1From the Save Sight Institute, University of Sydney; and the 2Sydney Eye Hospital, Sydney, Australia.
| Abstract |
|---|
|
|
|---|
METHODS. Twenty-eight patients with glaucoma and 30 healthy subjects underwent mfVEP on monocular and dichoptic stimulation. Dichoptic stimulation was presented with the use of virtual reality goggles (recording time, 7 minutes). Monocular mfVEPs were recorded sequentially for each eye (recording time, 10 minutes).
RESULTS. Comparison of mean relative asymmetry coefficient (RAC; calculated as difference in amplitudes between eyes/sum of amplitudes of both eyes at each segment) on monocular and dichoptic mfVEP revealed significantly lower RAC on dichoptic (0.003 ± 0.03) compared with monocular testing (–0.02 ± 0.04; P = 0.002). In all 28 patients, dichoptic mfVEP identified defects with excellent topographic correspondence. Of 56 hemifields (28 eyes), 33 had Humphrey visual field (HFA) scotomas, all of which were detected by dichoptic mfVEP. Among 23 hemifields with normal HFA, two were abnormal on monocular and dichoptic mfVEP. Five hemifields (five patients) normal on HFA and monocular mfVEP were abnormal on dichoptic mfVEP. In all five patients, corresponding rim changes were observed on disc photographs. Mean RAC of glaucomatous eyes was significantly higher on dichoptic (0.283 ± 0.18) compared with monocular (0.199 ± 0.12) tests (P = 0.0006).
CONCLUSIONS. Dichoptic mfVEP not only detects HFA losses, it may identify early defects in areas unaffected on HFA and monocular mfVEP while reducing testing time by 30%. Asymmetry was tighter among healthy subjects but wider in patients with glaucoma on simultaneous binocular stimulation, which is potentially a new tool in the early detection of glaucoma.
Multifocal visual evoked potential (mfVEP) recording is a sensitive objective perimetric technique for the detection of glaucomatous visual field loss; reported sensitivity is 80% to 95% in early glaucoma with the conventional black-white pattern-reversal stimulus.2 3 4 Since it was first described in 1994,5 several advances have made the technique less variable and more clinically applicable. Four-channel recording made it possible to reliably record VEPs from multiple stimulated areas of the visual field.6 7 Asymmetry analysis improved consistency and sensitivity by overcoming variability caused by differences in cortical anatomy between patients.8 9 Electroencephalogram (EEG)-based scaling of amplitudes used in multifocal objective perimetry greatly reduced intersubject variability by accounting for external factors (skull thickness, underlying tissue conductivity), possibly resulting in considerably better sensitivity.10 Recently, 100% sensitivity in early glaucoma has been reported with a newly developed blue-on-yellow pattern-onset stimulus.11
However, the technique as typically applied is relatively time consuming and requires 20 to 30 minutes per patient, primarily because recordings are conducted monocularly. Another disadvantage of sequential monocular recording is unequal psychophysical conditions during recordings of each eye. As was reported previously, asymmetry analysis of mfVEP amplitude is the most sensitive way to detect early changes.8 9 However, toward the end of the test, the patient may become fatigued or lose attention, which may negatively affect recording of the second eye, producing artificial asymmetry and therefore false-positive results. Simultaneous recording of both eyes would, therefore, be beneficial for decreasing recording time and improving asymmetry analysis.
Recording of binocular (dichoptic) mfVEPs has been recently reported by James et al.12 13 14 15 using a liquid crystal polarizing shutter with associated polarizing spectacles. Although it did permit simultaneous recording from multiple visual field areas of both eyes, this stimulation technique caused considerable reduction in the level of luminance contrast, resulting in lower amplitude of responses and, consequently, smaller dynamic range of the response, limiting clinical applicability.
We recently reported successful simultaneous binocular (dichoptic) recording of mfVEP using virtual reality goggles for stimulus display.16 This stimulation technique does not require the patient to wear polarizing glasses, and it presents true simultaneous recordings with both eyes continuously receiving stimuli as opposed to alternating ferro-electric shutters. To record dichoptic mfVEP using virtual reality goggles, different pseudorandom binary sequences were used to stimulate test locations in both eyes, and responses for multiple visual field areas in both eyes were derived from a single cortical signal. It yielded amplitude comparable to monocular stimulation and accurate topographic representation of the visual field.
The aims of the present study were to determine the ability of dichoptic mfVEP using virtual reality goggles to detect glaucomatous visual field loss and to test whether the potential advantages hypothesized earlier—namely, decreased testing time and lower intereye variability—can be demonstrated.
| Methods |
|---|
|
|
|---|
The diagnosis of glaucoma was made on the basis of glaucomatous cupping of the optic disc, as judged by stereoscopic ophthalmoscopy with a corresponding visual field defect on Humphrey visual field analysis (HFA). All patients fulfilled the following inclusion criteria: best-corrected visual acuity greater than 6/18 in the worse eye; reliable and repeatable visual field defects on Swedish Interactive Threshold Algorithm (SITA) standard 24–2 perimetry, which corresponded to disc excavation; and absence of other ocular abnormalities (except glaucoma). A scotoma on Humphrey field examination was defined as a set of three contiguous abnormal points in the same hemifield at P < 5%, with at least two points at P < 2% on the pattern deviation plot. Points immediately above and below the blind spot could not qualify as part of the scotoma. Peripheral rim points could qualify as part of the overall scotoma, but at least two of the points qualifying as the scotoma nucleus had to be non-rim. Patients with significant cataract or other media opacities, anisometropia greater than 1.5 D, and inconsistent or nonrepeatable visual field defects were excluded. Significant cataract was defined as cataract causing visual acuity to drop below 6/12 after best correction.
In addition, 30 healthy subjects were recruited to construct normative databases for monocular and binocular tests. All had best-corrected visual acuity of 6/6 or better in both eyes, anisometropia (if any) less than 1.5 D, stereoacuity of 40 seconds of arc on Titmus fly stereogram, and normal ophthalmic examination results. All subjects underwent mfVEP testing on monocular and dichoptic stimulation, and the order of tests was randomized.
Dichoptic Testing
Instrument Set-up.
Stimuli were presented simultaneously to both eyes using virtual reality goggles (nVisor SX; NVIS, Reston, VA), which are based on liquid crystal on silicon (LCOS) technology and have a 60-Hz display refresh rate for each screen, driven by separate video boards. The goggles screens were removed from their head mounts (provided by the manufacturer) and mounted on a chin-rest set-up to facilitate placement of the recording electrode cross on the occipital region. Figure 1 shows the instrument set-up used. The subject was seated comfortably in front of the machine, with chin placed on the chinrest, forehead against the forehead support, and full refractive correction in place for distance. The patient viewed the binocular display through two separate eyepieces. Vertical bars were marked to indicate the level of lateral canthus used to align the eyes vertically with the goggles screens. The distance between eyepieces was adjusted to match each persons interpupillary distance.
|
|
Monocular Testing
Instrument Set-up.
Monocular black-white multifocal VEPs were recorded for each eye (right eye first) using multifocal objective perimetry (Accumap, version 2.0; ObjectiVision Pty. Ltd., Sydney, Australia), described in detail elsewhere.2 Sparse stimulus presentation was also used for monocular stimulation. The dartboard stimulus was the same as that used for each eye of the dichoptic tests. The stimulus was generated on a 21-inch high-resolution display (Hitachi Ltd., Tokyo, Japan) with stimulation rate of 75 Hz. All subjects were given optimal refraction for near and were seated 30 cm from the screen.
Stimulus.
For monocular recording, each recording sequence had a duration of nine frames of the monitor (13.3 x 9 ms = 119.7 ms). Each element "1" of the binary sequence was represented by a pattern on state that lasted for two frames and a pattern off state that lasted for seven frames. For element "0" of the stimulating sequence, the pattern off state was active for 9 frames. Five runs were recorded for each eye, each run lasting 56 seconds.
Recording
For monocular and dichoptic tests, four gold cup electrodes (Grass, West Warwick, RI) mounted in an occipital cross-electrode holder were used for bipolar recording. Two electrodes were positioned 4 cm on either side of the inion: one electrode was in the midline 2.5 cm above the inion, and one electrode was 4.5 cm below the inion. Electrical signals were recorded along four channels as the difference between superior and inferior, left and right, and obliquely between left and inferior and right and inferior electrodes. A ground electrode was placed on one ear lobe. Cortical responses were amplified 100,000 times and band-pass filtered (1–20 Hz). Software (Opera; ObjectiVision Pty. Ltd.) correlated the responses with the stimulating PRBS and attributed the calculated signals to the respective segments of the visual field. This software also scaled the responses to the background EEG to reduce the interindividual variability, described elsewhere.10 For every segment, the largest peak-to-trough amplitude for each wave within the interval of 60 to 200 ms was determined for each channel. The wave of maximal amplitude from each segment in the field from the four channels was automatically selected, and the software created a combined topographic map.2
Statistical Analysis
For monocular and dichoptic mfVEP, this map was compared with a normal database of 30 healthy subjects, and the visual field defects were documented by constructing amplitude probability plots. The difference in intereye amplitude at each segment was also calculated, and this was compared with the expected difference in the normal population. Areas of visual asymmetry (or visual loss) were documented by constructing an asymmetry probability plot.8 A scotoma on the amplitude deviation plot was defined as a cluster of three abnormal points in the same hemifield at P < 0.02, with at least one of them at P < 0.01. For the asymmetry plot, a cluster of three contiguous points at P < 0.01 or two points at P < 0.005 was considered a scotoma. The four central superior rim points were excluded from the clusters.
Analysis was based on detection of scotomas by HFA, monocular and dichoptic mfVEP, and topographic correspondence between scotomas. To compare asymmetry obtained with monocular and dichoptic recording, the degree of asymmetry expressed as the RAC between the eyes of each subject was calculated as the difference in amplitude between the left eye and the right eye divided by the sum of amplitudes at each segment.
| Results |
|---|
|
|
|---|
Glaucoma
Twenty-eight persons with glaucoma were examined. Eighteen were men and 10 were women, and the mean age was 67.4 ± 8.03 years (52–80 years). Three of 28 patients had bilateral visual field defects on HFA; the rest had unilateral defects. The affected eye was chosen for patients with unilateral glaucoma, and one eye was chosen at random for patients with bilateral field defects. Fourteen patients had mean deviation (MD) less than 7 dB on HFA, 12 had MD 7 to 14 dB, and two had MD 14 to 16 dB. Therefore, most patients had early or moderate glaucoma, 24 patients had best-corrected visual acuity 6/9 or better in the worse eye, and the remaining four had vision of 6/12 to 6/18 in the worse eye. All patients had stereoacuity better than 200 seconds of arc.
In all 28 eyes with glaucomatous defects, monocular and dichoptic mfVEP identified defects, all which corresponded topographically to the scotomas on HFA. An example is shown in Figure 3 .
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
In the present study of dichoptic mfVEP in glaucoma, virtual reality goggles were used to simultaneously stimulate multiple areas of the visual field in both eyes up to 18° of eccentricity. All HFA scotomas were identified by dichoptic mfVEP (100% sensitivity). Although the severity of the visual field defect varied considerably in the study group, 50% of patients had MD less than 7 dB, indicating good sensitivity of the technique in early glaucoma. Although glaucoma was unilateral in most patients, three patients (10%) had bilateral visual field defects, all which were detected by dichoptic mfVEP.
In addition, when individual RACs were averaged across perimetrically affected hemifields, significantly larger values were demonstrated on dichoptic compared with monocular testing. In contrast, averaged RAC values in healthy subjects were significantly smaller on dichoptic stimulation. Furthermore, in one fifth of all patients tested (five patients), dichoptic mfVEP detected scotomas in hemifields that were normal on HFA. Four of the five defects were identified by asymmetry analysis. In all five patients, rim changes corresponding to the additional defect were noted.
The results of the present study suggest that dichoptic mfVEP not only accurately detects perimetric losses identified by subjective visual field analysis, it may also identify early functional loss in areas as yet unaffected on HFA and monocular mfVEP. The latter appears to be the result of more sensitive detection of abnormal asymmetry between eyes using dichoptic mfVEP. This is an unexpected finding and has not been reported before. Although the basis of this phenomenon cannot be conclusively explained, two possible explanations are suggested.
First, asymmetry was significantly smaller among healthy controls when dichoptic stimulation was used, possibly because signals from both eyes were recorded under identical conditions (simultaneously). Tighter asymmetry among normals produces a narrower statistical range in the normative database, which may in turn enhance the ability of the test to detect more subtle asymmetry. Second, under binocular viewing conditions in the presence of normal binocular vision, a relatively less-defined image from an eye with early ganglion cell damage may be further suppressed by the brain when the contralateral eye provides a better image. The traces shown in Figure 4 demonstrate this effect. Traces are suppressed in most segments on dichoptic stimulation compared with monocular testing. The other eye was normal (not shown in the figure).
Compared with monocular mfVEP, dichoptic stimulation reduces testing time by approximately 30% (7 minutes for dichoptic compared with 10 minutes for monocular testing). All patients reported good acceptance. None of the patients tested had untoward eye strain or diplopia associated with simultaneous binocular viewing. The use of identical rotating fixation targets and stationary intersegment background, together with random, briefly appearing pattern-onset stimuli, may facilitate comfortable fusion of the images presented to both eyes. Although most studies of mfVEP in glaucoma have used pattern-reversal stimulation, we chose to use pattern-onset stimulation for monocular and dichoptic tests to maintain methodological consistency. In addition to methodological issues, another advantages of pattern-onset stimulation is the production of larger signals in the central part of the field compared with pattern-reversal stimulation.21 This, combined with scaling of signals based on background EEG activity and the use of specialized algorithms for enhanced noise detection and filtering available in the multifocal objective perimetry (Accumap; ObjectiVision Pty. Ltd.) machine, facilitates very good signal-to-noise ratios, with 5-minute per eye recordings for monocular tests and 7.5-minute recordings for dichoptic tests.
The technique of dichoptic mfVEP is, however, not without potential limitations. Persons with disorders of binocular vision such as amblyopia or microtropia may demonstrate suppression scotomas on dichoptic testing that would confound the results of testing for another ocular abnormality. Because the advantage of the technique is based on detection of the asymmetry of damage, it may also not be as useful for early symmetrical glaucoma when both eyes are affected to relatively the same extent.
In conclusion, this study demonstrates the ability of dichoptic mfVEP to detect functional loss in glaucoma. It exploits the superior sensitivity of binocular vision to subtle unilateral defects and demonstrates its possible role as a new tool to detect early glaucomatous damage while also reducing testing time. Detailed investigation into the repeatability of this technique and its diagnostic performance in early glaucoma, in preperimetric glaucoma, and in the early detection of progression are required and will be the subject of future studies.
| Footnotes |
|---|
Submitted for publication March 15, 2007; revised April 29, 2007; accepted July 30, 2007.
Disclosure: H. Arvind, None; A. Klistorner, ObjectiVision Pty. Ltd. (R, P, C); S. Graham, ObjectiVision Pty. Ltd. (R, P, C); J. Grigg, None; I. Goldberg, None; A. Klistorner, None; F.A. Billson, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked "advertisement" in accordance with 18 U.S.C.
1734 solely to indicate this fact.
Corresponding author: Hemamalini Arvind, Save Sight Institute, 8, Macquarie Street, Sydney, NSW-2000 Australia; hemamalini_arvind{at}yahoo.com.