(Investigative Ophthalmology and Visual Science. 2005;46:4468-4473.)
© 2005 by The Association for Research in Vision and Ophthalmology, Inc.
DOI: 10.1167/iovs.05-0609
Postblink Changes in the Ocular Modulation Transfer Function Measured by a Double-Pass Method
Robert Montés-Micó,1,2
Jorge L. Alió,1,3 and
W. Neil Charman4
1From the Research, Development and Innovation Department, VISSUM Ophthalmologic Institute of Alicante, Alicante, Spain; the
2Department of Ophthalmology, Otorhinolaryngology, and Pathology, University of Murcia, Murcia, Spain;
3School of Medicine, University Miguel Hernández, Alicante, Spain; and the
4Department of Optometry and Neuroscience, Faculty of Life Sciences, University of Manchester, Manchester, United Kingdom.
 |
Abstract
|
|---|
PURPOSE. To examine the temporal changes in the modulation transfer function (MTF) of the eye after a blink.
METHODS. The distance MTF of a 5-mm pupil was derived from double-pass retinal images in 20 healthy young subjects at various intervals after a blink (1 second up to 15 seconds). Such measurements include the effects of wavefront aberration, scattered light, and errors in focus and are thus more relevant to real life than are estimates of MTF based on wavefront errors alone.
RESULTS. Optimal MTF by a variety of criteria was found some 6 seconds after a blink.
CONCLUSIONS. Inclusion of the effects of stray light and errors of focus does not affect the finding that optimal retinal image quality occurs some time after a blink. It does not appear that a loss in optical quality is the trigger of normal blinking.
The tear film is the most anterior optical element of the eye. The large refractive index change across the airtear interface means that any variation in the regularity of this surface caused by local changes in tear-film thickness can significantly alter the optical aberrations of the eye and hence the quality of the retinal image.1 2 Because the overall thickness of the tear film is usually accepted to be approximately 10 µm3 4 (although lesser thicknesses have been suggested5 ), substantial variations in optical path in wavelength terms across its surface are, in principle, possible. The basic, normal, dynamic changes in the tear film after its deposition during a blink have been modeled by Wong et al.,6 who show that they result from a complex interplay between factors such as lid forces, the structure of the film, its viscosity and surface tension,7 and forces due to the tear menisci at the lids. Experimental studies have given more detailed insights into the local variations across the surface of the film.8 9 10 Tear film changes are accelerated in patients with dry eye.11
Several experimental studies have confirmed that tear film changes influence ocular aberrations.12 13 14 15 16 17 18 19 20 It appears that the wavefront aberrations associated with both the anterior cornea and the whole eye pass through a minimum a few seconds after a blink. This occurs after approximately 6 seconds in normal eyes and 4 seconds in patients with dry eye.16 17 18 19 20 It is suggested that, after a blink, there is initially a fairly rapid decrease in the aberrations associated with the anterior surface of the tear film, as the tear film stabilizes and becomes smoother. This is followed by a gradual increase, as evaporation and other effects3 4 cause the tear film to become more irregular and to start to break up.16 17 18 19 20
If the wavefront aberration of the eye is known, it is, in principle, possible to calculate such descriptors of retinal image quality as the point-spread function (PSF) and optical transfer function (OTF) by using Fourier methods.21 A problem, however, is that other factors besides higher-order wavefront aberration can contribute to the time-dependent changes in retinal image quality. In particular, it is well known that local small-scale variations in anterior surface regularity, which are not adequately described by wavefront measurements, cause marked increases in forward light scattering with time, particularly after tear break-up starts to occur. This increased scattering is obvious in, for example, the decrease in clarity of the spot images given by Hartmann-Shack aberrometers12 13 17 and tend to broaden and depress the retinal PSF and reduce modulation transfer. Any scattering from other sourcesfor example, the bulk cornea, lens, and retinaproduces similar effects, although these do not change over the interblink interval. In general, estimates of retinal image quality based on wavefront aberration alone may be too optimistic22 and, if anterior corneal scattering increases at a different rate than wavefront aberration, the postblink time at which optimal image quality is obtained may be affected. A further relevant factor under real-life conditions, even in emmetropic or corrected eyes, is the possible presence of lags and leads in accommodation. The associated defocus blur (second-order aberration) on the retina reduces the relative importance of tear-induced changes in higher-order wavefront aberration.
In the present study, the changes in retinal image quality with time after a blink were measured directly under conditions of distance vision with free accommodation, by the well-established double-pass technique,23 24 to record the retinal point-spread function (PSF). The results are compared with those of studies based on wavefront aberration alone.
 |
Methods
|
|---|
Subjects
Twenty patients, 15 men and 5 women, participated in this study. Their ages ranged from 22 to 32 years (mean, 26.6 ± 3.1). All were emmetropes with a visual acuity of 20/20 or better and normal ocular health. Fluorescein tear break-up times (TBUTs) as measured by standard clinical methods25 were normal, ranging from 8 to 15 seconds among subjects. Mean TBUT was 10.4 ± 1.9 seconds and mean interblink interval was 5.7 ± 0.5 seconds. The study followed the tenets of the Declaration of Helsinki. Informed consent was obtained from all patients after the nature and possible consequences of the study had been explained.
MTF Measurement
The ocular MTF (the modulus of the OTF) was evaluated with a commercial system (Optical Quality Analysis System [OQAS]; Visiometrics, Terrassa, Spain). The instrument is based on the double-pass technique, which was developed to perform an objective optical quality measurement of the eye.23 24 The image of a monochromatic point source (780 nm) is first formed on the retina. The reflected light passes back through the eye, and the double-pass PSF in air is recorded by a low-noise digital camera. In the OQAS instrument, the entrance and exit pupils used to form the initial and final PSFs differ. The entrance pupil is small enough (1.5 mm) for the initial PSF to be diffraction limited, because ocular aberrations are negligible for small pupils.26 In contrast, a larger diameter (5 mm in the present study) is used for the exit pupil, so that the ocular aberrations affect the second stage of the formation of the double-pass PSF. Thus, when the modulus of the two-dimensional Fourier transform of the PSF is derived to yield the double-pass MTF, this function can be corrected for the known, monochromatic MTF of the first, diffraction-limited stage of image formation, to give the MTF due only to the second stage of the imaging processes, which gives the single-pass MTF of the eye for a pupil diameter equal to that of the exit pupil of the instrument. The advantage of this method is that it yields information on the asymmetric as well as the symmetric aberrations of the eye.27 28 Its disadvantage is that measurements cannot be made at higher spatial frequencies than the cut-off imposed by the 1.5-mm entrance pupil (34 cyc/deg). It is also necessary to assume an absence of scattering from lens sutures, the retina or other structures.29 For the purposes of the present study, the mean one-dimensional MTF was calculated as the average over all orientations of the two-dimensional MTF for the 5-mm pupil exit pupil used. It should be noted that averaging the MTF over all orientations leads to a loss in information on the presence of asymmetric aberrations like coma and astigmatism and tends to reduce intersubject differences associated with asymmetric aberrations. It was felt that this was justified in view of the greater ease in comparing effects at different times (described later).
Experimental Procedure
The subjects head was stabilized on a three-dimensional translating stage. The subject was instructed to blink three or four times and fixate on a distant image (a high-contrast Snellen E against a red background) created by the double-pass system while keeping the eyes wide open as long as possible. Subjects used natural pupils and were not under cycloplegia. However, an artificial pupil in the apparatus restricted the effective measurement pupil to 5 mm, the diameter of the natural pupil always being larger than this. During the period of nonblinking, 15 images were captured with the apparatus. The first was taken immediately after the blink (nominally t = 0) and the others at 1-second time intervals from 1 to 15 seconds after a blink. Only the left eye was used for the measurements, the other eye being occluded. To avoid possible longer-term changes in the cornea and/or tear film from successive periods of nonblinking, the experiments on each individual were performed on five separate days to yield five sets of measurements. The tests were run in a room with controlled temperature (23 ± 1°C) and humidity (40% ± 2%).
 |
Results
|
|---|
Figure 1 shows the changes in mean MTF of a 5-mm pupil (all subjects, all trials, and all orientations) as a function of time after a blink. The SD in modulation transfer at any time or spatial frequency was typically approximately 10% of the mean. Whereas there was little change in modulation transfer at lower spatial frequencies (e.g., 5 cyc/deg) with time, modulation transfer at higher spatial frequencies varied substantially, with optimal levels being reached approximately 6 seconds after a blink.

View larger version (44K):
[in this window]
[in a new window]
|
FIGURE 1. Calculated MTFs at each second post-blink for all subjects evaluated. Radial projection, averaged over all orientations, of the two-dimensional MTF (y-axis) versus spatial frequency (x-axis in cycles per degree).
|
|
This behavior is illustrated more clearly in Figure 2 , which plots the value of mean modulation transfer at a series of spatial frequencies as a function of time after a blink. Note that a high-contrast 30-cyc/deg grating is only likely to be visible over a limited period of
6 seconds after a blink.
 |
Discussion
|
|---|
Before discussing the variations in MTF with time in more detail, it is reasonable to ask how the present optimal mean MTFs for a 5-mm pupil, which are found approximately 6 seconds after a blink, compare with those in the literature, particularly those based on wavefront aberration alone. Figure 3 makes this comparison. The MTF by the method of Artal30 is based on the double-pass technique in monochromatic light and hence includes the effects of intraocular scattered light, while those of Walsh and Charman31 and Liang and Williams32 are derived from wavefront data alone and hence would be expected to be a little higher, due to the absence of the contrast-degrading effects of scattered light. All these earlier data were obtained with carefully corrected eyes under cycloplegia, so that errors of focus were minimized. Although the times after a blink at which these earlier data were obtained were not specified, it seems reasonable to assume that they represent near-optimal optical performance.

View larger version (24K):
[in this window]
[in a new window]
|
FIGURE 3. Comparison of MTFs with 5-mm pupils and the neural contrast threshold appropriate to this natural pupil size (found for a retinal illuminance of approximately 130 trolands, gray curve). The present data are the optimal MTF found approximately 6 seconds after a blink (780 nm). Other curves are from Artal30 (double-pass method under cycloplegia with optimal correction at 633 nm); Walsh and Charman3 (based on wavefront aberration data alone, 555 nm); and Liang and Williams32 (based on wavefront aberration data alone, 633 nm).
|
|
It can be seen that our optimal double-pass MTF was lower than all the others. It was also slightly more irregular. We attribute these effects mainly to a combination of errors of focus associated with accommodative lead for many of our subjects when viewing the distant target of our OQAS double-pass instrument under natural viewing conditions and the longitudinal chromatic aberration between the wavelengths of the red visible accommodation target and the infrared wavelength (780 nm) at which the PSFs were measured. The signs of these defocus effects are opposite one another. Although accommodative leads (excesses in ocular power) for a target at infinity vary between subjects, at photopic levels, they typically amount to approximately 0.50 D.33 In contrast, the reduction in apparent ocular power associated with the ocular longitudinal chromatic aberration and the fact that infrared is reflected from deeper retinalchoroidal layers than the receptors probably also amounts to approximately 0.5 D for a wavelength of approximately 780 nm,34 so that there is an approximate balance between the two defocus effects. The normal fluctuations in accommodation, over a range of approximately 0.5 D,35 will be superimposed on the mean defocus and cause variations in individual sampled MTFs, even in the absence of changes in higher-order aberration. Last, there may be some residual uncorrected second-order spherical and cylindrical wavefront errors in our "emmetropic" subjects, whereas second-order aberrations have been removed from the wavefront aberration-based MTFs in Figure 3 . Higher-order aberrations are essentially the same in visible and the infrared34 when expressed in micrometers, but the same wavefront errors will have less impact in the infrared because the wavelength is longer. In contrast, image-degrading light-scattering within the retina and choroid may be higher at the longer wavelength, so that overall the MTF measured in the near infrared would not be expected to differ greatly from that which would be measured if visible light were to be used. An error of focus of approximately 0.25 D or less would be sufficient to explain the discrepancy between our optimal MTF and that in the earlier work.36 Finally, we note that that the use of the small (1.5 mm) entrance pupil in our equipment and a wavelength of 780-nm limits spatial frequencies reaching the retina to a maximum of approximately 35 cyc/deg, so that we cannot make measurements at spatial frequencies higher than this; however, this effect does not explain our comparatively low readings at spatial frequencies of approximately 10 cyc/deg.
