METHODS. In a prospective study of 228 eyes, IOP measurements by GAT and DCT were compared, and the effects of central corneal thickness (CCT), corneal curvature, axial length, and anterior chamber depth were analyzed. To evaluate intra- and interobserver variability, IOP was measured in eight eyes by four observers.
RESULTS. There was a high concordance between the IOP readings obtained by DCT and GAT. However, IOP readings were consistently higher with DCT than with GAT (median difference: +1.7 mm Hg, interquartile range [25th75th percentile] = 0.82.7 mm Hg). In contrast to GAT, multivariable regression analysis showed no significant effect of corneal thickness, corneal curvature, astigmatism, anterior chamber depth, and axial length on DCT readings. For repeated measurements the intraobserver variability was 0.65 mm Hg for the DCT and 1.1 mm Hg for the GAT (P = 0.008). Interobserver variability was 0.44 mm Hg for the DCT and 1.28 mm Hg for the GAT (P = 0.017).
CONCLUSIONS. IOP measurements by DCT are highly concordant with IOP readings obtained from GAT but do not vary in CCT and have a lower intra- and interobserver variability. DCT seems to be an appropriate method of tonometry for routine clinical use.
With the dynamic contour tonometer (DCT), a new digital tonometer has been introduced as an alternative to adjustments of applanation tonometry readings based on CCT. The so-called contour-matched tonometer tip has a concave surface that allows the cornea to assume the shape that it naturally assumes when pressure on both sides of the cornea is equal and distortion of the cornea is minimal. Exposing a miniaturized pressure sensor closely to the contour of such a cornea is thought to measure IOP directly (i.e., without systematic errors resulting from force-to-pressure translations; Kanngiesser H, et al. IOVS 2002;43:ARVO E-Abstract 301). In a pilot study on patients before and after corneal refractive surgery (LASIK) DCT has been shown to measure IOP accurately, independent of corneal thickness.11
The purposes of this study were to compare the IOP readings obtained by DCT with those of GAT, to evaluate the ocular structural factors influencing IOP measurements obtained from both tonometers and to determine intra- and interobserver variability in a group of nonsurgical healthy subjects.
| Methods |
|---|
|
|
|---|
All measurements were taken by the same examiner in the following order: biometry, pachymetry, GAT, and DCT. First, axial length, corneal curvature, and anterior chamber depth were measured with an optical biometry system (IOL Master; Carl Zeiss AG, Feldbach, Switzerland). Second, CCT was measured with an ultrasonic pachymeter (model SP-2000; Tomey Corp., Cambridge, MA). The pachymeter probe was placed on the center of the cornea over an undilated pupil and the mean of three readings within a range of ±5 µm was calculated for each eye. Third, GAT was performed on a slit lamp (Haag-Streit, Köniz, Switzerland) with a tonometer calibrated according to the manufacturers guidelines. Before each reading, the measuring drum was reset to approximately 2 mm Hg, and the mean of three consecutive readings was recorded.
Fourth, DCT was performed using a technically identical prototype of the model launched in November 2003 (Pascal dynamic contour tonometer; Swiss Microtechnology AG, Port, Switzerland; slit-lampmounted, self-calibrating, 1 g appositional force, 100 Hz sampling rate, 7 mm tip diameter, 1.2 mm pressure sensor diameter). As DCT provides a digital readout of the IOP on a liquid crystal display (LCD), prior knowledge of the GAT result would not influence the result and made it unnecessary to randomize the order of IOP measurements (always GAT followed by DCT) or to mask the investigator. However, for the study of the intra- and interobserver variability, all four investigators were fully masked to all results. For this part of the study, three GAT readings followed by three DCT readings were taken in eight participants by each investigator, resulting in 192 measurements. The delay between readings by different investigators was kept as short as possible (<30 seconds).
Data are presented as medians and 25th to 75th percentile limits. Comparisons in pressure measurements were performed using the nonparametric Wilcoxon signed ranks test to account for the skewed and nonsymmetrical distribution of data points. P < 0.05 was considered significant. To correct for related data, when two eyes of the same subject were entered into analysis, we performed clustered analyses, using the subject identifier as the cluster variable. We fitted models in which DCT and GAT, respectively, acted as the dependent variable. To adjust for the skewed data distribution, DCT and GAT were transformed into their logarithms. Corneal thickness, corneal curvature, astigmatism, anterior chamber depth, and axial length were entered as continuous independent variables. First each independent variable was assessed in a univariate analysis. Then all independent variables were fitted into two multivariable models (one for GAT and one DCT).
