Orbscan;
Orbtek, Inc., Salt Lake City, UT).
 |
Results
|
|---|
In previous studies of tear PO2
behind rigid contact lenses in rabbits, simple instillation of
dyeprotein complex solution was made directly into the tears,
followed by lens insertion.3
This worked well in sedated
rabbits, because an adequate amount of dye was retained behind the lens
for 10 to 15 minutes. Using either rigid lenses or hydrogels, this
approach did not work in human subjects, presumably because of more
frequent blinking and rapid washout relative to the sedated rabbit.
Protein binding to hydrogels is a well-known clinical problem, and the
porphyrin dye is completely bound to BSA. We took advantage of this
property and bound the dye-protein complex to the lens surface by
incubation overnight. Large proteins are not expected to penetrate 38%
water hydrogels, and we verified this by sectioning lenses and viewing
at x200 magnification (data not shown).
The quenching constant, qk, and the
lifetime in the absence of oxygen (
0) of this
porphyrinprotein complex are well established.21
22
23
For
the eyecontact lens system we used
qk = 304 (in torr per second) and
0 = 581 µsec, which are the parameters for
this dye at 35°C and pH 7.2.22
We verified that these
parameters were appropriate for our instrumentation. A dye-coated thin
hydrogel was placed in saline in a cuvette and kept at 35°C. The
saline was bubbled with 100% nitrogen gas or air. The estimated
PO2 in nitrogen was less
than 0.1 torr and ranged from 140 to 165 torr in air. To verify that
the phosphorescence parameters were appropriate for the lens on the
eye, a tight-fitting goggle was placed over the eyes of a subject who
was wearing a thin hydrogel dye-coated lens. Humidified nitrogen gas or
air was passed through the goggle and the phosphorescence decay
measured. Again, for air the estimated
PO2 was 140 to 165 torr.
Under nitrogen gas the estimated
PO2 was 0.8 torr. This is a
reasonable level, because oxygen diffusion from the anterior chamber
and occasional oxygen fluxes from the palpebral conjunctiva during
blinks make it difficult to obtain absolute anoxia at the corneal
surface.
Figure 1
shows representative open-eye oxygen measurements for the three lenses
10 minutes after lens insertion. Figure 1A shows data from one subject
for the high-Dk/t lens taken at 2-second intervals for 60 seconds. The
mean ± SD for this data set was 105.6 ± 2.0 torr. Figure 1B
shows representative data for the thin hydrogel (mean ± SD;
31.1 ± 1.2 torr). Figure 1C
shows data for the thick hydrogel for
which the mean ± SD over 60 seconds was 6.7 ± 0.15 torr.
These data illustrate that the variability in
PO2 estimates increased with
increasing PO2 and in the four
subjects, the average of the SDs of each 60-second data set (30
readings) was 3.85, 2.78, and 0.21 torr, for the Balafilcon (Bausch &
Lomb) and thin and thick hydrogel lenses (Metroptics), respectively.
Figure 1D summarizes the open-eye data in the four subjects for each
lens.

View larger version (26K):
[in this window]
[in a new window]
|
Figure 1. Tear PO2 in the open eye.
(AC) Representative data sets of
tear PO2 taken every 2 seconds over
1 minute for the Balafilcon (Bausch & Lomb) and the thin and thick
hydrogel (Metroptics) lenses, respectively. (D)
Mean open-eye PO2 in the four
subjects for each lens. Error bars, SD.
|
|
In preliminary experiments we had subjects who were wearing test lenses
close their eyes for 1, 3, 5, and 10 minutes before opening the eyes,
to measure the PO2 to the open-eye
steady state value. These experiments indicated that approximately 3
minutes of eye closure was sufficient to establish the closed-eye
valuethat is, 5 or 10 minutes of eye closure did not produce lower
estimates. Therefore, for all closed-eye experiments we used 5 minutes
of eye closure to assure closed-eye equilibrium. Figure 2
shows representative data from the same subject shown in Figure 1
for
the three lenses, after eye closure. Figures 2A
and 2B
indicate the
initial estimate (I) of
PO2 for the Balafilcon (Bausch &
Lomb) and the thin hydrogel lenses (Metroptics) at t =
0, by fitting the data to a simple exponential increase to a new steady
state. Of interest, Figure 2C
shows that with the thick hydrogel lens
there was no change in PO2
for approximately 10 seconds after eye opening. The average of the
first six readings was taken as the closed-eye
PO2. Figure 2D
summarizes
the closed-eye data in the four subjects.

View larger version (29K):
[in this window]
[in a new window]
|
Figure 2. Tear PO2 in the closed eye.
(AC) Representative data sets of
tear PO2 after 5 minutes of
closed-eye lens wear for the Balafilcon (Bausch & Lomb) and the thin
and thick hydrogel (Metroptics) lenses, respectively.
(A, B, dashed
line) The fit to an exponential model. I is initial
PO2 at t =
0, SS is the steady state
PO2, and
r2 is the coefficient of
determination. For the thick hydrogel lens (C), the first
six data points were averaged to obtain the closed-eye
PO2 estimate.
(D) Mean closed-eye
PO2 in the four subjects
for each lens. Error bars, SD.
|
|
We suspected that the observed delayed increase in
PO2 just after eye opening with the
thick hydrogel lens was due to the time needed for oxygen to diffuse
across the 200-µm thickness of the lens. The diffusion coefficient
(D) for oxygen in water at 35°C is approximately 3 x
10-5
cm2/sec.24
From the relation,
t = x2/D,
where x is distance and t is time, we can
calculate that it would take approximately 13 seconds for a change in
oxygen to appear at the posterior lens surface, which is within a few
seconds of what we observed.
Figure 3
shows the mean QC calculated from the
estimated tear PO2 for the
six conditions (three lenses, open and closed eye). The data indicate
that QC varied significantly with tear
PO2.
QC ranged from 2.2 x
10-4 at approximately 100 torr surface
PO2 to 3.7 x
10-6 mL
O2/cm3 · sec at 4 torr).

View larger version (10K):
[in this window]
[in a new window]
|
Figure 3. QC as a function of
PO2. QC was
estimated for each of the six conditions (three lenses, open and closed
eye). Error bars, SD.
|
|
 |
Discussion
|
|---|
Our goal in this study was to show that human corneal oxygen
consumption could be estimated by a direct, noninvasive measure of tear
PO2 beneath contact lenses. Given
the oxygen transmissibility of a contact lens, the oxygen flux through
the lens can be calculated, because the boundary conditions (an assumed
front surface PO2 and the
measured back surface PO2) are known.
In the steady state, oxygen flux out of the lens must be equal to flux
into the cornea. From this relation and the corneal thickness, an
estimate of QC can be
obtained.10
At high surface
PO2 we estimated
QC to be 2.2 x
10-4 mL
O2/cm3 · sec. This is
approximately two to three times that reported in dissected rabbit
corneas.25
This difference could be due to species
differences or more likely because rabbit corneal oxygen consumption
was suppressed by the trauma of explantation to an in vitro measurement
apparatus. Weissman and Fazio26
estimated in vivo human
corneal oxygen fluxes based on the known lens Dk/t and estimated
surface PO2 from corneal swelling
experiments. At 25 torr, Weissman12
estimates human in
vivo QC to be 4.85 x 10-5. From our data, the calculated
QC was 5.8 to 6.2 x
10-5 mL
O2/cm3 · sec at a
PO2 of 25 to 30 torr, which
is reasonably close to Weissmans estimate. That
QC decreases with decreasing surface
PO2 is not unexpected.
Early in vivo studies11
indicated that at approximately 20
torr, QC began to decrease. Further,
recent mathematical modeling of oxygen distribution from the front to
the back surface of the cornea has shown that even at an open-eye
surface PO2 of 70 torr, a
small portion of the central stroma is anoxic and at surface
PO2 between 30 and 40 torr,
basal epithelial cells are hypoxic,27
which would
significantly suppress O2 consumption.
This study shows that tear PO2 can
be measured in human subjects using the oxygen quenching of the
phosphorescence probe Pd-meso tetra (4-carboxyphenyl) porphyrin. The
measurement is not completely noninvasive, because it requires hydrogel
lens wear. Lens wear itself can have mechanical effects on the surface
epithelium that could suppress
QC. This could be tested
by determining whether other mild forms of trauma (e.g., surface
drying, brief wear of a rigid contact lens, or brief touch with an
applanation tonometer) affects QC.
Because of the nature of measuring phosphorescence decay accelerated by
oxygen quenching, the measurement is most sensitive at a low
PO2. This is exemplified
by the greater variability of individual open-eye data sets at a high
PO2. The dyeprotein complex
immobilized to the surface of contact lenses could act as an oxygen
sensor for many hours; however, the individual lengths of the
experiments in this study were no more than 20 minutes. Binding the dye
complex to the lens was needed, because direct instillation into the
tears led to rapid loss of signal (<1 minute) due to washout.
