(Investigative Ophthalmology and Visual Science. 2006;47:5011-5016.)
© 2006 by The Association for Research in Vision and Ophthalmology, Inc.
DOI: 10.1167/iovs.06-0039
Automatic Retinal Oximetry
Sveinn Hakon Hardarson,1
Alon Harris,2
Robert Arnar Karlsson,3
Gisli Hreinn Halldorsson,3
Larry Kagemann,2,4
Ehud Rechtman,2
Gunnar Már Zoega,1
Thor Eysteinsson,1
Jon Atli Benediktsson,3
Adalbjorn Thorsteinsson,5
Peter Koch Jensen,6
James Beach,7 and
Einar Stefánsson1
1From the Departments of Ophthalmology and
5Anesthesiology, University of Iceland, National University Hospital, Reykjavík, Iceland; the
2Department of Ophthalmology, Indiana University/Purdue University School of Medicine, Indianapolis, Indiana; the
3Faculty of Electrical and Computer Engineering, University of Iceland, Reykjavik; the
4Department of Ophthalmology, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania; the
6Eye Department, National University Hospital of Copenhagen, Denmark; and the
7Institute for Technology Development, Stennis Space Center, Mississippi.
 |
Abstract
|
|---|
PURPOSE. To measure hemoglobin oxygen saturation (SO2) in retinal vessels and to test the reproducibility and sensitivity of an automatic spectrophotometric oximeter.
METHODS. Specialized software automatically identifies the retinal blood vessels on fundus images, which are obtained with four different wavelengths of light. The software calculates optical density ratios (ODRs) for each vessel. The reproducibility was evaluated by analyzing five repeated measurements of the same vessels. A linear relationship between SO2 and ODR was assumed and a linear model derived. After calibration, reproducibility and sensitivity were calculated in terms of SO2. Systemic hyperoxia (n = 16) was induced in healthy volunteers by changing the O2 concentration in inhaled air from 21% to 100%.
RESULTS. The automatic software enhanced reproducibility, and the mean SD for repeated measurements was 3.7% for arterioles and 5.3% venules, in terms of percentage of SO2 (five repeats, 10 individuals). The model derived for calibration was SO2 = 125 142 · ODR. The arterial SO2 measured 96% ± 9% (mean ± SD) during normoxia and 101% ± 8% during hyperoxia (n = 16). The difference between normoxia and hyperoxia was significant (P = 0.0027, paired t-test). Corresponding numbers for venules were 55% ± 14% and 78% ± 15% (P < 0.0001). SO2 is displayed as a pseudocolor map drawn on fundus images.
CONCLUSIONS. The retinal oximeter is reliable, easy to use, and sensitive to changes in SO2 when concentration of O2 in inhaled air is changed.
A noninvasive device to measure retinal oxygen levels is needed for the study of the role of oxygen levels in human retinopathy and glaucoma. Spectrophotometric measurements of hemoglobin oxygen saturation in human retinal blood vessels was initiated by Hickam et al.1 in the 1960s but has still not been mastered as a reliable and useful clinical or research tool. During the past 7 years, we have built on the work of Beach et al.,2 Delori,3 and others4 to develop a noninvasive automatic retinal oximeter, which we hope will allow studies of human retinal and optic nerve disease in research and clinical settings. We report the sensitivity and reproducibility of this device in human studies.
Michaelson (1948)5 suggested that hypoxia plays a role in the pathophysiology of diabetic and other retinopathies. Animal research and limited studies on humans suggest that oxygen may be an important factor in the pathogenesis and treatment of some retinal diseases. These include venular6 7 8 9 10 11 (Scibor M et al. IOVS 2002;43:ARVO E-Abstract 3305) and arteriolar12 13 occlusions, retinal detachment,14 15 16 17 18 19 and diabetic retinopathy20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 (Trick GL et al. IOVS 2005;46:ARVO E-abstract 1413). Animal studies have also indicated that both vitrectomy7 36 37 38 and laser photocoagulation8 9 28 36 38 39 40 41 42 43 44 45 raise the oxygen tension in the retina.
Although glaucoma therapy is currently aimed at lowering intraocular pressure, the beneficial effects may be due to better perfusion and oxygenation of the optic nerve rather than the direct effects of lower intraocular pressure. Experiments in monkeys,46 47 cats,48 49 rabbits,50 and pigs51 52 have demonstrated that optic nerve oxygenation decreases when the perfusion pressure declines and autoregulation of blood flow is overwhelmed (e.g., because of raised intraocular pressure). Poor regulation of blood flow has been postulated as the mechanism of glaucomatous optic atrophy.53 54
These theories are predominantly based on animal research, and few studies have been conducted to investigate oxygenation in humans, because reliable and noninvasive methods of measuring human retinal oxygenation have been lacking. In the past decades, various approaches to retinal oximetry have been tried (for a review of the methods, see Harris et al.4 ). Most of these methods are based on the fact that oxygenated and deoxygenated hemoglobin have different light absorption spectra (color). By analyzing the light absorbance of blood at two or more wavelengths, the oxygenation of hemoglobin can be estimated. Investigators in this field have run into numerous problems, however, that have prevented the development of a reliable and practical method. The problems include nonlinear sensors (photographic film), eye movement during lengthy measurements, small areas measured at one time, and optical complexities within the eye. Nevertheless, considerable progress has been made.2 3 55 56
In this study we compared the reproducibility of our manual and automatic methods for retinal oximetry. We present calibration for our automatic method and use this to display the reproducibility and sensitivity of our retinal oximeter to hemoglobin oxygen saturation.
 |
Methods
|
|---|
All procedures were approved by the institutional review boards of the respective institutions: Indiana University-Purdue University, Indianapolis, and the University of Iceland, National University Hospital, Reykjavík. Informed consent was provided by all subjects before participation in the studies. All procedures conformed to the tenets of the Declaration of Helsinki.
The Retinal Oximeter
Our method for retinal oximetry is based on the work of Beach et al.2 The setup and outcome are shown in Figure 1 .

View larger version (35K):
[in this window]
[in a new window]
|
FIGURE 1. (A) The retinal oximeter. A traditional fundus camera, coupled with a beam splitter and a digital camera is used to obtain four images simultaneously. (B) Each of the four images is obtained at a different wavelength of light.
|
|
Two retinal oximeters were used for the study, one based in Indianapolis and the other in Reykjavik. Both are composed of a fundus camera coupled with a beam splitter to a digital camera. (In Indianapolis, the fundus camera was a Topcon TRC50-VT; Topcon Co., Tokyo, Japan; the beam splitter was a MultiSpec Patho-Imager; Optical Insights, Tucson, AZ; and the digital camera was a CoolSNAP ES; Roper Scientific, Tucson, AZ. In Reykjavik, the same beam splitter was used, but the fundus camera was a Canon CR6-45NM; Canon Inc., Tokyo, Japan; the digital camera was an SBIG ST-7E; Santa Barbara Instrument Group, Santa Barbara, CA, used at 2 x 2 binning). The beam splitter separates the original image into four optical channels. In each channel, there is a different narrow band-pass filter, through which only light of specific wavelengths can pass. The center wavelengths of the filters are 542, 558, 586, and 605 nm and the half-bandwidth is 5 nm, except for the 542-nm filter, which has a 9-nm half-bandwidth. In the present study, only the 586- and 605-nm wavelengths were used.
With a specialized computer program, optical density (OD) can be calculated for every vessel at each of the four wavelengths. OD is a measure of the bloods light absorbance and is calculated as
 | (1) |
where I and Io are the brightness levels inside and just outside a vessel, respectively. It can be shown that the ratio of ODs at certain wavelengths (OD ratio, ODR) has an inverse and approximately linear relationship to hemoglobin oxygenation2 4 :
 | (2) |
In equation 2 , SO2 is the percentage of hemoglobin oxygen saturation; a and k are constants; ODX and ODY are ODs (no unit) at wavelengths X and Y, respectively; and ODR is the optical density ratio. Thus, in theory, hemoglobin oxygenation can be calculated using brightness inside and outside vessels at two wavelengths of light.
