1From the New England Eye Center and the 3Biostatistics Research Center, Division of Clinical Care Research, Tufts-New England Medical Center, Tufts University School of Medicine, Boston, Massachusetts; the 2UPMC Eye Center, Department of Ophthalmology, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania; and the 4Department of Electrical Engineering and Research Laboratory of Electronics, Massachusetts Institute of Technology, Cambridge, Massachusetts.
| Abstract |
|---|
|
|
|---|
METHODS. Ten normal subjects were imaged six times (three before and three after dilation) per day, and the series was repeated on three different days. The order of the scans before pupil dilation was randomized in each of the 3 days of scanning. After pupil dilation, the scans were also randomized in each of the 3 days of scanning. Each series was performed separately for standard-density (128 A-scans per macular and ONH image and 256 A-scans per NFL image) and high-density (512 A-scans per image for all three scan types) scanning.
RESULTS. The mean macular thickness was 235 ± 9.8 µm. A-scan density (or image acquisition speed) had a statistically significant effect (P < 0.05) on the reproducibility of the mean macular thickness, macular volume, and a few sectors of the macular map. No significant dilation effect was found for any of the macular parameters. The best intraclass correlation coefficient (ICC; 94%) for macular scans was found for dilated high-density scanning, with an intervisit SD of 2.4 µm and an intravisit SD of 2.2 µm. The mean NFL thickness for standard scanning was 98 ± 9 µm. NFL reproducibility showed mixed results and had interactions between scan density and dilation for some parameters. For most of the NFL parameters, reproducibility was better with dilated standard-density scanning. The mean NFL thickness ICC for dilated standard scanning was 79%, with an intervisit SD of 2.5 µm and an intravisit SD of 1.6 µm. For the ONH analysis, the reproducibility was better for dilated standard-density scanning for almost all the parameters, except for disc area, horizontal integrated rim volume, and vertical integrated rim area, which were better before dilation. The best reproducibility was found for cup-to-disc ratio (ICC = 97%, with intervisit SD of 0.04 µm and intravisit SD of 0.02 µm).
CONCLUSIONS. StratusOCT demonstrated reproducible measurements of NFL thickness, macular thickness, and optic nerve head parameters. The best reproducibility was found for dilated standard scanning for NFL and ONH parameters and for dilated high-density scanning for macular parameters.
In a previous studies, the reproducibility of measurements obtained with the first prototype OCT instrument11 and with the first generations of commercial OCT instruments was investigated.12 13 14 15 16 17 These devices acquired 100 (transverse) x 250 (axial) data points per image in the prototype version, and 100 (transverse) x 500 (axial) data points per image in the first and second generation of OCT acquired in 1 second.1 3 5 9 18 These studies showed that the OCT technology provides reproducible measurements of the thickness of the retinal layers. Recently, a third-generation OCT (StratusOCT, Carl Zeiss Meditec, Dublin, CA) was introduced and is currently in clinical use. StratusOCT provides images with 8- to 10-µm axial resolution and a maximum of 512 transverse x 1024 axial data points per image acquired in 1.25 seconds, as described in the user manual of the StratusOCT. Compared with previous generations of OCT systems, StratusOCT allows more A-scans to be acquired for increased image resolution. Increasing the number of A-scans produces an image with a high transverse pixel density, resulting in a better image quality, but increases the image acquisition time, which could lead to measurement errors due to eye motion. Also, the individual scan alignment necessary in the high-density scanning may introduce further errors due to the scan repositioning.
An essential quality in determining the utility of a device for clinical use is its measurement reproducibility. The goal of our present study was to determine the reproducibility of macular and nerve fiber layer (NFL) thickness, and optic nerve head (ONH) measurements obtained with OCT-3 in normal human eyes before and after pupil dilation. We also investigated the effect of different A-scan densities on the reproducibility of the measurements.
| Materials and Methods |
|---|
|
|
|---|
Image and Data Processing
The OCT images were obtained using the StratusOCT, with a typical 2-mm deep, 6-mm wide image used for macular and optic disc scans and a 3.4-mm diameter circumpapillary image used for the NFL scans. The StratusOCT performed 400 A-scans per second and could generate OCT images composed of different numbers of A-scans. Imaging was performed at two different transverse scan densities: standard density (high-speed acquisition) and high density (lower-speed acquisition). For each session, the scans were aligned guided by the fundus image provided by the OCT system. The OCT system did not allow additional alignment between visits. The macular scan is composed of six linear scans centered at the fovea equally spaced 30° apart. The ONH scan is similarly obtained from six linear scans centered on the ONH. For the NFL scans, three consecutive circles, with a diameter of 3.4 mm,11 centered on the ONH were acquired. All the standard scan types for the fast speed acquisition were attained simultaneously. For high-density scan types, each scan (each of the six lines and each of the three circles) was acquired sequentially after the scanner was repositioned before each acquisition. Internal fixation was used for all scan types. A description of all the scans performed as well as the scanning parameters is presented in Table 1 . The OCT images were automatically analyzed with the StratusOCT software (ver. A2; Carl Zeiss Meditec), to segment the retinal layers and quantitatively measure retinal morphology. Only good-quality scans with uniform OCT signal distribution along the entire image (an OCT expert examined the entire scan for uniform signal intensity along the highly reflective retinal layers such as NFL and RPE, as previously described Ishikawa H, et al. IOVS 2003;44:ARVO Abstract 3358) and without overt algorithm failure in detecting the retinal borders were included in the study.
|
![]() |
|
ONH parameters were automatically derived by the OCT software from the six linear scans.18 21 The machine automatically determined the edge of the ONH as the end of the retinal pigment epithelium (RPE)choriocapillaris layer. A straight line connected the edges of the RPEchoriocapillaris, and a parallel line was aligned 150 µm anteriorly. Structures below this line were defined as the disc cup, and above this line was the neuroretinal rim. For definition of 10 global parameters, automatically determined by StratusOCT that were included in the reproducibility study, see Table 5 .
|
Statistical Analysis
Only one eye per subject was included in this study. The eyes were analyzed, adjusting them so that the macular map sectors and the clock hours for NFL could be appropriately assessed. In macula tables, sectors 1 and 2 represented the nasal quadrant and 8 and 9 represented the temporal quadrant. In the NFL tables for clock hours, 1:00 to 5:00 represented the nasal clock hours and 7:00 to 11:00 represented the temporal clock hours. All data presented (all OCT macular, NFL, and ONH parameters described earlier for all sessions, all days, and all the subjects imaged in the study) were age corrected, although age was not a statistically significant parameter for all variables.
Age-adjusted variance components analysis with random-effects mixed models were used to determine intraclass correlation coefficients (ICCs) and intervisit (within-subject between-dates) and intravisit (within-subject within-date) standard deviations as measures of reproducibility. As four of these models did not converge because of low variance in our data set (cells marked with * in Table 6 ), variance components models that did not adjust for age were used to determine the ICCs and the SDs in question. In all models, the measured parameter given by OCT was considered to have three variance components: intersubject, intervisit, and intravisit. The ICC was the ratio of the intersubject component of the variance to the total variance. We calculated the intersubject standard deviation by taking the square root of the intersubject component of variance and the intrasubject standard deviation by taking the square root of the residual, or error, component of the variance.
|
| Results |
|---|
|
|
|---|
Included and Excluded Scans
One scan session was excluded from the macular analysis and two scan sessions from the NFL analysis due to algorithm failure. One NFL session was excluded due to poor scan quality. One subject was excluded from ONH analysis due to elevation of the optic disc in a hypermetropic eye.
In all NFL scans the NFL thickness profiles had the characteristic double-hump pattern previously described as the normal NFL pattern.9 12 22 In all ONH scans used in this study, the automated detection of the RPE ends of each of the ONH scans was performed correctly by StratusOCT without the need for operator intervention.
Macular Thickness
There was no effect of scan density or dilation on measured overall macular thickness. The difference between before and after dilation was less than 2 µm for all parameters except sector 5 (the central sector), which was 4 µm. Selected results for mean macular thickness are presented in Table 2 , and the corresponding sectoral results are presented in Figure 2 . A summary of the ICCs and SDs for mean macular thickness, macular volume, and sectoral analysis is presented in Table 3 .
|
|
|
Standard Density.
The ICC for the mean macular thickness after dilation (88%) was only 2% higher than before dilation. The results were similar for macular volume (Table 3) . The ICCs before dilation for macular sectors (range, 51%92%) were 3% to 18% higher than after dilation, except for the macular sectors mac3 and -9 (Table 3) . In general, intervisit and intravisit SDs were similar before and after dilation, ranging from 1 to 8 µm.
High Density.
The ICC for mean macular thickness was slightly higher after dilation (ICC = 94%). Sectoral measurements showed similar ICCs and SDs before and after dilation, with ICCs 1% to 6% higher before dilation (range, 65%96%), and the intra- and intervisit SDs of 1 to 7 µm, except for the sectors mac1, -2, and -7 (Table 3) .
Density and Dilation Effects.
In our data set, we found no statistically significant interactions between scan density and dilation effects on reproducibility. Density had a considerable effect on the reproducibility results for mean macula and total volume and for sectors mac3, -4, -8, and -9. Thus, with higher ICCs and lower SDs, the high-density scanning showed considerably better reproducibility of macular thickness parameters.
RNFL Thickness
The overall mean NFL thickness was higher for the standard-density scanning (9899 ± 9 µm) than the high-density scanning (94 ± 8 µm) both before and after dilation. The overall quadrant and clock hour values after dilation for both standard- and high-density scanning are presented in Figure 3 . Inter- and intravisit SDs for NFL quadrants and clock hours with both standard- and high-density scanning, before dilation, are presented in Table 4 and Figure 4 and after dilation in Table 4 and Figure 5 .
|
|
|
|
High-Density.
Nasal and temporal quadrants showed the thinnest NFL and had similar values before and after dilation (6570 ± 14 µm). Superior and inferior quadrants were thicker (120 ± 15 µm). The ICC for the mean NFL before and after dilation was 79%. The ICCs for the NFL quadrants were higher before dilation, and the values ranged between 71% and 80%, except for the nasal quadrant (60% and 68% before and after dilation). The ICCs for the clock hour positions were inconsistent, with the highest either before or after dilation for the various clock hour positions (Fig. 5) .
Density and Dilation Effects.
The effects of density and dilation showed a statistically significant interaction (P < 0.05) for mean NFL, inferior quadrant, and clock hours 3, 4, 5, 6, 9, and 11. In the case of a statistically significant interaction, the effects of density and dilation must be described separately for each category: scan density and dilation. For parameters without interaction effects, the reproducibility tended to be better with dilated, standard scanning (Table 4) , except for the superior quadrant, which was better with standard scanning before dilation. The effect of dilation and density on the NFL quadrants and clock hours was inconsistent. A density effect was found in all clock hours, except for clock hour 10. A dilation effect was found at clock hour positions 8, 10, and 12. The clock hour positions 1 and 7 showed no effect. The difference in the reproducibility between standard- and high-density scanning, before and after dilation, was not more than 9% for all the NFL parameters.
For both standard- and high-density scanning the quadrants and clock hours intravisit SDs were lower than the intervisit SDs (Figs. 4 5) . Most of the SDs for the quadrants and clock hours were lower after dilation than before dilation. The intravisit SDs ranged between 1 and 5 µm before and after dilation and intervisit SDs ranged between 3 and 15 µm before dilation and 3 and 8 µm after dilation.
Mean NFL thickness from three consecutive scans (obtained as one value by averaging the three NFL scans per session) showed a 3% and 5% increase in ICCs, compared with nonaveraged mean NFL. For quadrants and clock hours, we noted in general a 3% and 13% increase in ICCs for the averaged NFL data compared with the nonaveraged data. The intervisit SDs were less than 1% lower and the intravisit SDs were 1% to 3% higher for the NFL averaged data than for the nonaveraged NFL data.
ONH Parameters
Mean results for all ONH parameters for dilated standard- and high-density scanning are presented in Table 5 . Table 6 presents ICCs and SDs for standard- and high-density scanning before and after dilation. A noteworthy effect on reproducibility between standard- and high-density scanning was found only for the horizontal integrated rim volume and a dilation effect was found only for both vertical integrated rim area and cup volume (topo).
Standard Density.
Mean values for the 10 ONH parameters did not show any difference between undilated and dilated scans. The reproducibility (ICCs) for standard density after dilation was best for most ONH parameters (Table 6) . For disc area, horizontal integrated rim volume, vertical integrated rim area and cup area (topo) the ICC was highest for standard-density scanning, before dilation. Among all 10 ONH parameters, cup-to-disc area ratio (cdr) had the best ICCs (78% and 97% before and after dilation), and cup area (topo) (cat) the worst (65% and 45%).
High-Density.
Reproducibility was better before than after dilation, but was markedly worse than for the standard-density scanning. The 10 ONH parameters did not show any difference between mean values before dilation (0.152.09) and after dilation (0.122.12). Among all 10 ONH parameters, the two parameters that had the highest ICCs were horizontal integrated rim width (area) (hirw) (85% and 73% before and after dilation) and vertical integrated rim area (volume) (vira) (92% and 58%). The lowest ICC was obtained for rim area (ra) (2.9% after dilation).
Density and Dilation Effects.
Dilation and density showed mixed effects for ONH parameters, but no statistically significant interaction was found between the two. There was no dilation effect for any ONH parameters except vira, which showed better ICCs for standard scanning before dilation and for high-density scanning after dilation. No marked density effect was found for any ONH parameters except hirw, which showed a better ICC before dilation for both standard- and high-density scanning.
| Discussion |
|---|
|
|
|---|
In the young group of normal eyes in this study, OCT gave values for mean macular thickness of 235 ± 9.8 µm (Table 2) , mean NFL thickness of 96 ± 9 µm (Fig. 3) , disc area of 2.13 ± 0.4 µm, and cup area of 0.66 ± 0.5 µm (Table 5) . High-density measurements for NFL thickness and ONH parameters tended to be lower than in standard-density scans, perhaps due to the higher accuracy of detection of the retinal posterior borders with the high-density scanning.
This study demonstrated that scan density and dilation must be considered when OCT is used as a diagnostic tool. Reproducibility was good for most of NFL, macula, and ONH parameters evaluated (standard-density scanning). All ICCs were greater than 50% for almost all measures (except high-density ONH), and most were in the 80% to 90% range or higher.
Reproducibility of the mean macular thickness and macular volume was not significantly affected by dilation. Mean macular thickness had the best ICC of 94%, intervisit SD of 2.4 µm, and intravisit SD of 2.2 µm for high-density dilated scanning (Table 3) . In general, high-density scanning had higher ICCs and lower SDs, which are essential for accurate and reliable measurements of retinal thickness, perhaps because high-density scans allow more precise retinal border detection and segmentation. No effect of dilation on the reproducibility of macular parameters was found, probably either because fixation was easier for the subject in macular scans (primary gaze) than for NFL and ONH scans (nasal gaze) or because of the OCT softwares ability to detect retinal borders more easily on the well-defined and relatively uniform retinal features of the macular region.
Very different results were found for center foveal thickness with much higher SDs and lower ICCs than any other macular parameter. The best reproducibility for center foveal thickness was found for high-density before dilation (ICC = 71%) and lowest SDs (intravisit SD = 6 µm and intervisit SD = 12 µm). A similar trend was also noted for the center area of the macular map (mac5), which includes the foveal point. Massin et al.17 found average reproducibility of central foveal thickness of 11 µm in healthy subjects. This high variability is due to the large change in thickness between the foveal point and the thickest macula at the edge of the 1-mm center circle in the macular map. Therefore a small decentration of scans causes large variability in the measured thickness.
The NFL thickness analysis had inter- and intravisit SDs of 2 to 3 µm or less for mean thickness and 15 µm or less for each quadrant and clock hour (Table 4 , Figs. 4 5 ). There were interactions between scan density and dilation for some parameters. Reproducibility was best for dilated standard-density scanning. Similar results were found for nonaveraged and averaged NFL data and may indicate that good and reliable results can be obtained even for a single NFL scan, although further investigations are warranted.
The high-density scanning did not clearly show better reproducibility of the NFL measurements, contrary to the findings in the macular scan, perhaps because of the position of the scanned eye while the image was acquired. The macular scan is acquired in the primary-gaze position and thus is less prone to eye movement and scan placement variability. NFL scanning necessitates an eccentric fixation, which is more challenging for the scanned eye. Therefore, although the standard scanning provides a reduced number of sampling data it is balanced with a shorter acquisition duration that is less susceptible to eye movements.
For the ONH analysis, the reproducibility was best for dilated standard scanning for most parameters. The best reproducibility was found for cup-to-disc ratio (ICC = 97%, with intervisit SD = 0.04 µm and intravisit SD = 0.02 µm; Table 6 ). ONH reproducibility was best for standard scanning, probably because the fast acquisition of the OCT images are less affected by the eye movements and scan placement, which may be because of the eccentric fixation while acquiring the image, as explained earlier. The global ONH parameters are determined by interpolation from the six linear scans through the disc. These six lines may or may not be adequate for an accurate topographical map used for quantitative measurements (due to small sample density over this area). In this study, we used the commercial StratusOCT optic disc scanning procedure. The number of scans required or permitted for ONH topography is not variable. The evaluation of the appropriateness of this ONH mapping technique was beyond the scope of the present study; however, we have already reported the similarity in disease discrimination of the OCT and confocal scanning laser ophthalmoscopic ONH mapping.18
Koozekanani et al.15 showed intrasession variation of approximately 11 µm for overall macular thickness obtained with first-generation OCT. Our results for overall macular thickness showed best ICC of 94% and SD of 2.5 µm for dilated high-density scanning. The most recent publication regarding the reproducibility of the sectoral macular thickness map used the second generation of commercial OCT systems17 and found ICCs of 89% to 99% and SDs of 7 to 14 µm. We obtained ICCs of 86% to 95% and SDs of 2 to 6 µm in most sectors with dilated high-density scanning.
In a similar age group of normal subjects (mean age, 28 ± 5 years), a previous reproducibility study regarding NFL measurements with the prototype OCT,11 showed an ICC of 56% and SD of 10 to 20 µm for overall NFL thickness, an ICC of 24% to 64% and SD of 11 to 26 µm for quadrants, and ICC of 19% to 68% and SD of 11 to 30 µm for clock hours. A more recent publication by Jones et al.12 showed SD of 6 µm for mean NFL thickness, 10 µm for quadrants, and 11 µm for clock hours with OCT-2.
StratusOCT showed an SD of mean macular thickness measurements of 3 µm. The inter- and intravisit SDs were approximately 4 µm for mean NFL thickness, 7 µm for quadrants, and 1 to 15 µm for clock hours (Figs. 4 5) . The lower the standard deviations of the measurements, the smaller the changes in retinal thickness that can be measured accurately. We found that the reproducibility of the third generation of this technology significantly improved for mean NFL thickness (ICC of 79% on normal subjects) and mean macular thickness. This further emphasizes the ability of the OCT technology to measure small changes accurately, potentially resulting in earlier detection of disease or its progression before the onset of functional changes.
Study Advantages and Limitations
Although a direct comparison with previously published results cannot be made because the investigated subjects are not the same, our results in normal subjects of similar age show higher ICCs and lower SDs, which might indicate an improvement in the reproducibility of macular and NFL thickness measurements compared with previous generations of this technology.
Our results were obtained on a young normal group of subjects (30.5 ± 7 years). Caution should be used when applying these data to older subjects.
The protocol for standard scanningfast acquisition (all six linear scans through macula or optic disc and the three circular scans for NFL acquired in a rapid automatic sequence) allows for decreased measurement variability, compared with high-density scanningslow acquisition (each of the scans are acquired consecutively after manual repositioning), which is more susceptible to fixation problems, eye movements, and scanner placement by the operator. Therefore, the variability of high-density scanning measurements was higher probably because of the addition of an operator-induced variability component.
In addition, the scans were not aligned between visits because the OCT did not provide this feature. This adds to the intervisit variability, which is a limitation of the present study. If an additional variance component were included in the analysis due to the scan alignment variability, the intervisit variability of the OCT system might be smaller than reported herein.
For several parameters, the dilation and density effects on the reproducibility were not statistically significant. Although the differences between the dilated and undilated and standard- and high-density ICCs and SDs were not large, the lack of significance may be due to lack of power. Nevertheless, for this type of study, it is impractical to obtain a larger sample size, because no power calculation can be performed.
| Conclusions |
|---|
|
|
|---|
| Footnotes |
|---|
Submitted for publication May 23, 2003; revised December 5, 2003, January 12 and February 13, 2004; accepted February 18, 2004.
Disclosure: L.A. Paunescu, None; J.S. Schuman (P); L.L. Price, None; P.C. Stark, None; S. Beaton, None; H. Ishikawa, None; G. Wollstein, None; J.G. Fujimoto (P)
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Joel S. Schuman, UPMC Eye Center, Department of Ophthalmology, University of Pittsburgh School of Medicine, The Eye and Ear Institute Suite 816, 203 Lothrop Street, Pittsburgh, PA 15213; schumanjs{at}upmc.edu.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. N. Menke, P. Knecht, V. Sturm, S. Dabov, and J. Funk Reproducibility of Nerve Fiber Layer Thickness Measurements Using 3D Fourier-Domain OCT Invest. Ophthalmol. Vis. Sci., December 1, 2008; 49(12): 5386 - 5391. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K.-s. Leung, C. Y.-l. Cheung, R. N. Weinreb, G. Lee, D. Lin, C. P. Pang, and D. S. C. Lam Comparison of Macular Thickness Measurements between Time Domain and Spectral Domain Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., November 1, 2008; 49(11): 4893 - 4897. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K.-s. Leung, C. Y.-l. Cheung, D. Lin, C. P. Pang, D. S. C. Lam, and R. N. Weinreb Longitudinal Variability of Optic Disc and Retinal Nerve Fiber Layer Measurements Invest. Ophthalmol. Vis. Sci., November 1, 2008; 49(11): 4886 - 4892. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Huynh, X. Y. Wang, G. Burlutsky, E. Rochtchina, F. Stapleton, and P. Mitchell Retinal and Optic Disc Findings in Adolescence: A Population-Based OCT Study Invest. Ophthalmol. Vis. Sci., October 1, 2008; 49(10): 4328 - 4335. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Merle, S. Olindo, A. Donnio, R. Richer, D. Smadja, and P. Cabre Retinal Peripapillary Nerve Fiber Layer Thickness in Neuromyelitis Optica Invest. Ophthalmol. Vis. Sci., October 1, 2008; 49(10): 4412 - 4417. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Nagai-Kusuhara, M Nakamura, Y Tatsumi, Y Nakanishi, and A Negi Disagreement between Heidelberg Retina Tomograph and optical coherence tomography in assessing optic nerve head configuration of eyes with band atrophy and normal eyes Br. J. Ophthalmol., October 1, 2008; 92(10): 1382 - 1386. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Cettomai, M. Pulicken, E. Gordon-Lipkin, A. Salter, T. C. Frohman, A. Conger, X. Zhang, G. Cutter, L. J. Balcer, E. M. Frohman, et al. Reproducibility of Optical Coherence Tomography in Multiple Sclerosis Arch Neurol, September 1, 2008; 65(9): 1218 - 1222. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Costello, W Hodge, Y. Pan, E Eggenberger, S Coupland, and R. Kardon Tracking retinal nerve fiber layer loss after optic neuritis: a prospective study using optical coherence tomography Multiple Sclerosis, August 1, 2008; 14(7): 893 - 905. [Abstract] [PDF] |
||||
![]() |
A Martinez, N Proupim, and M Sanchez Retinal nerve fibre layer thickness measurements using optical coherence tomography in migraine patients Br. J. Ophthalmol., August 1, 2008; 92(8): 1069 - 1075. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Zaveri, A. Conger, A. Salter, T. C. Frohman, S. L. Galetta, C. E. Markowitz, D. A. Jacobs, G. R. Cutter, G.-S. Ying, M. G. Maguire, et al. Retinal Imaging by Laser Polarimetry and Optical Coherence Tomography Evidence of Axonal Degeneration in Multiple Sclerosis Arch Neurol, July 1, 2008; 65(7): 924 - 928. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. L. Gabriele, H. Ishikawa, G. Wollstein, R. A. Bilonick, K. A. Townsend, L. Kagemann, M. Wojtkowski, V. J. Srinivasan, J. G. Fujimoto, J. S. Duker, et al. Optical Coherence Tomography Scan Circle Location and Mean Retinal Nerve Fiber Layer Measurement Variability Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2315 - 2321. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. M. Tangelder, R. G. L. Van der Heijde, B. C. P. Polak, and P. J. Ringens Precision and Reliability of Retinal Thickness Measurements in Foveal and Extrafoveal Areas of Healthy and Diabetic Eyes Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2627 - 2634. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Kelty, J. F. Payne, R. H. Trivedi, J. Kelty, E. M. Bowie, and B. M. Burger Macular Thickness Assessment in Healthy Eyes Based on Ethnicity Using Stratus OCT Optical Coherence Tomography Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2668 - 2672. [Abstract] [Full Text] [PDF] |
||||
![]() |
V Pueyo, J Martin, J Fernandez, C Almarcegui, J Ara, C Egea, L Pablo, and F Honrubia Axonal loss in the retinal nerve fiber layer in patients with multiple sclerosis Multiple Sclerosis, June 1, 2008; 14(5): 609 - 614. [Abstract] [PDF] |
||||
![]() |
L Kagemann, T Mumcuoglu, G Wollstein, R Bilonick, H Ishikawa, K A Townsend, M Gabriele, J G Fujimoto, and J S Schuman Sources of longitudinal variability in optical coherence tomography nerve-fibre layer measurements Br. J. Ophthalmol., June 1, 2008; 92(6): 806 - 809. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Nolan, J. M. Stringham, S. Beatty, and D. M. Snodderly Spatial Profile of Macular Pigment and Its Relationship to Foveal Architecture Invest. Ophthalmol. Vis. Sci., May 1, 2008; 49(5): 2134 - 2142. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Patel, F. K. Chen, F. Ikeji, W. Xing, C. Bunce, L. Da Cruz, and A. Tufail Repeatability of Stratus Optical Coherence Tomography Measures in Neovascular Age-Related Macular Degeneration Invest. Ophthalmol. Vis. Sci., March 1, 2008; 49(3): 1084 - 1088. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Nagai-Kusuhara, M Nakamura, M Fujioka, Y Tatsumi, and A Negi Association of retinal nerve fibre layer thickness measured by confocal scanning laser ophthalmoscopy and optical coherence tomography with disc size and axial length Br. J. Ophthalmol., February 1, 2008; 92(2): 186 - 190. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P. Danis, M. R. Fisher, E. Lambert, A. Goulding, D. Wu, and L.-Y. Lee Results and Repeatability of Retinal Thickness Measurements From Certification Submissions Arch Ophthalmol, January 1, 2008; 126(1): 45 - 50. [Abstract] [Full Text] [PDF] |
||||
|
|