If we now consider the postblink changes in MTF recorded in the present experiment, Figure 2 suggests strongly that, if high spatial frequencies are important, optical performance is optimal after approximately 6 seconds. It is, however, of interest to use other performance criteria to see whether they lead to the same conclusion. A large number of MTF-based, single-figure performance metrics have been suggested in the literature.37 We have chosen to use the following: the cut-off frequency at which the mean MTF falls to zero; the visual high-contrast cut-off frequency at which the MTF intersects the neural contrast threshold curve; the area under the MTF; the area between the MTF and the neural contrast threshold; the volume under the two-dimensional MTF (which relates to the Strehl ratio2 ); and the volume between the two-dimensional MTF and the corresponding two-dimensional neural threshold (which relates to the visual Strehl ratio37 ). Objective refraction based on finding the correction that maximizes the volume under the MTF has been shown to correlate well with subjective refraction.38 The results of the six image metrics as a function of time after a blink are shown in Figure 4 . The neural contrast threshold used, which is plotted in Figure 3 , is that appropriate to the luminance level that typically leads to a 5-mm natural pupil and gives a retinal illuminance of 130 trolands.39 40

View larger version (21K):
[in this window]
[in a new window]
|
FIGURE 4. MTF metrics: cutoffs for MTF alone and MTF with neural contrast threshold (A), area under MTF and between MTF and neural threshold (B), and volume under MTF and between MTF and neural threshold (C).
|
|
It can be seen that all the performance measures peaked at approximately 5 to 6 seconds. The inclusion of neural as opposed to purely optical effects may tend to move the peak performance to slightly shorter times, presumably because it reduces the importance of modulation transfer at high spatial frequencies. In general, though, it appears that assessing optical performance by direct measurement of the retinal MTF leads to the same conclusion as was reached on the basis of wavefront measurements16 17 that optical performance in normal subjects is best some 6 seconds or so after a blink.
This result is perhaps not surprising, in that marked increases in light scattering from the eyes anterior surface would be expected to arise only after the tear film has started to break up, which occurred at an average of approximately 10 seconds in our subjects. Before this time, dynamic changes in both the wavefront aberration and the single-pass MTF are caused mainly by aberration changes associated with the local alterations in the thickness of the still-unbroken tear layer although, as noted earlier, the MTF is additionally degraded by time-invariant intraocular scattering and by focus errors. Note too that we cannot exclude the possibility that lid pressure during blinking has some effect on corneal contour and that the contour changes during the interval between blinks to produce local changes in optical path that are independent of tear film changes. Buehren et al.41 have interpreted their videokeratoscope data in terms of effects of this kind, and their results suggest that the upper and lower areas of a 5-mm pupil may be affected.
The results of this study therefore support those of earlier work based purely on wavefront aberration measurements. Optimal optical performance in normal subjects occurs some 5 to 6 seconds after a blink. If indeed optical quality continues to improve over this timescale, it may be wondered why blink intervals are usually comparable to or shorter than this (normal blink intervals were around 6 seconds in our subjects). Our results provide no evidence that, in normal subjects, blinking is delayed until tear film break-up starts within the corneal area corresponding to that of the pupil. The discomfort caused by spot-wise drying of the precorneal tear film has been suggested to be an initiating factor for blinking in some normal persons and patients with dry eye.42 The present results imply that most blinking is not triggered by any detectable deterioration in the retinal image (as is, of course, also evidenced by the fact that blink rates are similar in darkness), and it may be that the local dry spots involved in the initiation of some blinks are usually located in the peripheral cornea,42 where they have no impact on axial image quality. Another possible factor in blink initiation is that blinking may also be an unconscious part of the mechanism to correct for slow drifts in fixation, accommodation and convergence. The basic blink rate appears to be a characteristic of the individual, although it can, of course, be affected by a variety of external stimuli, such as stress or task difficulty.43 44 45
 |
Footnotes
|
|---|
Supported in part by a grant from the Spanish Ministry of Health, Instituto de Salud Carlos III, Red Tematica de Investigación Cooperativa en Oftalmología (Ref. C 03/13) Subproyecto Cirugía Refractiva y Calidad Visual.
Submitted for publication May 17, 2005; revised June 20, 2005; accepted September 14, 2005.
Disclosure: R. Montés-Micó, None; J.L. Alió, None; W.N. Charman, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Robert Montés-Micó, Research, Development and Innovation Department, VISSUM Instituto Oftalmológico de Alicante, Avda, Denia s/n, Edificio VISSUM, 03016, Alicante, Spain; roberto.montes{at}uv.es.
 |
References
|
|---|
- Smirnov MS. Measurement of the wave aberration of the human eye. Biofizika. 1961;6:687703. English translation. Biophysics. 1962;7:776795.
- Albarrán C, Pons AM, Lorente A, Montés-Micó R, Artigas JM. Influence of the tear film on optical quality of the eye. Contact Lens Ant Eye. 1997;20:129135.[CrossRef][Medline][Order article via Infotrieve]
- Holly FJ, Lemp MA. Tear physiology and dry eyes. Surv Ophthalmol. 1977;22:6987.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Craig J. Structure and function of the preocular tear film. The Tear Film: Structure, Function and Clinical Examination. 2002;1850. Butterworth-Heinemann Oxford, UK.
- King-Smith PE, Fink BA, Hill RM, Koelling KW, Tiffany JM. The thickness of the tear film. Curr Eye Res. 2004;29:357368.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Wong H, Fatt I, Radke CJ. Deposition and thinning of the human tear film. J Colloid Interface Sci. 1996;184:4451.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Berger RE, Corrsin S. A surface tension gradient mechanism for driving the precorneal tear film after a blink. J Biomech. 1974;7:225238.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Benedetto DA, Clinch TE, Laibson PR. In vivo observation of tear film dynamics using fluorophotometry. Arch Ophthalmol. 1984;102:410412.[Abstract]
- Licznerski TJ, Kasprzak HT, Kowalik W. Analysis of shearing interferograms of tear film using fast-Fourier transforms. J Biomed Opt. 1998;3:2337.
- Németh J, Erdélyi B, Csákány B, et al. High-speed videokeratographic measurement of tear film build-up time. Invest Ophthalmol Vis Sci. 2002;43:17831790.[Abstract/Free Full Text]
- Johnson ME, Murphy PJ. Changes in the tear film and ocular surface from dry eye syndrome. Prog Retin Eye Res. 2004;23:449474.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Thibos LN, Hong X. Clinical applications of the Shack-Hartmann aberrometer. Optom Vis Sci. 1999;76:817825.[ISI][Medline][Order article via Infotrieve]
- Tutt R, Bradley A, Begley C, Thibos LN. Optical and visual impact of tear break-up in human eyes. Invest Ophthalmol Vis Sci. 2000;41:41174123.[Abstract/Free Full Text]
- Koh S, Maeda N, Kuroda T, et al. Effect of tear film break-up on higher-order aberrations measured with wavefront sensor. Am J Ophthalmol. 2002;134:115117.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Himebaugh NL, Wright AR, Bradley A, Begley CG, Thibos LN. Use of retroillumination to visualize optical aberrations caused by tear film break-up. Optom Vis Sci. 2003;80:6978.[Medline][Order article via Infotrieve]
- Montés-Micó R, Alió JL, Muñoz G, Charman WN. Temporal changes in optical quality of air-tear film interface at anterior cornea after blink. Invest Ophthalmol Vis Sci. 2004;45:17521711.[Abstract/Free Full Text]
- Montés-Micó R, Alió JL, Muñoz G, Pérez-Santonja JJ, Charman WN. Postblink changes in total and corneal ocular aberrations. Ophthalmology. 2004;111:758767.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Montés-Micó R, Cáliz A, Alió JL. Wavefront analysis of higher-order aberrations in dry eye patients. J Refract Surg. 2004;20:243247.[ISI][Medline][Order article via Infotrieve]
- Montés-Micó R, Cáliz A, Alió JL. Changes in ocular aberrations after artificial tears instillation in dry eye patients. J Cataract Refract Surg. 2004;30:16491652.[Medline][Order article via Infotrieve]
- Montés-Micó R, Alió JL, Charman WN. Dynamic changes in the tear film in dry eyes. Invest Ophthalmol Vis Sci. 2005;46:16151619.[Abstract/Free Full Text]
- Welford WT. Aberrations of Optical Systems. 1986;240260. Adam Hilger Bristol, UK.
- Williams DR, Brainard DH, McMahon MJ, Navarro R. Double-pass and interferometric measures of the optical quality of the eye. J Opt Soc Am A. 1994;11:31233135.[ISI][Medline][Order article via Infotrieve]
- Campbell FW, Gubisch RW. Optical quality of the human eye. J Physiol (London). 1966;186:558578.[Abstract/Free Full Text]
- Santamaría J, Artal P, Bescos J. Determination of the point spread function of the human eye using a hybrid optical-digital method. J Opt Soc Am A. 1987;4:11091114.[ISI][Medline][Order article via Infotrieve]
- Cho P, Leung L, Lam A, Choi A. Tear break-up test: clinical procedures and their effects. Ophthalmic Physiol Opt. 1998;18:319324.[Medline][Order article via Infotrieve]
- Berny F, Slansky S. Wavefront determination resulting from Foucault test as applied to the human eye and visual instruments. Optical Instruments and Techniques. 1969;375385. Oriel Press London.
- Artal P, Marcos S, Navarro R, Williams DA. Odd aberrations and double-pass measurements of retinal image quality. J Opt Soc Am A. 1995;12:195201.[ISI][Medline][Order article via Infotrieve]
- Artal P, Iglesias I, Lopez-Gil N. Double-pass measurements of retinal-image quality with unequal entrance and exit pupil sizes and the reversibility of the eyes optical system. J Opt Soc Am A. 1995;12:23582366.[ISI][Medline][Order article via Infotrieve]
- Navarro R, Losada MA. Phase transfer and point-spread function of the human eye determined by a new asymmetric double-pass method. J Opt Soc Am A. 1995;12:23852392.
- Artal P. Calculations of two-dimensional foveal retinal images in real eyes. J Opt Soc Am A. 1990;7:13741381.[Medline][Order article via Infotrieve]
- Walsh G, Charman WN. Measurement of the axial wavefront aberration of the human eye. Ophthalmic Physiol Opt. 1985;5:2331.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Liang J, Williams DR. Aberrations and retinal image quality of the normal human eye. J Opt Soc Am A. 1997;14:28732883.[ISI][Medline][Order article via Infotrieve]
- Johnson CA. Effects of luminance and stimulus distance on accommodation and visual resolution. J Opt Soc Am. 1976;66:138142.[Medline][Order article via Infotrieve]
- Llorente L, Diaz-Santana L, Lara-Saucedo D, Marcos S. Aberrations of the human eye in visible and near infrared illumination. Optom Vis Sci. 2003;80:2635.[Medline][Order article via Infotrieve]
- Charman WN, Heron G. Fluctuations in accommodation: a review. Ophthalmic Physiol Opt. 1988;8:153164.[ISI][Medline][Order article via Infotrieve]
- Charman WN, Jennings JAM. The optical quality of the monochromatic retinal image as a function of focus. Br J Physiol Opt. 1976;31:119134.[Medline][Order article via Infotrieve]
- Thibos LN, Hong X, Bradley A, Applegate RA. Accuracy and precision of objective refraction from wavefront aberrations. J Vision. 2004;4:329351.[CrossRef]
- Guirao A, Williams DR. A method to predict refractive errors from wave aberration data. Optom Vis Sci. 2003;80:3642.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Charman WN, Chateau N. The prospects for super-acuity: limits to visual performance after correction of monochromatic ocular aberration. Ophthalmic Physiol Opt. 2003;23:479493.[CrossRef][Medline][Order article via Infotrieve]
- Charman WN. Ablation design in relation to spatial frequency, depth-of-focus and age. J Refract Surg. 2004;20:S542S549.[Medline][Order article via Infotrieve]
- Buehren T, Collins MJ, Iskander DR, Davis B, Lingelbach B. The stability of corneal topography in post-blink interval. Cornea. 2001;20:826833.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Brown SI. Further studies on the pathophysiology of keratitis sicca of Rollet. Arch Ophthalmol. 1970;83:542547.[ISI][Medline][Order article via Infotrieve]
- Davson H. Physiology of the Eye. 1990;791. Macmillan London.
- York M, Ong J, Robbins JC. Variation in blink rate associated with contact lens wear and task difficulty. Am J Optom Arch Am Acad Optom. 1971;48:461466.[Medline][Order article via Infotrieve]
- Patel S, Henderson R, Bradley L, Galloway B, Hunter L. Effect of visual display unit use on blink rate and tear stability. Optom Vis Sci. 1991;68:888892.[Medline][Order article via Infotrieve]
This article has been cited by other articles:

|
 |

|
 |
 
N. Lopez-Gil, V. Fernandez-Sanchez, R. Legras, R. Montes-Mico, F. Lara, and J. L. Nguyen-Khoa
Accommodation-Related Changes in Monochromatic Aberrations of the Human Eye as a Function of Age
Invest. Ophthalmol. Vis. Sci.,
April 1, 2008;
49(4):
1736 - 1743.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Koh, N. Maeda, Y. Hirohara, T. Mihashi, S. Ninomiya, K. Bessho, H. Watanabe, T. Fujikado, and Y. Tano
Serial measurements of higher-order aberrations after blinking in normal subjects.
Invest. Ophthalmol. Vis. Sci.,
August 1, 2006;
47(8):
3318 - 3324.
[Abstract]
[Full Text]
[PDF]
|
 |
|