1From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France; 3Center of MicroNanoTechnology, Ecole Polytechnique Fédérale de Lausanne, Lausanne, Switzerland; 4Laboratory of Signal Treatment and Instrumentation, Université Jean Monnet, Saint-Etienne, France; and 5French Blood Center/Eye Bank of Saint Etienne, Saint Etienne, France.
| Abstract |
|---|
|
|
|---|
METHODS. A standard slide, the Keratotest, was developed with 10 laser-engraved mosaics and different predetermined "cell" densities representing the range of ECDs observed routinely. Horizontal and vertical micrometric scales were etched adjacently to each mosaic, and a standard microscopy resolution test pattern was included. The Keratotest was applied to assess the reliability of a computer-assisted analyzer developed for corneal endothelial evaluation based on light microscopy images.
RESULTS. The Keratotest consisted of 10 microlithographic homogeneous mosaics of 1-mm2 printed area and 1.2-µm cell boundary thickness. The micrometric scale associated with each mosaic aided in simultaneous verification of microscope calibration, and the test pattern aided in checking the microscope resolution. The design was unalterable and reproducible, and the glass slide incorporated in a carbon fiber support ensured easy handling and safe transport. Evaluation of the Keratotest mosaics by the computer-assisted analyzer found a high level of agreement (error margin between +0.12 and 0.46%) with the laser-engraved cell density.
CONCLUSIONS. This prototype device enabled assessment of reliability of ECD measurement in eye banks. It also allowed verification of the calibration and resolution of light microscopes. Periodic validation of counting procedure in eye banks with mosaics of known "cell" densities should be useful for standardization of donor corneal tissue quality control.
In Europe, where organ culture is the storage method of choice, the endothelium is viewed directly under a light microscope after osmotic preparation consisting of incubation of the endothelial face of the cornea for 1 to 4 minutes in 0.9% sodium chloride or 1.8% sucrose.4 This dilates the intercellular spaces and enables the visualization of cell contours that are otherwise invisible. From this preparation, cells are most often counted manually in real time by observation through a calibrated reticule fitted in the microscope eyepiece or by indication of the cells in a known surface on endothelial photographs. Recent years have seen the emergence of prototypes and commercially available computer-assisted counting systems, currently used in some European eye banks,3 in which analysis of digitized optical microscopy images allows rapid and effective measurement of ECD of organ-cultured corneas.5 6 7 8
Previous studies analyzing ECD measurement methods in 22 eye banks in France have emphasized the need for standardization of counting methods.9 10 Lack of microscope calibration had been identified as a major factor responsible for gross variation in ECD estimates between eye banks. Differences in manual counting techniques were also shown to influence the ECD values. Until now, the only method of calibration available was a certified graduated micrometer, a traditional method in microscopy. Essentially, this measures only the magnification ratio between the tissue observed and its image projected in a calibrated reticule or by image analysis software. Whatever the method used, manual or computerized, periodical calibration of the measuring equipment and standardization of counting procedures are essential to ensure reliability of intrabank and interbank cell counts. To address this issue, we present a prototype standard slide, called Keratotest, carrying a series of hexagonal mosaics of different densities mimicking a range of human corneal endothelia.
| Methods |
|---|
|
|
|---|
Manufacturing Process
The Keratotest consisted of a series of laser-engraved regular microlithographic mosaics with predetermined CDs. The steps of design and manufacture are described briefly here.
Calculation of CD
Given that a hexagon is made of six equilateral triangles of side length (L) and height (H), from Pythagorean theory it is known that H =
3/2;L. Therefore, the area of each triangle is given by
LH = 
L2, and the area of a hexagon is 6 x 
L2 = 3
3/2L2. From this, we determined mathematically the number of the hexagons that could be engraved within a given surface measuring 1 mm2 so as to obtain a pattern of 10 mosaics of different CDs. The cell border area was calculated according to the Bourne hypothesis, by which the intercellular space is assumed to be negligible and the true cell border is located at the center of each border line.11 Therefore, the total surface area of each cell was given by its internal surface area plus half the cell border area. Next, from the CD, we calculated the best-fit circle diameter of the hexagon, which enabled us to obtain its vectorial coordinates and to generate the holographic pattern.
Creation of Holographic Pattern
High-quality, two-dimensional drawings of 10 mosaic patterns with adjacent horizontal and vertical micrometer were obtained in DXF format (Drawing eXchange Format; AutoCAD software; Autodesk SA, Neuchatel, Switzerland) containing the holographic data. A simplified microscopy resolution test pattern conforming to the National Bureau of Standards (NBS 1010a Microscopy Test Chart) consisting of three bars of equal spatial frequency in the horizontal and vertical directions was also designed. These were next converted to the machine-specific CIF (Caltech Intermediate File format) format using a translator (CAD; LinkCAD, Burlington, IA).
Photolithography
Photolithographic mosaics were made at the Swiss Federal Institute of MicroNanoTechnology in Lausanne, Switzerland. A glass wafer was coated with a 32-nmthick chromium layer in a vacuum chamber on the principle of evaporation. The chromium layer was covered with a 1.5-µmthick layer of S1813 photoresist (Rohm and Haas Electronic Materials, Coventry, UK) with the use of a spin coater. The holographic pattern was generated by a high-resolution, laser beam direct writing lithography machine (DWL200 Laser Writer; Heidelberg Instruments Mikrotechnik GmbH, Heidelberg, Germany) that uses a Kr laser light source (413-nm wavelength). The target substrate was placed on an XY table with Mach-Zender interferometer position feedback, which permitted high-precision movement with a 50-nm encoding resolution and a 2-µm depth of focus.
Generation of Test Slide
The exposed pattern was transferred to the glass wafer by standard photoresist development and selective etching of exposed chromium. Etched samples were characterized by optical microscopy and scanning electron microscopy. The photoresist, developed by analogy to a photographic film, was baked at 120°C on a hotplate to make it hard and resilient to the etching solution. The substrate was then immersed in an etching solution containing ceric ammonium nitrate. At the end of the process, resin residues were eliminated with a stripping solution. The engraved surface was shielded in a coverslip mounted with resin (Eukitt; ProSciTech, Kirwan, Australia) to ensure mechanical protection. The mount was then cut and incorporated into an unalterable carbon-fenestrated support of dimensions similar to those of a microscope slide (76 x 26 mm).
Example of Application to Computer-Assisted Counting System
We used the Keratotest to assess a tri-image endothelial analyzer (Sambacornea; Samba Technologies, Meylan, France) developed in our laboratory specifically for the evaluation of ECD and morphometry based on light microscopy images.7 The analyzer itself was calibrated with a graduated micrometer of 1-mm certified length (Leica Microsystems, Rueil-Malmaison, France) different from that included in the Keratotest. The 10 mosaics were counted with the tri-image endothelial analyzer five times each by two independent observers. Each time, three random images of a mosaic were captured after displacement and rotation of the Keratotest to ensure that each view was unique. Images were analyzed in a masked fashion to obtain a mean CD for that observation. Each observer selected a different counting zone on each mosaic each time to enable analysis of widely varying CDs. Selections were subjected to automated analysis, as described elsewhere.7 Analysis was performed only by threshold adjustment, and no manual touch-up of contour segments was made. For each mosaic, the (mean ± SD) CD was calculated, and the percentage error was compared with the actual CD of the laser-engraved mosaic. Results of the two analyses were compared.
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Quality assessment of donor tissue and accreditation of medical personnel are important considerations in eye banks worldwide, and the reliability and precision of endothelial cell counting and its periodic standardization are essential prerequisites. In contrast to standardization methods prevalent in other medical disciplines, in which a sample of predetermined value is used for comparison or calibration, no similar tool is available for application to endothelial cell count. A corneal donor tissue sample cannot be used for calibration because no two corneas are alike. We have previously demonstrated that ECD estimation was neither precise nor reproducible in French eye banks.9 10 Inadequate microscope calibration and nonuniform counting procedures (difference in number of cells assessed, number of reticule units counted, and count strategy"border" or "zone") had been identified as the major factors for the high intrabank and interbank variations of ECD. Human error might also have contributed to this variation because a tendency to underestimate was noted in corneas with high ECDs. Based on these findings, we advocated the urgent need for standardization of cell counting methods to prevent the delivery of corneas with low ECD at certain centers and wastage in others.
As a first step in this direction, we designed Alizarin redstained test corneas to evaluate the reproducibility of cell counts between eye banks.10 Endothelial cells were rendered perfectly visible because of membrane staining, thereby eliminating visualization problems encountered during routine light microscopy resulting from nonuniform response of the endothelial monolayer to the osmotic challenge.5 Flat mounting of the dissected endothelium enabled visualization of the entire endothelium in a single plane, thus eliminating field depthrelated measurement errors caused by the combined effects of physiologic corneal curvature and posterior corneal folding. Stained endothelial flat mounts, though representative of the real human endothelium, had certain limitations: (1) the ECD could not be predetermined mathematically, and (2) the slides were not reproducible because each human endothelium is unique. Hence, manufacture of a series to facilitate interbank exchange was not possible. A different standard that would represent the real endothelium but would be mathematically accurate and reproducible and would be useful for training new technicians was required to assess the reliability of eye bank counting methods.
The importance of calibrating specular microscopes for in vivo ECD counts has been emphasized in previous studies.12 13 Different analysis techniques of specular microscopy imagesi.e., use of a fixed-frame or a center methodled to important interbank ECD variation in the Cornea Donor Study.14 Light microscopic evaluation used for organ-cultured corneas also has similar prerequisites. Computer-assisted analyzers also require calibration of the optical and the computing devices. In the tri-image endothelial analyzer, this is achieved by use of a calibration factor derived from a standard micrometric slide. However, verification of counting procedures or detection of human errors is not spontaneously possible. Our photolithographic device was designed for simultaneous calibration of the microscope and the counting device (reticule, printed grid, or digital image analyzer) and could be adapted to manual and computer-assisted cell counts. The extremely narrow limits of difference between the mathematically determined Keratotest ECD and that obtained after repeated counts by different observers with the computer-assisted analyzer permitted assessment of the reliability of counting methods.
Patterning of flat surfaces by use of photolithography is a high-precision technology widely used in industrial applications, such as in the manufacture of integrated circuits and microsystems.15 Choice of this particular method guaranteed the etching precision and hence the accuracy of the CDs obtained. Mathematical predetermination allowed creation of the exact CD desired, thus permitting a large range of ECDs (8003200 cells/mm2) likely to be encountered in eye banks. The manufacturing process made it possible to create reproducible slides on a semi-industrial scale to meet the requirements of potential users.
The current prototype allows only validation of the ECD count. Further modifications with mosaics having "cells" of different surface area and shape are ongoing to enable assessment of morphometric parameters increasingly recognized as valuable indicators of donor tissue quality.16 17
| Acknowledgements |
|---|
| Footnotes |
|---|
Supported by Grant Recherche et greffes 2004 from lEtablissement français des Greffes.
Submitted for publication May 15, 2006; revised June 5, 2006; accepted August 23, 2006.
Disclosure: N. Deb-Joardar, None; G. Thuret, None; G.-A. Racine, None; D. Pons, None; G. Brun, None; O. Parriaux, None; M. Peoch, None; S. Acquart, None; P. Gain, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Philippe Gain, Service dOphtalmologie (pavillon 50A), CHRU de Bellevue, 25 Boulevard Pasteur, F 42055 Saint Etienne Cedex 2, France; philippe.gain{at}univ-st-etienne.fr.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Zhao, G. Thuret, S. Piselli, A. Pipparelli, S. Acquart, M. Peoc'h, J.-M. Dumollard, and P. Gain Use of Poloxamers for Deswelling of Organ-Cultured Corneas Invest. Ophthalmol. Vis. Sci., February 1, 2008; 49(2): 550 - 559. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Deb-Joardar, G. Thuret, M. Zhao, S. Acquart, M. Peoc'h, O. Garraud, and P. Gain Comparison of Two Semiautomated Methods for Evaluating Endothelial Cells of Eye Bank Corneas Invest. Ophthalmol. Vis. Sci., July 1, 2007; 48(7): 3077 - 3082. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Deb-Joardar, G. Thuret, Y. Gavet, S. Acquart, O. Garraud, H. Egelhoffer, and P. Gain Reproducibility of Endothelial Assessment during Corneal Organ Culture: Comparison of a Computer-Assisted Analyzer with Manual Methods Invest. Ophthalmol. Vis. Sci., May 1, 2007; 48(5): 2062 - 2067. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||