Proteindye binding to the lens surface had no effect on lens Dk/t.
Also, binding to the lens did not alter the dyes quenching
parameters, presumably because the primary interaction of the porphyrin
is with BSA. We assume that the dye complex is indicating
PO2 at the lens surface and not
from inside the lens. Because BSA has a molecular weight of
66 kDa
it is not expected to penetrate beyond the lens surface. Light
microscopy of lens sections indicated that only the surface was
stained. Furthermore, the delay between eye opening and a change in
measured PO2 when subjects wore
the thick hydrogel would have been significantly shorter if dye complex
had penetrated the lens. Last, because both front and back lens
surfaces are stained, the front surface phosphorescence had to be
removed. The front surface of the lens is exposed to air (155 torr) and
the decay constant at 155 torr is 20 µsec. Thus significant
phosphorescence from the front surface could be avoided by delaying
data acquisition after the flash by approximately 40 µsec.
Estimates of closed-eye PO2 have
been of interest to contact lens researchers for many years. In the
current study, we used closed-eye
PO2 estimates to extend the
range of surface PO2 to examine its effect
on QC. After eye opening,
the fit of the oxygen measurement data to a simple exponential model
was very good. Because of the fitting process, the data, especially in
the first 10 seconds, had to be of high quality to get an accurate
estimate of the initial PO2. In
practice, data are often rejected because of poor fixation after eye
opening or, more commonly, excessive blinking. However, because of the
relatively short time of eye closure, it is convenient to repeat the
closed-eye data collection until an acceptable data set is obtained.
Previously, in vivo measurements of corneal oxygen consumption were
very cumbersome and somewhat invasive (use of fluid-filled tight
fitting goggles)11
or they were lengthy procedures that
relied on the relationship between surface oxygen and corneal
swelling.26
In contrast, the current procedure is
relatively simple and quick and has broad applicability. We anticipate
that very few human subjects would be unable to tolerate wearing a
hydrogel lens for 20 minutes. Within this time,
QC can be determined at two
PO2s (open and closed
eye). With this technique, the effects of disease, drugs, or surgical
interventions on QC could
be determined. For example, it could be used as a measure of the depth
of the effects of a disease on the cornea (e.g., diabetes) and of the
effects of topical drug use on metabolic activity, recovery of
QC after photoablative surgery, as a
determinant in the wound-healing process, or to study the relation
between sensory innervation and corneal metabolism (e.g., neurotrophic
keratopathy, cataract surgery, or penetrating keratoplasty).
 |
Footnotes
|
|---|
Supported by Grant EY12934 from the National Institutes of Health.
Submitted for publication May 30, 2001; revised September 25, 2001;
accepted October 8, 2001.
Commercial relationships policy: N.
The publication costs of this article were defrayed in part by page
charge payment. This article must therefore be marked
"advertisement" in accordance with 18 U.S.C.
1734
solely to indicate this fact.
Corresponding author: Joseph A. Bonanno, Indiana University, School of
Optometry, 800 E. Atwater Avenue, Bloomington, IN 47405;
jbonanno{at}indiana.edu
 |
References
|
|---|
-
Masters, B. (1985) A noninvasive optical measurement of tear oxygen tension at the corneal epithelium Curr Eye Res 4,725-727[Medline][Order article via Infotrieve]
-
Piston, DW, Masters, BR, Webb, WW (1995) Three-dimensionally resolved NAD(P)H cellular metabolic redox imaging of the in situ cornea with two-photon excitation laser scanning microscopy J Microsc 178,20-27[Medline][Order article via Infotrieve]
-
Harvitt, D, Bonanno, J. (1996) Direct noninvasive measurement of tear oxygen tension beneath gas-permeable contact lenses in rabbits Invest Ophthalmol Vis Sci 37,1026-1036[Abstract/Free Full Text]
-
McLaren, JW, Dinslage, S, Dillon, JP, Roberts, JE, Brubaker, RF (1998) Measuring oxygen tension in the anterior chamber of rabbits Invest Ophthalmol Vis Sci 39,1899-1909[Abstract/Free Full Text]
-
Shonat, RD, Wilson, DF, Riva, CE, Pawlowski, M. (1992) Oxygen distribution in the retinal and choroidal vessels of the cat as measured by a new phosphorescence imaging method Appl Opt 31,3711-3718
-
Sarver, M, Polse, K, Baggett, D. (1983) Intersubject difference in corneal edema response to hypoxia Am J Optom Physiol Opt 60,128-131[Medline][Order article via Infotrieve]
-
Herse, P, Petchell, M. (1998) Oxygen consumption and ATPase activity in the corneal epithelium of rabbits with alloxan induced hyperglycemia Acta Ophthalmol Scand 76,528-532[Medline][Order article via Infotrieve]
-
Rubinstein, MP, Parrish, ST, Vernon, SA (1990) Corneal epithelial oxygen uptake rate in diabetes mellitus Eye 4,757-759
-
Fatt, I, Freeman, R, Lin, D. (1974) Oxygen tension distributions in the cornea: a re-examination Exp Eye Res 18,357-365[Medline][Order article via Infotrieve]
-
Fatt, I, Bieber, M, Pye, S. (1969) Steady state distribution of oxygen and carbon dioxide in the in vivo cornea of an eye covered by a gas-permeable contact lens Am J Optom 46,3-14
-
Jauregui, MJ, Fatt, I. (1972) Estimation of the in vivo oxygen consumption rate of the human corneal epithelium Am J Optom Arch Acad Optom 49,507-511
-
Weissman, BA (1984) Oxygen consumption of whole human corneas Am J Optom Physiol Opt 61,291-292[Medline][Order article via Infotrieve]
-
Benjamin, W, Hill, R. (1980) Ultrathins: the case for continuous care J Am Optom Assoc 54,277-279
-
Huth, SW, Lannom, CS, Lannom, S. (1984) Effect in in vivo and in vitro surface protein deposits on the oxygen permeability of polyhydroxyethylmethacrylate gel contact lenses Am J Optom Physiol Opt 61,232-238[Medline][Order article via Infotrieve]
-
Refojo, MF, Leong, FL, Ueno, N, Herman, J. (1982) Oxygen transmissibility and hydration of new and used hydrogel contact lenses J Am Optom Assoc 53,215-218[Medline][Order article via Infotrieve]
-
American National Standards Institute. Merrifield, VA: Optical
Laboratories Association, Publication no. ANSI
Z80.20-1998:9,1617,34,4357.
-
Fatt, I, Bieber, M. (1968) The steady-state distribution of oxygen and carbon dioxide in the in vivo cornea Exp Eye Res 7,103-112[Medline][Order article via Infotrieve]
-
Fatt, I. (1968) Steady-state distribution of oxygen and carbon dioxide in the in vivo cornea: part II Exp Eye Res 7,413-430[Medline][Order article via Infotrieve]
-
Freeman, R, Fatt, I. (1972) Oxygen permeability of the limiting layers of the cornea Biophys J 1972,237-247
-
Weisman, B, Selzer, K, Duffin, R, Petit, T (1983) Oxygen permeability of rabbit and human corneal stroma Invest Ophthalmol Vis Sci 24,645-647[Abstract/Free Full Text]
-
Vanderkooi, JM, Wilson, DF (1982) A new method for measuring oxygen concentration in biological systems Adv Exp Med Biol 200,189-193
-
Vanderkooi, J, Maniara, G, Green, T, Wilson, D (1987) An optical method for measurement of dioxygen concentration based upon quenching of phosphorescence J Bio Chem 262,5476-5482[Abstract/Free Full Text]
-
Wilson, DF, Vanderkooi, JM, Green, TJ, Maniara, G, DeFeo, SP, Bloomgarden, DC (1987) A versatile and sensitive method for measuring oxygen Adv Exp Med Biol 215,71-77[Medline][Order article via Infotrieve]
-
Weast, RC (1990) CRC Handbook of Chemistry and Physics 70th ed. CRC Press Boca Raton, FL.
-
Freeman, R. (1972) Oxygen consumption by the component layers of the cornea J Physiol 225,15-32[Abstract/Free Full Text]
-
Weissman, BA, Fazio, DT (1982) Human corneal oxygen flux under soft contact lenses Am J Optom Physiol Opt 59,635-638[Medline][Order article via Infotrieve]
-
Harvitt, D, Bonanno, J. (1999) Re-assessment of corneal oxygen distribution during contact lens wear Optom Vis Sci 76,712-719[Medline][Order article via Infotrieve]