To study the variability between the different investigators and the two measurement readings, we performed analyses of variance (ANOVAs). We used the variance components procedure to estimate the contribution of an independent variable (observer, test, subjects, observer subject interaction, and residual error) to the variance of the dependent variable (pressure measurement). Based on the variance components, we calculated the intraclass correlation coefficients (ICC) for GAT and DCT using the variance component of the subjects in the numerator and the sum of all variances in the denominator. To test for significant differences between the ICCs, we calculated the sum of variance of all noise components for the two tests (all variance components except the subjects variance component). The division of the two noise components was bootstrapped and tested using a one-sample t-test.
To assess the intra- and interobserver variability of DCT and GAT, we calculated the interobserver variability for each of the two tests as the sum of the variance components of the investigator and the investigatorsubject interaction. The variance component of the residual error was used as the intraobserver variability.
Statistical analysis was performed on computer (SPSS statistical software, ver. 10; SPSS Inc., Chicago, IL).
| Results |
|---|
|
|
|---|
The IOP measurements obtained by DCT and GAT demonstrate a high concordance between the two techniques (Fig. 1) . IOP readings obtained by DCT were consistently higher (median difference: +1.7 mm Hg, interquartile range [25th75th percentile] = 0.82.7 mm Hg) than readings obtained by GAT. Multivariable regression analysis on DCT readings showed no effect of corneal thickness (P = 0.65), corneal curvature (P = 0.51), astigmatism (P = 0.89), anterior chamber depth (P = 0.44), or axial length (P = 0.06). In contrast to DCT, IOP readings obtained by GAT depended on corneal thickness (P = 0.012), but were also not significantly influenced by corneal curvature (P = 0.24), anterior chamber depth (P = 0.60), or axial length (P = 0.89).
|
|
| Discussion |
|---|
|
|
|---|
An important factor for the accuracy of IOP measurements is the variability that occurs between measurements performed by the same observer (intraobserver) or by different observers (interobserver), when measuring IOP in the same eye. This study found a smaller intra- and interobserver variability for DCT (0.65 and 0.44 mm Hg) than for GAT (1.11 and 2.38 mm Hg), despite the fact that all examiners had used GAT routinely for at least 5 years, but had no prior experience with DCT. The intra- and interobserver variability found with GAT is very similar to the results found in other studies.1 15 A potential problem of such studies is that repetitive IOP measurements over a short period, which are necessary to calculate the inter- and intraobserver variability, can decrease IOP between the first and the subsequent sets of measurements.14 15 To correct for a possible decline in IOP by the first six measurements the data were reanalyzed without the results obtained during the first sets of GAT and DCT readings. This reduced the amount of inter- and intraobserver variability for both tonometers but did not change the finding that the intra- and interobserver variability found for DCT (0.79 and 0.12 mm Hg, respectively) was smaller than the variability found for GAT (0.97 and 0.50 mm Hg, respectively).
In contrast to GAT, for which the index on the scale is set at 2-mm increments to take into account the intra- and interobserver variability, DCT gives pressure readings with a precision of one decimal place. Despite the considerably lower variability found for DCT in this study, displaying IOP readings with a precision of one decimal place may reflect a "pseudoprecision" that may be clinically misleading. Based on the findings in this study, we recommend recording IOP readings obtained with the DCT in a patients file only after rounding them up or down to a value of a full unit (mm Hg). This is still more precise than the GAT readings used to date.
In the present study DCT was found to be suitable as a routine clinical tool for measuring IOP. Because the DCT can be attached to any slit lamp fitted with a normal GAT stand, the new tonometer can be used on most rigs without the need for modification. The examination technique with the DCT is similar to the technique used with the GAT, except that it does not require the sometimes cumbersome tuning of a knob to adjust two oscillating or melting semicircles, which leaves room for observer-dependent interpretation. All examiners involved in this study managed to obtain consistent readings with the DCT right from the beginning, without any learning curve. Measuring IOP with the DCT requires the tip of the tonometer to rest on the patients cornea for approximately 5 seconds. This is slightly longer than the contact time that an experienced examiner would require with the GAT. However, the acoustic signal of the DCT that informs the examiner about the correct alignment of the tonometer tip seems to encourage patients to remain still for the time needed.
In summary, this study found IOP measurements taken with the new DCT to have an excellent concordance with measurements obtained by the GAT. In the subjects studied, IOP measurements with the DCT did not depend on corneal thickness, corneal curvature, or axial length. With regard to increasing awareness of the effect of corneal thickness on IOP readings by GAT and the increasing number of patients with a history of corneal refractive surgery, DCT may offer some clinically relevant advantages over conventional Goldmann-type applanation tonometers for screening and management of patients with suspected or known glaucoma.
| Footnotes |
|---|
Disclosure: C. Kaufmann, None; L.M. Bachmann, None; M.A. Thiel, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Michael A. Thiel, Department of Ophthalmology, University of Zurich, Frauenklinikstr. 24, CH-8091 Zurich, Switzerland; michael.thiel{at}usz.ch.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. A. Read, M. J. Collins, and D. R. Iskander Diurnal Variation of Axial Length, Intraocular Pressure, and Anterior Eye Biometrics Invest. Ophthalmol. Vis. Sci., July 1, 2008; 49(7): 2911 - 2918. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. G. Boehm, A. Weber, L. E. Pillunat, R. Koch, and E. Spoerl Dynamic Contour Tonometry in Comparison to Intracameral IOP Measurements Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2472 - 2477. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Romppainen, L. M. Bachmann, C. Kaufmann, C. Kniestedt, M. Mrochen, and M. A. Thiel Effect of Riboflavin-UVA Induced Collagen Cross-linking on Intraocular Pressure Measurement Invest. Ophthalmol. Vis. Sci., December 1, 2007; 48(12): 5494 - 5498. [Abstract] [Full Text] [PDF] |
||||
![]() |
A R Ismail, M Lamont, S Perera, D Khan-Lim, R Mehta, J D A Macleod, and D F Anderson Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties Br. J. Ophthalmol., July 1, 2007; 91(7): 980 - 981. [Full Text] [PDF] |
||||
![]() |
M. C. Grieshaber, A. Schoetzau, C. Zawinka, J. Flammer, and S. Orgul Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle Glaucoma Arch Ophthalmol, June 1, 2007; 125(6): 740 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Cervino Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure. Br. J. Ophthalmol., December 1, 2006; 90(12): 1444 - 1446. [Full Text] [PDF] |
||||
![]() |
J M Gonzalez-Meijome, J Jorge, A Queiros, P Fernandes, R Montes-Mico, J B Almeida, and M A Parafita Age differences in central and peripheral intraocular pressure using a rebound tonometer Br. J. Ophthalmol., December 1, 2006; 90(12): 1495 - 1500. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Ravin and E. J. Higginbotham Are we there yet? Celebrating the centennial of the schiotz tonometer. Arch Ophthalmol, September 1, 2006; 124(9): 1337 - 1338. [Full Text] [PDF] |
||||
![]() |
C. Kaufmann, L. M. Bachmann, Y. C. Robert, and M. A. Thiel Ocular Pulse Amplitude in Healthy Subjects as Measured by Dynamic Contour Tonometry. Arch Ophthalmol, August 1, 2006; 124(8): 1104 - 1108. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Kotecha, E T White, J M Shewry, and D F Garway-Heath The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry Br. J. Ophthalmol., December 1, 2005; 89(12): 1572 - 1575. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kniestedt, S. Lin, J. Choe, A. Bostrom, M. Nee, and R. L. Stamper Clinical Comparison of Contour and Applanation Tonometry and Their Relationship to Pachymetry Arch Ophthalmol, November 1, 2005; 123(11): 1532 - 1537. [Abstract] [Full Text] [PDF] |
||||
![]() |
P-A Tonnu, T Ho, T Newson, A El Sheikh, K Sharma, E White, C Bunce, and D Garway-Heath The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry Br. J. Ophthalmol., July 1, 2005; 89(7): 851 - 854. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |