(Investigative Ophthalmology and Visual Science. 2006;47:4373-4377.)
© 2006 by The Association for Research in Vision and Ophthalmology, Inc.
DOI:  10.1167/iovs.06-0536

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Deb-Joardar, N.
Right arrow Articles by Gain, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Deb-Joardar, N.
Right arrow Articles by Gain, P.

Standard Microlithographic Mosaics to Assess Endothelial Cell Counting Methods by Light Microscopy in Eye Banks Using Organ Culture

Nilanjana Deb-Joardar,1,2 Gilles Thuret,1,2 Georges-André Racine,3 David Pons,4 Gerald Brun,4 Olivier Parriaux,4 Michel Peoc’h,1 Sophie Acquart,1,5 and Philippe Gain1

1From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France; 3Center of MicroNanoTechnology, Ecole Polytechnique Fédérale de Lausanne, Lausanne, Switzerland; 4Laboratory of Signal Treatment and Instrumentation, Université Jean Monnet, Saint-Etienne, France; and 5French Blood Center/Eye Bank of Saint Etienne, Saint Etienne, France.


    Abstract
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
PURPOSE. To develop standard microscopic hexagonal mosaics mimicking the human corneal endothelium for quality control of endothelial cell density (ECD) measurement and verification of cell counting strategy by light microscopy in eye banks using organ culture.

METHODS. A standard slide, the Keratotest, was developed with 10 laser-engraved mosaics and different predetermined "cell" densities representing the range of ECDs observed routinely. Horizontal and vertical micrometric scales were etched adjacently to each mosaic, and a standard microscopy resolution test pattern was included. The Keratotest was applied to assess the reliability of a computer-assisted analyzer developed for corneal endothelial evaluation based on light microscopy images.

RESULTS. The Keratotest consisted of 10 microlithographic homogeneous mosaics of 1-mm2 printed area and 1.2-µm cell boundary thickness. The micrometric scale associated with each mosaic aided in simultaneous verification of microscope calibration, and the test pattern aided in checking the microscope resolution. The design was unalterable and reproducible, and the glass slide incorporated in a carbon fiber support ensured easy handling and safe transport. Evaluation of the Keratotest mosaics by the computer-assisted analyzer found a high level of agreement (error margin between +0.12 and –0.46%) with the laser-engraved cell density.

CONCLUSIONS. This prototype device enabled assessment of reliability of ECD measurement in eye banks. It also allowed verification of the calibration and resolution of light microscopes. Periodic validation of counting procedure in eye banks with mosaics of known "cell" densities should be useful for standardization of donor corneal tissue quality control.


Endothelial cell density (ECD) is the principal criterion for quality evaluation of donor corneal tissue in eye banks. The ECD of the grafted cornea determines its survival with time in terms of maintenance of transparency.1 2 An ECD of 2000 to 2400 cells/mm2 is considered by most eye banks in Europe to be the standard cutoff for corneal delivery.3 ECD is estimated on a small sample of few hundred cells out of an estimated 250,000 to 350,000 cells present in the cornea. It is therefore imperative to have a precise and reproducible method of ECD evaluation that would help standardize donor tissue evaluation methods and exchange between eye banks.

In Europe, where organ culture is the storage method of choice, the endothelium is viewed directly under a light microscope after osmotic preparation consisting of incubation of the endothelial face of the cornea for 1 to 4 minutes in 0.9% sodium chloride or 1.8% sucrose.4 This dilates the intercellular spaces and enables the visualization of cell contours that are otherwise invisible. From this preparation, cells are most often counted manually in real time by observation through a calibrated reticule fitted in the microscope eyepiece or by indication of the cells in a known surface on endothelial photographs. Recent years have seen the emergence of prototypes and commercially available computer-assisted counting systems, currently used in some European eye banks,3 in which analysis of digitized optical microscopy images allows rapid and effective measurement of ECD of organ-cultured corneas.5 6 7 8

Previous studies analyzing ECD measurement methods in 22 eye banks in France have emphasized the need for standardization of counting methods.9 10 Lack of microscope calibration had been identified as a major factor responsible for gross variation in ECD estimates between eye banks. Differences in manual counting techniques were also shown to influence the ECD values. Until now, the only method of calibration available was a certified graduated micrometer, a traditional method in microscopy. Essentially, this measures only the magnification ratio between the tissue observed and its image projected in a calibrated reticule or by image analysis software. Whatever the method used, manual or computerized, periodical calibration of the measuring equipment and standardization of counting procedures are essential to ensure reliability of intrabank and interbank cell counts. To address this issue, we present a prototype standard slide, called Keratotest, carrying a series of hexagonal mosaics of different densities mimicking a range of human corneal endothelia.


    Methods
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
Specifications
The design of the Keratotest had to meet certain essential requirements. It had to correspond to microscopic hexagonal mosaics and mimic the human corneal endothelium, particularly "cell" area and "intercellular" spaces, as close to the reality as possible when observed under standard light microscopes. Cell density (CD) had to be perfectly mathematically predetermined to avoid any bias and had to represent the range of ECDs observed commonly in eye banks (we previously showed that errors in manual counting could partly depend on ECD10 ). Possibility of simultaneous calibration of the microscope by a classic graduated micrometer would facilitate a complete system check with a single device. Verification of the resolving power of the optical components (eyepiece and objective) of the microscope would be an additional advantage. To withstand mechanical transportation without structural damage, the material had to be durable and unalterable. The manufacturing process had to be reliable and reproducible to facilitate production of a series.

Manufacturing Process
The Keratotest consisted of a series of laser-engraved regular microlithographic mosaics with predetermined CDs. The steps of design and manufacture are described briefly here.

Calculation of CD
Given that a hexagon is made of six equilateral triangles of side length (L) and height (H), from Pythagorean theory it is known that H = {surd}3/2;L. Therefore, the area of each triangle is given by 1/2LH = {surd}3/4L2, and the area of a hexagon is 6 x {surd}3/4L2 = 3{surd}3/2L2. From this, we determined mathematically the number of the hexagons that could be engraved within a given surface measuring 1 mm2 so as to obtain a pattern of 10 mosaics of different CDs. The cell border area was calculated according to the Bourne hypothesis, by which the intercellular space is assumed to be negligible and the true cell border is located at the center of each border line.11 Therefore, the total surface area of each cell was given by its internal surface area plus half the cell border area. Next, from the CD, we calculated the best-fit circle diameter of the hexagon, which enabled us to obtain its vectorial coordinates and to generate the holographic pattern.

Creation of Holographic Pattern
High-quality, two-dimensional drawings of 10 mosaic patterns with adjacent horizontal and vertical micrometer were obtained in DXF format (Drawing eXchange Format; AutoCAD software; Autodesk SA, Neuchatel, Switzerland) containing the holographic data. A simplified microscopy resolution test pattern conforming to the National Bureau of Standards (NBS 1010a Microscopy Test Chart) consisting of three bars of equal spatial frequency in the horizontal and vertical directions was also designed. These were next converted to the machine-specific CIF (Caltech Intermediate File format) format using a translator (CAD; LinkCAD, Burlington, IA).

Photolithography
Photolithographic mosaics were made at the Swiss Federal Institute of MicroNanoTechnology in Lausanne, Switzerland. A glass wafer was coated with a 32-nm–thick chromium layer in a vacuum chamber on the principle of evaporation. The chromium layer was covered with a 1.5-µm–thick layer of S1813 photoresist (Rohm and Haas Electronic Materials, Coventry, UK) with the use of a spin coater. The holographic pattern was generated by a high-resolution, laser beam direct writing lithography machine (DWL200 Laser Writer; Heidelberg Instruments Mikrotechnik GmbH, Heidelberg, Germany) that uses a Kr laser light source (413-nm wavelength). The target substrate was placed on an XY table with Mach-Zender interferometer position feedback, which permitted high-precision movement with a 50-nm encoding resolution and a 2-µm depth of focus.

Generation of Test Slide
The exposed pattern was transferred to the glass wafer by standard photoresist development and selective etching of exposed chromium. Etched samples were characterized by optical microscopy and scanning electron microscopy. The photoresist, developed by analogy to a photographic film, was baked at 120°C on a hotplate to make it hard and resilient to the etching solution. The substrate was then immersed in an etching solution containing ceric ammonium nitrate. At the end of the process, resin residues were eliminated with a stripping solution. The engraved surface was shielded in a coverslip mounted with resin (Eukitt; ProSciTech, Kirwan, Australia) to ensure mechanical protection. The mount was then cut and incorporated into an unalterable carbon-fenestrated support of dimensions similar to those of a microscope slide (76 x 26 mm).

Example of Application to Computer-Assisted Counting System
We used the Keratotest to assess a tri-image endothelial analyzer (Sambacornea; Samba Technologies, Meylan, France) developed in our laboratory specifically for the evaluation of ECD and morphometry based on light microscopy images.7 The analyzer itself was calibrated with a graduated micrometer of 1-mm certified length (Leica Microsystems, Rueil-Malmaison, France) different from that included in the Keratotest. The 10 mosaics were counted with the tri-image endothelial analyzer five times each by two independent observers. Each time, three random images of a mosaic were captured after displacement and rotation of the Keratotest to ensure that each view was unique. Images were analyzed in a masked fashion to obtain a mean CD for that observation. Each observer selected a different counting zone on each mosaic each time to enable analysis of widely varying CDs. Selections were subjected to automated analysis, as described elsewhere.7 Analysis was performed only by threshold adjustment, and no manual touch-up of contour segments was made. For each mosaic, the (mean ± SD) CD was calculated, and the percentage error was compared with the actual CD of the laser-engraved mosaic. Results of the two analyses were compared.


    Results
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
Presentation of the Keratotest
The Keratotest consisted of homogeneous hexagons of 10 different CDs identified by the letters A to J engraved in a central window of a single glass slide (Fig. 1) . High-precision laser engraving allowed drawing of cells with boundaries of 1.2 µm, as observed for human endothelial cells after osmotic dilation (Fig. 2) . The exact CD of each of the 10 mosaics was recalculated after taking into consideration the necessary approximations (for design of hexagons, calculations were rounded off to four decimal places), precision of engraving, and tolerance of the laser writer; the difference between the theoretical CD and the final CD was negligible (Table 1) . The graduated micrometer, placed adjacently to each mosaic, enabled simultaneous vertical and horizontal calibration of the microscope. The test pattern helped to verify the optical resolution of the microscope. Resolution was determined by observing the line pair pattern, where all three lines and spaces in both the horizontal and the vertical direction could be discerned. Spatial frequency, expressed in terms of cycles per millimeter, was indicated by the numeric value shown with each test element. The glass slide incorporated in the carbon fiber support ensured easy maintenance, handling, and safe transport.


Figure 1
View larger version (97K):
[in this window]
[in a new window]
 
FIGURE 1. Global view of the Keratotest. Small gray circles: ten laser-engraved mosaics (AJ) of various cell densities. Cell densities (in cells/mm2) are: (A) 800; (B) 2400; (C) 3200; (D) 1800; (E) 2000; (F) 2200; (G) 1200; (H) 2800; (I) 1600; and (J) 3600. Large circle: position of the microscopy resolution test pattern. Original magnification, x4.

 

Figure 2
View larger version (89K):
[in this window]
[in a new window]
 
FIGURE 2. (A) Magnified view of the mosaic, G, of the Keratotest showing the regular hexagonal pattern with a cell density of 1200/mm2 and the horizontal and vertical micrometer axes with 100-µm primary, 50-µm secondary, and 10-µm tertiary graduations. Original magnification, x4. (B) Magnified view of inset showing the intercellular margins (thickness, 1.2 µm) and micrometer details. Original magnification, x10.

 

View this table:
[in this window]
[in a new window]
 
TABLE 1. Comparison between the Desired Mathematically Calculated Cell Density and the Final Cell Density Engraved on Keratotest

 
Application to Computer-Assisted Counting System
The CD estimation of each mosaic (mean ± SD of five measures) of the Keratotest performed by the tri-image endothelial analyzer (Fig. 3) found a level of agreement ranging from +0.12% to –0.46% compared with the laser-engraved CD (Table 2) .


Figure 3
View larger version (81K):
[in this window]
[in a new window]
 
FIGURE 3. Printout of the final analysis results of Keratotest mosaic A by the tri-image endothelial analyzer. Blue line, arrow: selected analysis zone. Green: contour recognition pattern as obtained after appropriate thresholding. Orange: cells represent automatic detection of probable errors in contour recognition. The section on the right shows the cell density (including number of cells counted) and morphometric data (coefficient of variation [CV] of cell area, minimum/maximum/mean cell area, SD, and percentage of hexagonal cells). Note that for mosaic A containing 800 cells, the calculated cell density was 797 cells/mm2, and CV was 1% based on a total of 1859 cells counted on three images. Hexagonality is 100% based on a total of 1578 cells counted on three images (cells intersected by the blue line were excluded for hexagonality measurement).

 

View this table:
[in this window]
[in a new window]
 
TABLE 2. Comparison of the Cell Density of Keratotest and That Calculated by the Sambacornea Analyzer

 

    Discussion
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 
We developed a prototype standard slide to assess the reliability of ECD measurement with light microscopy in eye banks using organ culture. Additionally, the slide allowed verification of microscope calibration and optical resolution. Reliability of the counting procedure could be assessed by a periodic validation of eye bank counting methods using the range of mosaics with known CDs.

Quality assessment of donor tissue and accreditation of medical personnel are important considerations in eye banks worldwide, and the reliability and precision of endothelial cell counting and its periodic standardization are essential prerequisites. In contrast to standardization methods prevalent in other medical disciplines, in which a sample of predetermined value is used for comparison or calibration, no similar tool is available for application to endothelial cell count. A corneal donor tissue sample cannot be used for calibration because no two corneas are alike. We have previously demonstrated that ECD estimation was neither precise nor reproducible in French eye banks.9 10 Inadequate microscope calibration and nonuniform counting procedures (difference in number of cells assessed, number of reticule units counted, and count strategy—"border" or "zone") had been identified as the major factors for the high intrabank and interbank variations of ECD. Human error might also have contributed to this variation because a tendency to underestimate was noted in corneas with high ECDs. Based on these findings, we advocated the urgent need for standardization of cell counting methods to prevent the delivery of corneas with low ECD at certain centers and wastage in others.

As a first step in this direction, we designed Alizarin red–stained test corneas to evaluate the reproducibility of cell counts between eye banks.10 Endothelial cells were rendered perfectly visible because of membrane staining, thereby eliminating visualization problems encountered during routine light microscopy resulting from nonuniform response of the endothelial monolayer to the osmotic challenge.5 Flat mounting of the dissected endothelium enabled visualization of the entire endothelium in a single plane, thus eliminating field depth–related measurement errors caused by the combined effects of physiologic corneal curvature and posterior corneal folding. Stained endothelial flat mounts, though representative of the real human endothelium, had certain limitations: (1) the ECD could not be predetermined mathematically, and (2) the slides were not reproducible because each human endothelium is unique. Hence, manufacture of a series to facilitate interbank exchange was not possible. A different standard that would represent the real endothelium but would be mathematically accurate and reproducible and would be useful for training new technicians was required to assess the reliability of eye bank counting methods.

The importance of calibrating specular microscopes for in vivo ECD counts has been emphasized in previous studies.12 13 Different analysis techniques of specular microscopy images—i.e., use of a fixed-frame or a center method—led to important interbank ECD variation in the Cornea Donor Study.14 Light microscopic evaluation used for organ-cultured corneas also has similar prerequisites. Computer-assisted analyzers also require calibration of the optical and the computing devices. In the tri-image endothelial analyzer, this is achieved by use of a calibration factor derived from a standard micrometric slide. However, verification of counting procedures or detection of human errors is not spontaneously possible. Our photolithographic device was designed for simultaneous calibration of the microscope and the counting device (reticule, printed grid, or digital image analyzer) and could be adapted to manual and computer-assisted cell counts. The extremely narrow limits of difference between the mathematically determined Keratotest ECD and that obtained after repeated counts by different observers with the computer-assisted analyzer permitted assessment of the reliability of counting methods.

Patterning of flat surfaces by use of photolithography is a high-precision technology widely used in industrial applications, such as in the manufacture of integrated circuits and microsystems.15 Choice of this particular method guaranteed the etching precision and hence the accuracy of the CDs obtained. Mathematical predetermination allowed creation of the exact CD desired, thus permitting a large range of ECDs (800–3200 cells/mm2) likely to be encountered in eye banks. The manufacturing process made it possible to create reproducible slides on a semi-industrial scale to meet the requirements of potential users.

The current prototype allows only validation of the ECD count. Further modifications with mosaics having "cells" of different surface area and shape are ongoing to enable assessment of morphometric parameters increasingly recognized as valuable indicators of donor tissue quality.16 17


    Acknowledgements
 
The authors thank Samuel Proton for assistance with vectorization and Souad Herrag for technical help.


    Footnotes
 
2 Contributed equally to the work and therefore should be considered equivalent authors. Back

Supported by Grant Recherche et greffes 2004 from l’Etablissement français des Greffes.

Submitted for publication May 15, 2006; revised June 5, 2006; accepted August 23, 2006.

Disclosure: N. Deb-Joardar, None; G. Thuret, None; G.-A. Racine, None; D. Pons, None; G. Brun, None; O. Parriaux, None; M. Peoc’h, None; S. Acquart, None; P. Gain, None

The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.

Corresponding author: Philippe Gain, Service d’Ophtalmologie (pavillon 50A), CHRU de Bellevue, 25 Boulevard Pasteur, F 42055 Saint Etienne Cedex 2, France; philippe.gain{at}univ-st-etienne.fr.