The Software
Two versions of software were used. Both versions register all four output images in the same coordinate system, so that each pair of points, used for calculation of ODs, is at the same fundus location for all four images. The registration algorithm utilizes the edges of the four images.57 The older software version requires the user to select measurement points manually, inside and outside vessels to calculate the ODR. The newer version automatically locates vessels on the fundus image and selects measurement points inside and outside vessels. Morphologic filters are used to enhance vessellike structures in the images. After morphologic enhancement, a skeleton of a thresholded fundus image is used to locate the approximate centerlines of the arterioles and venules. To avoid error caused by specular reflection, the new software uses the darkest point on the vessels cross section. To reduce variation in calculated ODRs due to random system noise and variable optical properties of the fundus, the software averages measurements from several pixels adjacent to the vessel. When the user has specified the vessel segment of interest, the mean ODR for that segment is given. The automatic software also displays a color-coded map of the vessels, where the colors indicate hemoglobin oxygen saturation (after calibration, described later).
Reproducibility
In Reykjavik, five images (each composed of four spectral images) were taken of one eye of each of 11 individuals. One individual was excluded from analysis because of poor-quality images. The images used were of diabetics, with (n = 2) and without (n = 3) retinopathy, and healthy volunteers (n = 5). The reproducibility was very similar in diabetics and healthy individuals. One first-degree arteriole and one first-degree venule were chosen for analysis in each eye. Each vessel was analyzed with both software versions and the SD was calculated between the five images. When the manual software was used, four pairs of measurement points were averaged for each vessel.
Sensitivity
The sensitivity of the system was tested by comparing results obtained during inhalation of different concentrations of O2. In Reykjavik, five healthy volunteers inhaled either 21% (room air) or 100% O2. In Indianapolis, 15 healthy volunteers inhaled either room air or 100% O2. Finger pulse oximeters were used to monitor arterial oxygenation (Medical/GE Healthcare, Milwaukee WI in Reykjavík, and POET II model 602-3; Criticare Systems, Waukesha, WI in Indianapolis). During normoxia, the finger oximetry values were 97% to 99% but 98% to 100% during hyperoxia. Measurements were made on one first-degree arteriole and one first-degree venule in each eye. Individuals were excluded from further analysis if their images did not contain the same vessels during inhalation of different concentrations of oxygen (one individual in Reykjavik and three in Indianapolis). A paired t-test was used for statistical analysis.
 |
Results
|
|---|
Reproducibility
Table 1 shows the mean and SD of the five repeated measurements of the same vessel, using either the manual or the automatic software. All other results (below) were generated with the automatic software.
Calibration
Earlier measurements on healthy volunteers (for review, see Harris et al.4 ) have yielded retinal arteriolar oxygenation of 92%,58 97%,59 and 98%3 and venular oxygenation of 45%,3 55%,2 58%,58 and 58%.1 The mean of these results is 96% for arterioles and 54% for venules. From our data, we calculated a mean ODR of 0.209 for arterioles and 0.502 for venules (18 healthy individuals, one first-degree vessel pair in each). Assuming a linear relationship between ODR and SO2 and using the above means from the literature for SO2 and our mean ODR gives the following:
 | (3) |
and
 | (4) |
Solving equations 3 and 4 together gives a = 125 and k = 142, and so the equation for SO2 is
 | (5) |
Reproducibility of Calibrated Results
Table 2 shows the reproducibility of the automatic measurements in Table 1 after calibration with equation 5 .
Oxygen Sensitivity of the Automatic Approach
Four healthy volunteers in Reykjavik and 12 in Indianapolis inhaled either room air or 100% oxygen. Table 3 shows the results of analysis with the automatic software, using equation 5 .
There is a significant difference between normoxia and hyperoxia, both in arterioles and venules. The mean difference in arterioles is 5% in terms of %SO2 (P = 0.0027, 95% confidence interval [CI] 2%8%, pooled data, n = 16, paired t-test). The mean difference for venules is 24% (P < 0.0001, 95% CI: 16%32%, n = 16).
Pseudocolor Presentation
When calibrated, the automatic software can display the results as a pseudocolor fundus image as shown in Figure 2 . Different colors represent different levels of oxygenation in retinal vessels.

View larger version (56K):
[in this window]
[in a new window]
|
FIGURE 2. A screen image produced by the automatic software from a diabetic patient with no retinopathy. The color code indicates the oxygenation in the vessels.
|
|
 |
Discussion
|
|---|
The automatic software yields better reproducibility than does the older manual software. The system is sensitive to changes in hemoglobin oxygen saturation in both arterioles and venules when oxygen concentration in the inhaled air is changed. Furthermore, the system is easy to use and can display results as %SO2, in numeric or graphic format.
Table 1 shows that the standard deviation for repeated measurements is much lower when the automatic software is used but the average ODR values are similar between manual and automatic analysis. When using the automatic software, the standard deviation (in terms of %SO2) for repeated measurements is 3.7% and 5.3% for arterioles and venules, respectively. The mean saturation for this dataset was 99% in arterioles and 52% in venules. Comparing these means with prior results from researchers using other retinal oximeters is of limited value, because we used these prior means to calibrate our oximeter (see the Results section). The reproducibility can, however, be compared with prior results, although this comparison is somewhat limited by the different methodologies applied. Delori3 used an oximeter that scanned a small region of the fundus with three wavelengths of light. The mean standard deviation between two consecutive measurements of the same vessel was 2.1% SO2, similar for arterioles and venules (means, 98% and 45%, respectively). As 1.4 seconds were needed to achieve a single measurement, a vessel-tracking system was needed to compensate for eye movement. Using an oximeter based on scanning laser ophthalmoscopy, Smith et al.60 reported that in preliminary measurements on humans at a single retinal position, the measurement repeatability was typically within ±5%SO2 (venular SO2, typically 45%55%). The same oximeter is reported to complete a scan in one-thirtieth of a second,61 and a typical scan appears to involve two adjacent vessels.60 Schweitzer et al.58 used an oximeter that scans a 1.5-mm by 40-µm slit of the fundus (one or two adjacent vessels) and measures the reflectance from the vessel at wavelengths from 450 to 700 nm in 2-nm increments. Nineteen measurements on each vessel gave an SD of 4.6% SO2 for arterioles and 4% SO2 for venules (means, 92% and 58%, respectively). Therefore, the reproducibility of our oximeter is comparable to that in earlier reports.
The results also demonstrate the sensitivity of the oximeter. The measured arteriolar hemoglobin oxygen saturation follows changes in inspired oxygen concentration. In hyperoxia the increase in hemoglobin saturation is limited by the fact that hemoglobin is almost 100% saturated in normoxia. In hyperoxia, oxygen that is dissolved in choroidal blood diffuses into the retina in larger amounts22 62 63 64 and reduces the extraction of oxygen from the retinal circulation, thus elevating the venular oxygen saturation, as our results show.
The retinal oximeters mentioned herein have also been shown to be sensitive to changes in oxygen saturation. This sensitivity is essential, of course, if an oximeter is to be of value for measuring changes associated with disease or treatment. Whether the oximeters show the true oxygen saturation is another matter that is difficult to resolve because absolute values are very difficult to obtain in humans. Measurements on several mammalian species with vascular retinas revealed that oxygen tension (PO2) of the retina and optic nerve is around 20 mm Hg when this is measured with an electrode at the vitreal surface of the retina and optic disc.7 65 66 67 Assuming normal blood parameters, our venular SO2 converts to PO2 of approximately 29 mm Hg.68 As could be expected, our venular PO2 value is a little higher than the tissue PO2 from the animal experiments (the PO2 gradient should be toward the tissue and away from the arterioles, capillaries, and probably venules).
We have room to improve both sensitivity and reproducibility, because we have not yet incorporated correction for confounding factors such as light-scattering of blood cells,3 different light paths through vessels,69 vessel width,2 and fundus pigmentation,2 60 all of which have been shown to affect retinal oximetry to some extent. Some of the corrections require more than two wavelengths of light. Our system already has the capability to use four wavelengths simultaneously although the results presented were generated with only two. We have found better reproducibility with data based on 586 and 605 nm, compared with those based on 542 and 558 nm.
In summary, we have shown that the automatic software improves the reproducibility of our oximeter and that the oximeter is sensitive to changes in hemoglobin oxygenation. The advantages of our retinal oximeter are that the measurement area is large, and little expertise is needed to acquire a pseudocolor fundus image of retinal vessel oxygenation. We hope this device will be useful in the evaluation of ischemic retinal diseases and their treatment.
 |
Footnotes
|
|---|
Supported by Icelandic Research Council Grants 050409011 and 051220005, The LandspítaliUniversity Hospital Fund for Science, The Research Fund of The University of Iceland, and The Icelandic Fund for Prevention of Blindness.
Submitted for publication January 13, 2006; revised May 9 and July 19, 2006; accepted September 18, 2006.
Disclosure: S.H. Hardarson, Oxymap (E, P); A. Harris, None; R.A. Karlsson, Oxymap (E, P); G.H. Halldorsson, Oxymap (E, P); L. Kagemann, None; E. Rechtman, None; G.M. Zoega, None; T. Eysteinsson, Oxymap (I, P); J.A. Benediktsson, Oxymap (I, P); A. Thorsteinsson, None; P.K. Jensen, None; J. Beach, Oxymap (I, P); E. Stefánsson, Oxymap (I, P)
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Einar Stefánsson, Department of Ophthalmology, University of Iceland, National University Hospital, Reykjavík, Iceland; einarste{at}landspitali.is.
 |
References
|
|---|
- Hickam JB, Frayser R, Ross JC. A study of retinal venous blood oxygen saturation in human subjects by photographic means. Circulation. 1963;27:375385.[Free Full Text]
- Beach JM, Schwenzer KJ, Srinivas S, Kim D, Tiedeman JS. Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation. J Appl Physiol. 1999;86:748758.[Abstract/Free Full Text]
- Delori FC. Noninvasive techniques for oximetry of blood in retinal vessels. Appl Opt. 1988;27:11131125.
- Harris A, Dinn RB, Kagemann L, Rechtman E. A review of methods for human retinal oximetry. Ophthalmic Surg Lasers Imaging. 2003;34:152164.[ISI][Medline][Order article via Infotrieve]
- Michaelson IC. The mode of development of the vascular system of the retina with some observations on its significance for certain retinal diseases. Trans Ophthalmol Soc UK. 1948;68:137180.
- Pournaras CJ, Tsacopoulos M, Strommer K, Gilodi N, Leuenberger PM. Experimental retinal branch vein occlusion in miniature pigs induces local tissue hypoxia and vasoproliferative microangiopathy. Ophthalmology. 1990;97:13211328.[ISI][Medline][Order article via Infotrieve]
- Stefansson E, Novack RL, Hatchell DL. Vitrectomy prevents retinal hypoxia in branch retinal vein occlusion. Invest Ophthalmol Vis Sci. 1990;31:284289.[Abstract/Free Full Text]
- Pournaras CJ, Tsacopoulos M, Strommer K, Gilodi N, Leuenberger PM. Scatter photocoagulation restores tissue hypoxia in experimental vasoproliferative microangiopathy in miniature pigs. Ophthalmology. 1990;97:13291333.[ISI][Medline][Order article via Infotrieve]
- Pournaras CJ. Retinal oxygen distribution. Its role in the physiopathology of vasoproliferative microangiopathies. Retina. 1995;15:332347.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Yoneya S, Saito T, Nishiyama Y, et al. Retinal oxygen saturation levels in patients with central retinal vein occlusion. Ophthalmology. 2002;109:15211526.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Pournaras JA, Petropoulos IK, Munoz JL, Pournaras CJ. Experimental retinal vein occlusion: effect of acetazolamide and carbogen (95% O2/5% CO2) on preretinal PO2. Invest Ophthalmol Vis Sci. 2004;45:36693677.[Abstract/Free Full Text]
- Alder VA, Ben-Nun J, Cringle SJ. PO2 profiles and oxygen consumption in cat retina with an occluded retinal circulation. Invest Ophthalmol Vis Sci. 1990;31:10291034.[Abstract/Free Full Text]
- Braun RD, Linsenmeier RA. Retinal oxygen tension and the electroretinogram during arterial occlusion in the cat. Invest Ophthalmol Vis Sci. 1995;36:523541.[Abstract/Free Full Text]
- Lewis G, Mervin K, Valter K, et al. Limiting the proliferation and reactivity of retinal Muller cells during experimental retinal detachment: the value of oxygen supplementation. Am J Ophthalmol. 1999;128:165172.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Stefansson E. Ocular Oxygenation and Neovascularization. 1981; Duke University Durham, NC.
- Mervin K, Valter K, Maslim J, et al. Limiting photoreceptor death and deconstruction during experimental retinal detachment: the value of oxygen supplementation. Am J Ophthalmol. 1999;128:155164.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Linsenmeier RA, Padnick-Silver L. Metabolic dependence of photoreceptors on the choroid in the normal and detached retina. Invest Ophthalmol Vis Sci. 2000;41:31173123.[Abstract/Free Full Text]
- Lewis GP, Talaga KC, Linberg KA, Avery RL, Fisher SK. The efficacy of delayed oxygen therapy in the treatment of experimental retinal detachment. Am J Ophthalmol. 2004;137:10851095.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Sakai T, Lewis GP, Linberg KA, Fisher SK. The ability of hyperoxia to limit the effects of experimental detachment in cone-dominated retina. Invest Ophthalmol Vis Sci. 2001;42:32643273.[Abstract/Free Full Text]
- Dean FM, Arden GB, Dornhorst A. Partial reversal of protan and tritan colour defects with inhaled oxygen in insulin dependent diabetic subjects. Br J Ophthalmol. 1997;81:2730.[Abstract/Free Full Text]
- Drasdo N, Chiti Z, Owens DR, North RV. Effect of darkness on inner retinal hypoxia in diabetes. Lancet. 2002;359:22512253.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Wangsa-Wirawan ND, Linsenmeier RA. Retinal oxygen: fundamental and clinical aspects. Arch Ophthalmol. 2003;121:547557.[Abstract/Free Full Text]
- Tiedeman JS, Kirk SE, Srinivas S, Beach JM. Retinal oxygen consumption during hyperglycemia in patients with diabetes without retinopathy. Ophthalmology. 1998;105:3136.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Ernest JT, Goldstick TK, Engerman RL. Hyperglycemia impairs retinal oxygen autoregulation in normal and diabetic dogs. Invest Ophthalmol Vis Sci. 1983;24:985989.[Abstract/Free Full Text]
- Isenberg SJ, McRee WE, Jedrzynski MS. Conjunctival hypoxia in diabetes mellitus. Invest Ophthalmol Vis Sci. 1986;27:15121515.[Abstract/Free Full Text]
- Sutherland FS, Stefansson E, Hatchell DL, Reiser H. Retinal oxygen consumption in vitro; the effect of diabetes mellitus, oxygen and glucose. Acta Ophthalmol (Copenh). 1990;68:715720.[Medline][Order article via Infotrieve]
- Alder VA, Yu DY, Cringle SJ, Su EN. Changes in vitreal oxygen tension distribution in the streptozotocin diabetic rat. Diabetologia. 1991;34:469476.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Augsten R, Konigsdorffer E, Schweitzer D, Strobel J. Multisubstance analysis of reflection spectra before and after laser photocoagulation for proliferative diabetic retinopathy. Eur J Ophthalmol. 1997;7:317321.[Medline][Order article via Infotrieve]
- Nguyen QD, Shah SM, Van Anden E, et al. Supplemental oxygen improves diabetic macular edema: a pilot study. Invest Ophthalmol Vis Sci. 2004;45:617624.[Abstract/Free Full Text]
- Berkowitz BA, Ito Y, Kern TS, McDonald C, Hawkins R. Correction of early subnormal superior hemiretinal DeltaPO(2) predicts therapeutic efficacy in experimental diabetic retinopathy. Invest Ophthalmol Vis Sci. 2001;42:29642969.[Abstract/Free Full Text]
- Berkowitz BA, Kowluru RA, Frank RN, et al. Subnormal retinal oxygenation response precedes diabetic-like retinopathy. Invest Ophthalmol Vis Sci. 1999;40:21002105.[Abstract/Free Full Text]
- Berkowitz BA, Luan H, Gupta RR, et al. Regulation of the early subnormal retinal oxygenation response in experimental diabetes by inducible nitric oxide synthase. Diabetes. 2004;53:173178.[Abstract/Free Full Text]
- Luan H, Leitges M, Gupta RR, et al. Effect of PKCbeta on retinal oxygenation response in experimental diabetes. Invest Ophthalmol Vis Sci. 2004;45:937942.[Abstract/Free Full Text]
- Trick GL, Berkowitz BA. Retinal oxygenation response and retinopathy. Prog Retin Eye Res. 2005;24:259274.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Berkowitz BA, Roberts R, Luan H, et al. Drug intervention can correct subnormal retinal oxygenation response in experimental diabetic retinopathy. Invest Ophthalmol Vis Sci. 2005;46:29542960.[Abstract/Free Full Text]
- Stefansson E, Landers MB, 3rd, Wolbarsht ML. Increased retinal oxygen supply following pan-retinal photocoagulation and vitrectomy and lensectomy. Trans Am Ophthalmol Soc. 1981;79:307334.[Medline][Order article via Infotrieve]
- Stefansson E. Oxygen and diabetic eye disease. Graefes Arch Clin Exp Ophthalmol. 1990;228:120123.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Stefansson E. The therapeutic effects of retinal laser treatment and vitrectomy; a theory based on oxygen and vascular physiology. Acta Ophthalmol Scand. 2001;79:435440.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Landers MB, 3rd, Stefansson E, Wolbarsht ML. Panretinal photocoagulation and retinal oxygenation. Retina. 1982;2:167175.[ISI][Medline][Order article via Infotrieve]
- Molnar I, Poitry S, Tsacopoulos M, Gilodi N, Leuenberger PM. Effect of laser photocoagulation on oxygenation of the retina in miniature pigs. Invest Ophthalmol Vis Sci. 1985;26:14101414.[Abstract/Free Full Text]
- Stefansson E, Hatchell DL, Fisher BL, Sutherland FS, Machemer R. Panretinal photocoagulation and retinal oxygenation in normal and diabetic cats. Am J Ophthalmol. 1986;101:657664.[ISI][Medline][Order article via Infotrieve]
- Novack RL, Stefansson E, Hatchell DL. The effect of photocoagulation on the oxygenation and ultrastructure of avascular retina. Exp Eye Res. 1990;50:289296.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Stefansson E, Machemer R, de Juan E, Jr, McCuen BW, 2nd, Peterson J. Retinal oxygenation and laser treatment in patients with diabetic retinopathy. Am J Ophthalmol. 1992;113:3638.[ISI][Medline][Order article via Infotrieve]
- Funatsu H, Wilson CA, Berkowitz BA, Sonkin PL. A comparative study of the effects of argon and diode laser photocoagulation on retinal oxygenation. Graefes Arch Clin Exp Ophthalmol. 1997;235:168175.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Yu DY, Cringle SJ, Su E, et al. Laser-induced changes in intraretinal oxygen distribution in pigmented rabbits. Invest Ophthalmol Vis Sci. 2005;46:988999.[Abstract/Free Full Text]
- Armaly MF, Araki M. Optic nerve circulation and ocular pressure. Invest Ophthalmol. 1975;14:724731.[Abstract/Free Full Text]
- Khoobehi B, Beach JM, Kawano H. Hyperspectral imaging for measurement of oxygen saturation in the optic nerve head. Invest Ophthalmol Vis Sci. 2004;45:14641472.[Abstract/Free Full Text]
- Novack RL, Stefansson E, Hatchell DL. Intraocular pressure effects on optic nerve-head oxidative metabolism measured in vivo. Graefes Arch Clin Exp Ophthalmol. 1990;228:128133.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Shonat RD, Wilson DF, Riva CE, Cranstoun SD. Effect of acute increases in intraocular pressure on intravascular optic nerve head oxygen tension in cats. Invest Ophthalmol Vis Sci. 1992;33:31743180.[Abstract/Free Full Text]
- Selbach MJ, Wonka F, Hoper J, Funk RH. Effects of elevated intraocular pressure on haemoglobin oxygenation in the rabbit optic nerve head: a microendoscopical study. Exp Eye Res. 1999;69:301309.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Blumenroder S, Augustin AJ, Koch FH. The influence of intraocular pressure and systemic oxygen tension on the intravascular pO2 of the pig retina as measured with phosphorescence imaging. Surv Ophthalmol. 1997;42(Suppl 1)S118S126.[Medline][Order article via Infotrieve]
- la Cour M, Kiilgaard JF, Eysteinsson T, et al. Optic nerve oxygen tension: effects of intraocular pressure and dorzolamide. Br J Ophthalmol. 2000;84:10451049.[Abstract/Free Full Text]
- Flammer J, Orgul S, Costa VP, et al. The impact of ocular blood flow in glaucoma. Prog Retin Eye Res. 2002;21:359393.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Flammer J, Haefliger IO, Orgul S, Resink T. Vascular dysregulation: a principal risk factor for glaucomatous damage?. J Glaucoma. 1999;8:212219.[ISI][Medline][Order article via Infotrieve]
- Schweitzer D, Thamm E, Hammer M, Kraft J. A new method for the measurement of oxygen saturation at the human ocular fundus. Int Ophthalmol. 2001;23:347353.[CrossRef][Medline][Order article via Infotrieve]
- Denninghoff KR, Smith MH, Lompado A, Hillman LW. Retinal venous oxygen saturation and cardiac output during controlled hemorrhage and resuscitation. J Appl Physiol. 2003;94:891896.[Abstract/Free Full Text]
- Halldorsson GH, Benediktsson JA, Zoega GM, Stefansson E, Eysteinsson T. Automatic registration of spectrophotometric retinal images. IEEE Nordic Signal Processing Workshop (Norsig 2004). 2004.58.
- Schweitzer D, Hammer M, Kraft J, et al. In vivo measurement of the oxygen saturation of retinal vessels in healthy volunteers. IEEE Trans Biomed Eng. 1999;46:14541465.[CrossRef][ISI][Medline][Order article via Infotrieve]
- de Kock JP, Tarassenko L, Glynn CJ, Hill AR. Reflectance pulse oximetry measurements from the retinal fundus. IEEE Trans Biomed Eng. 1993;40:817823.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Smith MH, Denninghoff KR, Lompado A, Woodruff JB, Hillman LW. Minimizing the influence of fundus pigmentation on retinal vessel oximetry measurements. Proc SPIE. 2001;4245:135145.[CrossRef]
- Denninghoff KR, Smith MH, Hillman L. Retinal imaging techniques in diabetes. Diabetes Technol Ther. 2000;2:111113.[CrossRef][Medline][Order article via Infotrieve]
- Wolbarsht ML, Stefansson E, Landers MB, 3rd. Retinal oxygenation from the choroid in hyperoxia. Exp Biol. 1987;47:4952.[ISI][Medline][Order article via Infotrieve]
- Linsenmeier RA, Yancey CM. Effects of hyperoxia on the oxygen distribution in the intact cat retina. Invest Ophthalmol Vis Sci. 1989;30:612618.[Abstract/Free Full Text]
- Pournaras CJ, Riva CE, Tsacopoulos M, Strommer K. Diffusion of O2 in the retina of anesthetized miniature pigs in normoxia and hyperoxia. Exp Eye Res. 1989;49:347360.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Ahmed J, Braun RD, Dunn R, Jr, Linsenmeier RA. Oxygen distribution in the macaque retina. Invest Ophthalmol Vis Sci. 1993;34:516521.[Abstract/Free Full Text]
- Stefansson E, Jensen PK, Eysteinsson T, et al. Optic nerve oxygen tension in pigs and the effect of carbonic anhydrase inhibitors. Invest Ophthalmol Vis Sci. 1999;40:27562761.[Abstract/Free Full Text]
- Yu DY, Cringle SJ. Oxygen distribution and consumption within the retina in vascularised and avascular retinas and in animal models of retinal disease. Prog Retin Eye Res. 2001;20:175208.[CrossRef][ISI][Medline][Order article via Infotrieve]
- Kelman GR. Digital computer subroutine for the conversion of oxygen tension into saturation. J Appl Physiol. 1966;21:13751376.[Free Full Text]
- Smith MH, Denninghoff KR, Lompado A, Hillman LW. Effect of multiple light paths on retinal vessel oximetry. Appl Opt. 2000;39:11831193.[Medline][Order article via Infotrieve]