1From AMO Groningen, Groningen, The Netherlands; and 2Laboratorio de Optica, Departamento de Fisica, Universidad de Murcia, Murcia, Spain.
| Abstract |
|---|
|
|
|---|
METHODS. An adaptive optics vision simulator was constructed that allows psychophysical tests to be performed while viewing targets through any desired ocular wave-front profile. With this simulator, the subjective visual performance of four subjects was assessed by letter acuity and contrast sensitivity (at 3, 6, and 15 cyc/deg) for two different values of induced spherical aberration. The values of spherical aberration were chosen to reproduce two conditions: the average amount measured in pseudophakic patients with implanted IOLs having spherical surfaces and the complete correction of the individuals spherical aberration. Visual performance was assessed in both white and green light, at best focus and for defocus of ±0.5 and ±1.0 D.
RESULTS. There was an average improvement in visual acuity associated with the correction of spherical aberration of 10% and 38% measured in white and green light, respectively. Similarly, average contrast sensitivity measurements improved 32% and 57% in white and green light. When spherical aberration was corrected, visual performance was as good as or better than for the normal spherical aberration case for defocus as large as ±1 D.
CONCLUSIONS. Correcting ocular spherical aberration improves spatial vision in the best-focus position without compromising the subjective tolerance to defocus.
Vision research has been expanded appreciably through the use of adaptive optics in recent years. Liang et al.4 used adaptive optics to correct higher-order aberration, with resultant improved contrast sensitivity. Yoon and Williams7 conducted more in-depth experiments to study the impact on visual performance of the correction of the wave-front aberration with adaptive optics. They extended measurements to include monochromatic and polychromatic contrast sensitivity and visual acuity testing and found improvements in the contrast sensitivity (at 16 and 24 cyc/deg) on the order of two times in polychromatic light and on the order of three times in monochromatic light. More recently, Fernandez et al.13 used adaptive optics to investigate the impact of different aberration patterns on letter acuity and found that even when a subjects wave-front aberration is altered in a manner that results in similar modulation transfer functions (MTFs) and root mean squared (RMS) wave-front aberration errors, visual acuity could be surprisingly different. These results demonstrate that adaptive optics is a powerful tool for a controlled laboratory study of the visual benefit of correcting higher-order aberrations.
IOLs are a natural candidate for practical static correction of higher order aberrations. The Tecnis IOL (AMO, Santa Ana, CA) attempts partial correction of the wave-front aberration by removing spherical aberration. It was designed with aspheric optics, to correct the average corneal spherical aberration of an average patient with cataract.21 Recent studies have revealed that this lens successfully eliminates spherical aberration in the average pseudophakic eye.8 The average postoperative Zernike coefficient for spherical aberration (Z(4,0)) for a 4-mm pupil was 0.007 µm in eyes with Tecnis IOL implants and 0.081 µm in eyes with spherically surfaced IOL implants (SI40NB; AMO). These results showed that when currently available spherically surfaced IOLs are implanted, significant amounts of positive spherical aberration remain uncorrected. As a result of the correction provided by the Tecnis lens, the postoperative contrast vision of patients was improved.8 19 20 At present, this is the only IOL that attempts to reduce the postoperative wave-front aberration of pseudophakic eyes. However, it corrects for the average corneal spherical aberration. It is therefore unclear what the visual benefit would be with either a customized IOL designed to correct all wave-front aberrations or an aspheric lens customized for the individual and designed to completely correct the spherical aberration. Along with the potential benefits, it has also been suggested that ocular spherical aberration increases depth of focus and therefore increases tolerance to defocus.22 23
To investigate both the potential advantages and disadvantages of new theoretical designs of IOLs, an adaptive optics vision simulator (AOVS) employing a membrane deformable mirror has been developed at the University of Murcia.24 In this study, we used this system to investigate the potential of IOLs that provide customized correction of spherical aberration (correct the spherical aberration for each individuals eye). We also investigated how this correction would affect subjective tolerance to defocus. Measurements of visual acuity and contrast sensitivity were performed to determine whether an IOL that perfectly corrects both refractive error (defocus and astigmatism) and spherical aberration would have a measurable positive impact on visual performance. Measurements were performed in both monochromatic and polychromatic light to evaluate the role of chromatic aberration.
| Methods |
|---|
|
|
|---|
|
0.2 seconds. Additional details concerning the calibration and testing of this type of deformable mirror can be found elsewhere.24 Once the aberrations have been modified, a second channel is used for testing the subjects visual quality. The targets used for these vision tests are displayed using a high-resolution monitor and a visual stimulus generator (VSG2/5; Cambridge Research Systems, Rochester, UK). This system, placed in the visual path of the subject provides the simulator with the ability to perform sophisticated psychophysical testing.
The range of defocus that can be induced is quite large (generated with the motorized optometer). However, the maximum amount of higher-order aberration modes that can be produced is approximately 0.5 µm. The actual limit depends on the mode and is lower for spherical aberrations. For this reason, spherical aberration was corrected or introduced by placing aspheric phase plates with different amounts of spherical aberration in the setup at a plane conjugate with the subjects pupil (by lenses L1L4). The AOVS requires precise alignment of the subjects pupil with both the membrane mirror and the wave-front sensor. This is achieved by stabilizing the subjects head with a bite bar and monitoring the pupil position with a video camera through beam splitters BS1 and BS2 when the front of the eye is illuminated with a set of infrared LEDs. The system is robust enough to accommodate blinking, which ensures the simulators usefulness for long-duration visual testing.
Subjective visual testing was performed using both polychromatic (white) and green targets. Green targets were created by viewing the polychromatic targets through a narrow-band interference filter centered on 540 nm. Because the wave-front aberrations and the defocus of the subject were measured in infrared light, the optometer was positioned so as to take into account the difference in focus between the green and infrared light brought about by longitudinal chromatic aberration. The luminance at the eye measured for the white light viewing condition was 35 cd/m2, whereas the luminance measured for the green light condition was 24 cd/m2. Although both luminance levels were in the photopic range, the fact that the white and green luminances were different means that no direct comparison of these measurements can be made.
Figure 1b shows an actual picture of the AOVS apparatus. The complete set of optical elements serves to simulate precisely the effect of an IOL with a desired aberration profile (as schematically indicated in the figure).
Subjects and Experimental Conditions
The experiment was performed in four normal young subjects (JF, PP, SM, and EV) aged 27, 29, 31, and 29, respectively. The subjects were all white and were selected after passing a complete ophthalmic examination. The study adhered to the tenets of the Declaration of Helsinki, and signed informed consent was obtained from every subject after the nature and all possible consequences of the study had been explained. Measurements of subjective visual quality were performed in both green light and white light for a 4.8-mm (artificial) pupil size. Two drops of 1% tropicamide were instilled approximately 0.5 hour before the experiments, to initiate cycloplegia and mydriasis.
Subjective measurements of visual performance (acuity and contrast sensitivity) were then obtained for two cases.
Case 1.
The spherical aberration of the subjects eye was corrected with a combination of phase plates and the deformable mirror. All other higher-order wave-front aberrations present in the subjects eyes remained unmodified.
Case 2.
After introducing the spherical aberration typically measured in a pseudophakic population with spherical IOL implants (Z(4,0) = 0.149 µm for a 4.8-mm pupil) in the subjects eye. This value was based on the spherical aberration measured in the pseudophakic population studied by Mester et al.8 and from the corneal spherical aberration found in older eyes.21 27 This spherical aberration was also introduced by using a combination of phase plates and the deformable mirror. All other higher-order wave-front aberrations present in the subjects eyes remained unmodified.
In both cases, astigmatism was corrected by using the membrane mirror, and the subjective best-focus position was determined by allowing the subject to move the optometer (in 0.1-D steps) until a target letter E was judged to be in focus (the subjective best-focus position was determined separately for both cases). This location is referred to as the 0-D position on all figures appearing in this article. All defocused measurements were performed by defocusing the system by the desired amount relative to this best-focus position (for both cases).
In all four subjects the difference in spherical aberration between cases 1 and 2 was approximately the same. The adaptive optics vision simulator provided a reliable control to maintain this difference throughout the experiment. For both cases 1 and 2, the effect of defocus on subjective visual performance was also investigated. Contrast sensitivity (at 6 cyc/deg) and visual acuity were measured at four different defocus positions, ±0.5 and ±1.0 D (achieved by moving the motorized optometer). At each focus position, including best focus, the ocular wave-front aberration was measured.
Visual Acuity Testing
To determine visual acuity experimentally, subjects were asked to perform a four-alternative, forced-choice illiterate E test. The subject identified the orientation (facing left, right, up or down) of a 100% contrast letter presented for 300 ms on the high-resolution monitor of the AOVS. Five letter sizes were presented randomly, and each size was repeated 12 times. Using a nonlinear fit, these data were then applied to a Boltzman growth sigmoidal psychometric function, and the letter acuity was determined to be the stroke thickness of the letter size where 75% of the responses were correct. This procedure was performed using both green and white light and was repeated three times for the best-focus position and two times for each defocused position. Low-contrast visual acuity (LCVA) was measured with the same setup and procedure for letters with 10% contrast. LCVA was measured only in green light.
Contrast Sensitivity Measurements
Contrast thresholds were determined for three spatial frequencies (3, 6, and 15 cyc/deg) using a two-alternative, forced-choice method and the QUEST search procedure.28 Subjects were presented with two images, one after the other, for a duration of 300 ms. A sine wave grating in a flat-topped Gaussian envelope subtending 24 minutes of arc was displayed in one of these images, and the subject was asked to determine which one. The grating image was randomly presented in one of four different orientations (45°, 0°, 45°, and 90°) to minimize orientation effects.
This procedure was performed in the best-focus position in both green and white light and was repeated three times for each spatial frequency. For defocused positions, measurements of the contrast threshold at 6 cyc/deg were taken in green light only, and each measurement was repeated twice.
Theoretical Optical Calculations
The wave-front aberration measurements recorded for each subject at each focus position during the subjective tests were used to compute the point spread function (PSF) and the two-dimensional monochromatic modulation transfer function (MTF). The one-dimensional MTF was then calculated from the radial projection (averaged over all angles) of the two-dimensional MTF.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
As only four subjects were assessed, each of which was relatively young, with low degrees of refractive error, generalizations cannot be made to all patients with cataract. Young subjects were used in this experiment due to the desire to collect reliable psychophysical data, but the improvements presented here may not be representative of the improvements attainable in older subjects because of possible differences in modulation threshold functions (retinal contrast transfer abilities). The larger amount of aberrations in the eyes of older subjects,31 32 compared with younger subjects may also influence the visual improvements attainable in older and younger subjects. Despite these limitations, the results obtained in this study can be compared with already published clinical results obtained with the Tecnis lens.
In the clinical study performed by Mester et al.,8 the average difference in spherical aberration was the same as occurred in the current study in individual subjects. The average improvement in photopic contrast sensitivity measured using a Functional Acuity Contrast Chart (FACT chart; Vision Sciences Research Corp., San Ramon, CA) at 6 cyc/deg was 30% as opposed to the 47% measured in this study (for the white light condition). This relative increase in visual performance can be explained by a number of factors. First, in this study, all subjects had customized correction of their individual spherical aberration, whereas in the Mester et al. study all patients received an average correction. The luminance conditions and pupil sizes were also different in the two studies. In the clinical study by Mester et al. the photopic contrast sensitivity was measured at 85 while 35 cd/m2 was used in this study. In addition, the pupil size of the patients with cataract was controlled by the ambient lighting conditions and was therefore a natural pupil size, whereas in the simulator the pupils of all subjects were pharmacologically dilated and an artificial pupil size was imposed (4.8 mm). This most likely played a role in the fact that the contrast sensitivity values measured in this study were lower than values that are typically measured in a clinical setting for contrast sensitivity. These low values can also be explained by the fact that, although the luminance conditions were photopic, they were on the low end of the photopic scale, and the pupil sizes were relatively large for photopic conditions. In addition, although patients are given no time restrictions on identifying a target in clinical measurements of contrast sensitivity (using a contrast sensitivity chart), the identification task performed on the simulator is made more difficult by the fact that subjects are presented with the target for only 0.3 seconds.
The different levels of improvement measured for different psychophysical tests administered during this study draws attention to the fact that the potential benefit derived from improvements in the visual system varies, depending on the target chosen for the assessment. In white light the average improvement in visual acuity associated with customized correction of spherical aberration was 10%, whereas for the same conditions, the average improvement in the CSF was 32% (as an average of the three spatial frequencies). Similarly, in green light, the average improvement in visual acuity associated with customized correction of spherical aberration was 38%, whereas, for the same conditions, the average improvement in the CSF was 57% (again as an average of the three spatial frequencies). This reflects the fact that contrast sensitivity testing is more sensitive to changes in the retinal contrast than is visual acuity testing.33 Also, retinal images of contrast sensitivity targets contain one pure spatial frequency (even for a defocused eye), whereas high-contrast letters are complex targets that contain more than one spatial frequency. As a result, retinal images of defocused letters may contain phase-contrast reversals. In fact, one subject commented that phase-contrast reversals under defocused conditions made it easier to identify letter orientation. In other studies of the potential of customized wave-front aberration correction,4 7 higher spatial frequencies were chosen for subjective performance tests, because theoretical optical calculations of the MTF reveal that higher spatial frequencies have more potential for improvement. However, we chose to use lower spatial frequencies or the areas around the peak of the CSF curve, because these peak frequencies are more closely correlated with functional vision.34 In a normal population, this peak often occurs at
6 cyc/deg,35 whereas recent studies of the contrast sensitivities of pseudophakic individuals reveal that their peak is
3 cyc/deg.8 19 20
It is not surprising that correcting spherical aberration provides a visual benefit.36 From the results obtained in this study, it seems clear that there are potential benefits to customized correction of spherical aberration in cataract patients who use IOLs. Naturally, patients who have large amounts of spherical aberration, either due to normal aging or to LASIK for myopia correction6 (in this case, LASIK that does not provide wave-front correction) would benefit more from customized correction. Correction with an IOL is sensitive to tilt and decentration of the IOL. Bench measurements of the Tecnis lens indicate that if it is decentered 0.5 mm and tilted <7°. Its optical performance will exceed that of a conventional spherical IOL.21 A higher level of spherical aberration correction would be even more sensitive to correct placement in the eye. The potential benefits associated with the customized correction of spherical aberration are reduced when the pupil is small. As a result, the improvement in visual performance would be largest in younger patients where the pupil is large or in situations where the pupil is dilated due to low light conditions such as driving at night. Yoon and Williams7 revealed that there are large visual benefits associated with the correction of all monochromatic higher-order wave-front aberrations. A study of whether the correction of the remaining higher-order aberrations is practical and realizable using customized IOLs should be performed. The effect of the correction of all these aberrations on depth of focus should also be investigated.
In this study, through-focus measurements of visual performance were conducted to achieve a better understanding of the impact of spherical aberration on pseudoaccommodation due to depth of focus. The classic definition of depth of focus is the range of defocus values for which there is a blur of small enough size that it will not adversely affect the performance of the system. Thus, when the MTF is used for assessment the 80% criterion is often used to determine depth of focus, which is the defocus range over which the MTF is greater than 80% of the peak MTF.29 Figure 11 illustrates the application of this criterion to model eyes with spherical aberrations that are the same as in cases 1 and 2 in the experiments conducted in this study. Using the MTF of idealistic on-axis eye models (containing no higher-order aberrations other than spherical aberration) and the 80% criterion, depth of focus is increased by adding spherical aberration to the eye model. Similar behavior is observed in the measurements. If the 80% criterion is applied to the contrast sensitivity data collected in the AOVS (Fig. 9) , depth of focus with spherical aberration corrected (case 1) again is less than depth of focus with spherical aberration introduced (case 2). As a result, we see that this definition of depth of focus may not be the most relevant definition when functional vision is considered. Indeed, with spherical aberration correction our results suggest that the range of functional vision is not decreased and that the average visual performance is as good or better when spherical aberration is corrected for as much as 1 D of defocus. This may be because even in the corrected state other higher order aberrations remain (average RMS error was still 0.25 µm after correcting the spherical aberration). This serves to flatten the MTF versus defocus curves shown in Figure 11 so that they appear more like those in Figure 10b and allow the subjects to have more tolerance to defocus. The 80% criterion is an arbitrary criterion that may not reflect the nature of tolerance to defocus in the visual system. It may be more relevant to use an absolute threshold metric that is individually determined. More investigation is needed to determine a functional definition of the subjective depth of focus.
|
Through the use of adaptive optics it is possible to correct precisely the higher-order wave-front aberrations of the eye. Although this technique can only be used in laboratory simulations with complicated and expensive equipment, it provides us with a noninvasive method for exploring the potential of new ophthalmic devices. Adaptive optics also provides us with a powerful tool for the interactive design of new ophthalmic devices and a way to test the relationship between ocular optical quality and visual performance under different conditions.
| Footnotes |
|---|
Supported by AMO Groningen and Grant BFM2001-0391 from Ministerio de Ciencia y Tecnología, Spain.
Submitted for publication March 2, 2004; revised June 17, 2004; accepted June 21, 2004.
Disclosure: P.A. Piers, AMO (E, P); E.J. Fernandez, AMO (C); S. Manzanera, AMO (C); S. Norrby, AMO (E, P); P. Artal, AMO (C, P)
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Patricia A. Piers, AMO Groningen, P.O. Box 901, 9700 AX Groningen, The Netherlands; patricia.piers{at}amo-inc.com.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. A. Villegas, E. Alcon, and P. Artal Optical Quality of the Eye in Subjects with Normal and Excellent Visual Acuity Invest. Ophthalmol. Vis. Sci., October 1, 2008; 49(10): 4688 - 4696. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.-H. Wong, S. A. Koopmans, T. Terwee, and A. C. Kooijman Changes in Spherical Aberration after Lens Refilling with a Silicone Oil Invest. Ophthalmol. Vis. Sci., March 1, 2007; 48(3): 1261 - 1267. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Tabernero, P. Piers, A. Benito, M. Redondo, and P. Artal Predicting the Optical Performance of Eyes Implanted with IOLs to Correct Spherical Aberration. Invest. Ophthalmol. Vis. Sci., October 1, 2006; 47(10): 4651 - 4658. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Diaz-Douton, A. Benito, J. Pujol, M. Arjona, J. L. Guell, and P. Artal Comparison of the retinal image quality with a hartmann-shack wavefront sensor and a double-pass instrument. Invest. Ophthalmol. Vis. Sci., April 1, 2006; 47(4): 1710 - 1716. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||