    References
 Top
 Abstract
 Methods
 Results
 Discussion
 References
 

  1. Nishimura JK, Hodge DO, Bourne WM. Initial endothelial cell density and chronic endothelial cell loss rate in corneal transplants with late endothelial failure. Ophthalmology. 1999;106:1962–1965.[CrossRef][ISI][Medline][Order article via Infotrieve]
  2. Bourne WM. Cellular changes in transplanted human corneas. Cornea. 2001;20:560–569.[CrossRef][ISI][Medline][Order article via Infotrieve]
  3. European Eye Bank Association Directory. 2006; 14th ed. European Eye Bank Association Amsterdam.
  4. Pels L. Organ culture: the method of choice for preservation of human donor corneas. Br J Ophthalmol. 1997;81:523–525.[Free Full Text]
  5. Barisani-Asenbauer T, Baumgartner I, Grabner G, Stur M. Automated digital image analysis of organ culture preserved donor corneas. Ophthalmic Res. 1993;25:94–99.[ISI][Medline][Order article via Infotrieve]
  6. Reinhard T, Spelsberg H, Holzwarth D, Dahmen N, Godehardt E, Sundmacher R. Endothelial evaluation of corneal transplants by digital imaging. Klin Monatsbl Augenheilkd. 1999;214:407–411.[Medline][Order article via Infotrieve]
  7. Gain P, Thuret G, Kodjikian L, et al. Automated tri-image analysis of stored corneal endothelium. Br J Ophthalmol. 2002;86:801–808.[Abstract/Free Full Text]
  8. Ruggeri A, Grisan E, Jaroszewski J. A new system for the automatic estimation of endothelial cell density in donor corneas. Br J Ophthalmol. 2005;89:306–311.[Abstract/Free Full Text]
  9. Thuret G, Manissolle C, Acquart S, et al. Is manual counting of corneal endothelial cell density in eye banks still acceptable? The French experience. Br J Ophthalmol. 2003;87:1481–1486.[Abstract/Free Full Text]
  10. Thuret G, Manissolle C, Acquart S, et al. Urgent need for normalization of corneal graft quality controls in French eye banks. Transplantation. 2004;78:1299–1302.[CrossRef][ISI][Medline][Order article via Infotrieve]
  11. Bourne WM. Morphologic and functional evaluation of the endothelium of transplanted human corneas. Trans Am Ophthalmol Soc. 1983;81:403–450.[Medline][Order article via Infotrieve]
  12. van Schaick W, van Dooren BT, Mulder PG, Volker-Dieben HJ. Validity of endothelial cell analysis methods and recommendations for calibration in Topcon SP-2000P specular microscopy. Cornea. 2005;24:538–544.[CrossRef][ISI][Medline][Order article via Infotrieve]
  13. Ohno K, Nelson LR, McLaren JW, Hodge DO, Bourne WM. Comparison of recording systems and analysis methods in specular microscopy. Cornea. 1999;18:416–423.[CrossRef][ISI][Medline][Order article via Infotrieve]
  14. Lass J, Gal R, Ruedy K, et al. An evaluation of image quality and accuracy of eye bank measurement of donor cornea endothelial cell density in the Specular Microscopy Ancillary Study. Ophthalmology. 2005;112:431–440.[CrossRef][ISI][Medline][Order article via Infotrieve]
  15. Madou M. Fundamentals of Microfabrication. 1997; CRC Press Boca Raton, FL.
  16. Yee RW, Matsuda M, Schultz RO, Edelhauser HF. Changes in the normal corneal endothelial cellular pattern as a function of age. Curr Eye Res. 1985;4:671–678.[ISI][Medline][Order article via Infotrieve]
  17. Wiffen SJ, Nelson LR, Ali AF, Bourne WM. Morphologic assessment of corneal endothelium by specular microscopy in evaluation of donor corneas for transplantation. Cornea. 1995;14:554–561.[ISI][Medline][Order article via Infotrieve]



This article has been cited by other articles:


Home page
IOVSHome page
M. Zhao, G. Thuret, S. Piselli, A. Pipparelli, S. Acquart, M. Peoc'h, J.-M. Dumollard, and P. Gain
Use of Poloxamers for Deswelling of Organ-Cultured Corneas
Invest. Ophthalmol. Vis. Sci., February 1, 2008; 49(2): 550 - 559.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
N. Deb-Joardar, G. Thuret, M. Zhao, S. Acquart, M. Peoc'h, O. Garraud, and P. Gain
Comparison of Two Semiautomated Methods for Evaluating Endothelial Cells of Eye Bank Corneas
Invest. Ophthalmol. Vis. Sci., July 1, 2007; 48(7): 3077 - 3082.
[Abstract] [Full Text] [PDF]


Home page
IOVSHome page
N. Deb-Joardar, G. Thuret, Y. Gavet, S. Acquart, O. Garraud, H. Egelhoffer, and P. Gain
Reproducibility of Endothelial Assessment during Corneal Organ Culture: Comparison of a Computer-Assisted Analyzer with Manual Methods
Invest. Ophthalmol. Vis. Sci., May 1, 2007; 48(5): 2062 - 2067.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Deb-Joardar, N.
Right arrow Articles by Gain, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Deb-Joardar, N.
Right arrow Articles by Gain, P.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS